Selectable: 100-120 VAC, 50/60 HZ, 6 A typical; 200-240 VAC 50/60 HZ, 3 A typical; single phase
Motor thermal protection
Automatic thermal protection
Operating speed
50 Hz: 1450 RPM
60 Hz: 1750 RPM
Cooling system
Air-cooled
Weight
Pump only: 24.5 kg (54 lbs)
Shipping weight: 28.1 kg (62 lbs)
Leak rate (with exhaust sealed)
<1 x 106 secs helium
Noise Level (per ISO 11201)
52 +/- 2 dB(A)
Vibration level at inlet (per ISO 10816-1)
Class 1B, 7 mm/s
Fuse Type
M5 x 20 mm, 10 A, Slo-Blo, Littelfuse Model 215010 or equivalent
Hour meter
Integral Hour Meter provided, displays running time to closest 1 hour
Emissions and Immunity
EN 61236-1:2006 as a Class A device
Safety
EN/IEC 61010 3rd Edition CSA certified for US and Canada
Ambient Relative Humidity
5-95%; non-condensing
Water Vapor Handling
120 g/hr
Installation
Indoor use, Installation Category II, Pollution Degree 2
Altitude
2000 m (6562’)
*If the pump and/or vacuum system is new, or has been exposed to atmosphere for an extended period of time, it is possible that water can be present which can affect the pump’s ability to achieve ultimate pressure. To remove water from the system it is recommended to open GB1 for 15 minutes and/or flush with dry nitrogen. **Mains supply voltage fluctuations shall not exceed 10% of the nominal voltage.
Table 4 Troubleshooting Chart
Problem
Possible Causes
Corrective Actions
Will not start
Fuse blown
Replace fuse.
Identify cause of overload.
Check the line voltage and the voltage configuration of the pump.
Motor thermal protector open
Allow motor to cool.
Identify cause of overload.
Excessive voltage drop
Check size and length of cable.
Defective motor
Inspect. Contact Agilent.
Poor ultimate pressure
System leak
Locate and repair leak.
Water in pump
Flush pump with air or dry nitrogen.
Gas ballast plugged
Replace breather vent. Contact Agilent.
Solvent in pump
Flush pump with air or dry nitrogen.
Install trap or filter.
Seals worn out
Rebuild pump.
Poor conductance to pump
Replumb with shorter and/or larger diameter tubing.
Hammering noise
Pump overheated
Check ventilation to pump.
Check ambient temperature.
Debris in pump
Check intake screen.
Flush pump.
Disassemble pump and inspect.
Pump runs intermittently
Motor thermal protector is cycling open and closed.*
Allow the motor to cool.
Identify the cause of the overload.
Service, Maintenance, Tip Seal, or Part Replacement
Wear or damage
See Table 5 on page 18 and Table and Table 6 on page 18.
* The IDP-7 is equipped with an auto-reset thermal motor protector. This protector will automatically shut down the pump when it detects an overload condition, and will automatically restart the pump when the motor has cooled to within an acceptable temperature range.
Table 5 Accessories and Replacement Kits
Part Number
Description
Contents
X3807-67002
Front Cowling Assembly
Replacement front cowling material to replace a broken plastic cowling.
X3807-67003
Rear Cowling Assembly
Replacement rear cowling material to replace a broken plastic cowling.
X3807-68008
Gas Ballast Kit
Spare Parts
RETURN THE COMPLETED REQUEST FOR RETURN FORM TO YOUR NEAREST LOCATION:
EUROPE:
NORTH AMERICA:
PACIFIC RIM:
Fax: 00 39 011 9979 330
Fax: 1 781 860 9252
please visit our website for individual office information
Fax Free: 00 800 345 345 00
Toll Free: 800 882 7426
http://www.agilent.com
Toll Free: 00 800 234 234 00
vpl-ra@agilent.com
vpt-customercare@agilent.com
Table 3 Typical IDP-7 Maximum Starting Current (A)
Frequency
High Voltage Range 100 V
120 V
127 V
200 V
230 V
240 V
60 Hz
22
25
25
12
N/A
14
50 Hz
23
N/A
N/A
12
14
15
Engineering Facts
14 facts · every claim linked to a page
Selection
The pump is only suitable for air or inert gases and must not be used with corrosive, explosive, or particulate-forming gases.
Process gas compatibility determines whether this pump can be selected for the application at all.
Source
“This pump is suitable for pumping air or inert gases. The pump is not intended to pump corrosive, explosive, or particulate-forming gases.”
Installation
The pump must be powered through an IEC-320 cord rated at least 10 A.
Undersized power cords can overheat or cause voltage drop during startup.
Source
“Connect the pump to the power supply using an IEC-320 style power cord of at least 10 A capacity”
Installation
The pump requires at least 60 cm clear space on all sides for ventilation, maintenance access, and breaker access.
Insufficient clearance causes overheating and prevents safe access to the power disconnect.
Source
“There must be a clear space of at least 60 cm (24") on all sides of the equipment.”
Installation
The pump is delivered configured for high voltage and must be reconfigured for low-voltage mains.
Connecting a high-voltage-configured pump to low-voltage mains will cause malfunction or damage.
Source
“The pump as delivered from the factory is configured for high voltage.”
Installation
The pump must be grounded through a properly installed grounding plug; adapters are not permitted.
Proper grounding is required to reduce electric shock risk in case of short circuit.
Source
“This product must be grounded.”
Installation
Extension cords, if used, must be at least 16-gage wire and no longer than 25 feet (7.6 m).
Undersized or overly long extension cords cause voltage drop and overheating.
Source
“Agilent recommends using only extension cords with a minimum of 16-gage wire and a maximum length of 25 feet (7.6 m).”
Performance
Maximum starting current can reach 25 A at 120 V, so supply and protection must be sized accordingly.
Inrush current affects breaker selection and verifies the supply can handle startup without nuisance trips.
Source
“Table 3 lists the typical IDP-7 maximum starting current and duration.”
Installation
Use short, large-diameter NW25 or larger inlet hardware and a bellows to isolate vibration and relieve strain.
Long or narrow inlet lines reduce effective pumping speed and transmit vibration to the chamber.
Source
“Insert a bellows between the pump and vacuum chamber to provide both vibration isolation and strain relief.”
Installation
Do not restrict exhaust flow with long small-diameter tubing; use short NW16 or larger hardware to avoid overheating.
Exhaust restriction raises back pressure and can overheat the pump.
Source
“To avoid overheating the pump, do not restrict the exhaust flow with long lengths of small diameter tubing.”
Operation
Gas ballast is required when pumping water vapor or condensable gases; both ports together handle up to 120 g/hr.
Exceeding water vapor handling capability shortens tip seal life and degrades ultimate pressure.
Source
“With both gas ballast 1 and gas ballast 2 in use, up to 120 g/hr water vapor can be handled.”
Operation
To reach maximum vacuum, the pump must run at least 24 hours with the inlet sealed.
Acceptance testing before this stabilization period may falsely indicate poor performance.
Source
“In order to reach maximum vacuum, the pump must be left running for at least 24 hours with the inlet sealed.”
Maintenance
Newly installed tip seals require several hours of run time to seat and meet speed and base pressure specifications.
Performance should not be judged immediately after tip seal replacement.
Source
“Newly installed Tip Seals may require several hours of run time to seat properly and enable the pump to meet speed and base pressure specifications.”
Safety
The outer surface can exceed 60 °C; allow the pump to cool before maintenance.
Hot surfaces present a burn hazard during maintenance after extended operation.
Source
“allow it to cool as the temperature of the outer surface may be in excess of 60 °C.”
Safety
Always disconnect electrical power before beginning maintenance work.
Prevents electric shock and accidental start during service.
Source
“Always disconnect your power supply to the pump before beginning maintenance work.”