Protector Premier Laboratory Hoods User Manual

Technical Reference Summary

Covered Models & Document Scope

Labconco Protector® Premier® Laboratory Fume Hoods, including standard vertical-rising sash models, A-Style Combination Sash models, by-pass block models, Eco Energy Saving Air Foil and Flush Air Foil variants, 4', 5', 6' and 8' wide hoods, and Integral Blower models (4' integral blower uses 10" exhaust). Optional accessories covered include Guardian Airflow Monitors, sash stops, ceiling enclosures, rear panels, additional service fixtures, distillation grids, upper bypass, electrical duplex outlets, and 8' dual exhaust adapter. Appendix D covers the Protector I-S Hood with Intelli-Sense Blower 3-speed operation.

Section-by-Section Technical Breakdown

pp. 9-12
Prerequisites: Location, Support, Exhaust, Electrical and Service Line Requirements

Specifies site preparation requirements: locate the hood away from traffic, doors, windows, fans and ventilation registers; installations must be permanent and stationary on a base cabinet and chemically resistant work surface. Exhaust connection is 12" nominal (12.75" OD) duct for 60–100 fpm face velocity and 1000–2500 fpm transport velocity. Electrical service is normally 115 V 50/60 Hz 20 A or 230 V 50/60 Hz 10 A. Service lines are ¼" OD copper (brass for natural gas) with 40 PSI recommended and 200 PSI maximum pressure.

  • Do not install on a cart, dolly or mobile bench; all installations must be permanent and stationary.
  • Exhaust duct designed for 12" nominal (12.75" OD) pipe; 4' integral blower models use 10" (10.75" OD).
  • Face velocity range 60 to 100 fpm; transport velocity away from hood 1000 to 2500 fpm.
  • Electrical: 115 V, 50/60 Hz, 20 A or 230 V, 50/60 Hz, 10 A; dedicated branch circuits recommended.
  • Service lines: ¼" OD copper (brass for natural gas), recommended operating pressure 40 PSI, maximum 200 PSI.
pp. 13-16
Installation: Unpacking, Shipping Skid Removal, Sash Weight Release and Hood Mounting

Covers unpacking and inspection, removal of the shipping skid, release of the sash weight, and mounting the hood on a supporting structure and work surface. Hoods weigh 400 to 800 lbs (182–363 kg). The work surface must be at least as wide as the hood, aligned flush with the back of the hood, and supported by a cross support installed after leveling the base cabinets and before placing the hood.

  • Leave hood attached to shipping skid until as close to final location as possible; move with floor jack or furniture dolly, never by tilting onto a hand truck.
  • Sash weight is individually matched to the specific hood and must not be exchanged with another unit.
  • Work surface depth 30", hood depth 31.7", hood internal depth 24", filler panel depth 8".
  • Install cross support after base cabinets and work surface are leveled and before installing the hood.
  • Do not lift by the front air foil.
pp. 17-21
Connecting Exhaust, Electrical Supply and Service Lines

Describes connecting the hood to the exhaust system, the electrical supply, and the service lines, plus sealing the hood to the work surface and certifying face velocity. Exhaust ductwork weight must be supported independently of the hood. Electrical supply must include overload protection with a marked disconnecting device in close proximity. Service fixtures are plumbed from valve to hose connector or gooseneck; supply tubing is provided by the installer.

  • Exhaust connection should be installed by a qualified HVAC contractor; duct material must match chemicals used.
  • Building electrical supply must include overload protection; switch or circuit breaker marked as disconnecting device and within easy reach.
  • Ground to MAINS protective earthing lug only; 115 V: black hot, white neutral; 230 V: brown Phase 1, blue Phase 2.
  • Maximum combined load if circuits wired together is sum of outlets plus unit rating (e.g. 2 A remote, 8 A integral blower), not exceeding 10 A or 20 A service.
  • Service fixtures support air, hot water, vacuum, cold water and natural gas; natural gas only in fixture pre-plumbed with brass tubing.
  • Do not use oxygen with any standard service fixture.
  • Seal hood to work surface with silicone sealant; cupsinks sealed with silicone to prevent leaks.
  • Certify average face velocity at sash opening to 60–100 fpm; factory ASHRAE 110-1995 rating AM<0.05 ppm at 4 lpm.
pp. 22-25
Performance Features and Safety Precautions

Describes the hood's performance features including by-pass airflow, sash travel, interior dimensions, LED lighting, service fixture capacity, duplex receptacles, Guardian airflow monitor preparation, and optional sash and blower configurations. Safety precautions include verifying blower operation and face velocity before use, and warnings for hot blower motor surface and flammable gas.

  • Sash provides 37.5" high visibility while limiting travel to 28" operating height.
  • Interior height 48"; by-pass airflow design maintains containment as sash moves.
  • Factory prepared for up to 8 service fixtures; duplex receptacles mounted on right and left corner posts.
  • LED T8 bulbs rated 50,000 hours and use approximately half the power of fluorescent bulbs.
  • Guardian Digital Airflow Monitor or Guardian Airflow Monitor optional; right corner post factory prepared.
  • Sash options: smooth anti-racking cable, chain, auto-return, or Intelli-Sash.
  • Exhaust connection 12" (12.75" OD); 4' integral blower models use 10" (10.75" OD).
  • Blower motor can attain temperatures that could cause burns.
pp. 28-30
Maintaining Your Protector Premier: Routine Maintenance Schedule and Service Operations

Chapter 6 covers service safety precautions, a routine maintenance schedule, and routine service operations for the Protector Premier fume hood.

  • Routine maintenance schedule and routine service operations are defined in Chapter 6.
  • Service safety precautions apply before performing maintenance.
pp. 31-36
Modifying Your Protector Premier: Work Surfaces, Enclosures, Panels, Fixtures and Accessories

Chapter 7 describes field installation of work surfaces, ceiling enclosures above the fume hood, rear panels behind the fume hood, additional service fixtures, Guardian Airflow Monitors, distillation grids, sash stop kits, upper bypass, electrical duplex outlets, and the 8' dual exhaust adapter.

  • Work surfaces, ceiling enclosures, rear panels, additional service fixtures, Guardian Airflow Monitors, distillation grids, sash stop kits, upper bypass, electrical duplex outlets and 8' dual exhaust adapter are field-installable.
  • Guardian Airflow Monitors are installed per Chapter 7.
pp. 37-39
Troubleshooting

Provides a table of problems, probable causes and corrective actions for the fume hood, covering blower and light operation, contamination, sash operation, electrical duplex outlets and service valves.

  • Problems covered include remote blower and lights not operating, blower-only or light-only failures, contaminants outside the hood, sash smoothness issues, duplex outlet power loss and service valve failures.
  • Corrective actions include checking wiring and switches, resetting circuit breakers, replacing belts, motors, lamps, ballasts, cables, pulleys and valves.
  • Hood should have average face velocity of 60–100 fpm.
pp. 40-47
Components, Dimensions and Specifications (Appendices A–C)

Appendix A contains labeled component diagrams, Appendix B contains dimensional diagrams, and Appendix C contains electrical requirements and wiring diagrams for the Protector Premier fume hood.

  • Appendix A: labeled diagrams of all fume hood components.
  • Appendix B: comprehensive dimensional diagrams for the laboratory fume hoods.
  • Appendix C: electrical requirements and wiring diagrams.
pp. 48-55
Protector I-S Hood and Intelli-Sense Blower

Appendix D provides information relating to the automatic Intelli-Sense Blower 3-speed operation for the Protector I-S Hood.

  • Covers automatic Intelli-Sense Blower 3-speed operation.
  • Includes current rating information referenced by caution notes.
pp. 65
VAV Open Area Guidelines

Appendix G provides guidelines to properly assess fume hood effective area whenever variable air volume (VAV) mechanical systems are installed, allowing the VAV system to be programmed at various sash positions.

  • Used to assess fume hood area opening for VAV system programming at various sash positions.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Aplica a
Exhaust duct connection is designed for 12" nominal pipe (12.75" OD); 4' integral blower models use 10" (10.75" OD). p.10 ↗ Size exhaust ductwork and blower transitions to these connection sizes to minimize static pressure loss. Protector Premier series; 4' integral blower models use 10"
Recommended average face velocity at the sash opening is 60 to 100 fpm. p.21 ↗ Certify and periodically verify face velocity within this range for safe containment. Whole series
Transport velocity away from the hood should be in the 1000 fpm to 2500 fpm range. p.10 ↗ Select blower and duct size to maintain transport velocity in this range. Whole series
Electrical service is normally 115 V, 50/60 Hz, 20 A or 230 V, 50/60 Hz, 10 A. p.18 ↗ Provide a dedicated branch circuit matching the hood I.D. plate; verify voltage behind front panel. Protector Premier series
If circuits are wired together, maximum load is the sum of individual outlets plus unit rating (e.g. 2 A remote, 8 A integral blower), not to exceed the 10 A or 20 A service. p.18 ↗ Do not exceed the service rating when combining outlet and blower loads on one circuit. Protector Premier series
Service lines should be ¼ inch outside diameter copper (brass for natural gas), with recommended operating pressure 40 PSI and maximum allowable pressure 200 PSI. p.12 ↗ Use a pressure regulator to reduce line pressure to 40 PSI where necessary; provide an accessible shut-off valve. Hoods with service fixtures
Hood models weigh between 400 and 800 lbs (182–363 kg). p.13 ↗ Verify supporting structure and work surface can safely support the combined weight of hood and related equipment. Protector Premier series
Work surface depth 30", hood depth 31.7", hood internal depth 24", filler panel depth 8". p.16 ↗ Provide a work surface at least as wide as the hood and aligned flush with the back of the hood for proper bypass airflow. Protector Premier series
Sash provides 37.5" high visibility while limiting travel to 28" operating height. p.22 ↗ Design working height and sash stop accessories around the 28" maximum operating opening. Standard vertical-rising sash models
Interior height is 48". p.22 ↗ Account for 48" interior height when planning apparatus and distillation setups. Protector Premier series
All hoods are factory prepared for up to 8 service fixtures. p.23 ↗ Plan service fixture additions up to the factory-prepared maximum of 8. Whole series
LED T8 bulbs are rated 50,000 hours and use approximately half the power of fluorescent bulbs. p.23 ↗ Consider LED lighting for energy savings and reduced lamp replacement frequency. Protector Premier series
Factory ASHRAE 110-1995 rating is less than 0.05 ppm at 4 liters per minute (AM<0.05). p.21 ↗ Use as manufactured containment baseline; field ASHRAE testing requires a certified on-site contractor. Whole series
Natural gas should be used only in the service fixture pre-plumbed with brass tubing. p.21 ↗ Do not connect natural gas to standard copper supply lines due to sulfur content deterioration. Hoods with natural gas service fixture
Do not use oxygen with any standard service fixture. p.20 ↗ Contact Labconco Customer Service for oxygen fixture information before any oxygen connection. Hoods with service fixtures
Exhaust ductwork weight must be supported independently of the hood superstructure. p.17 ↗ Provide independent duct supports to avoid damage to the hood structure. Whole series
The sash weight is individually matched to the specific hood and should not be exchanged on any other unit. p.15 ↗ Keep sash weights with their original hood during installation and service. Whole series
The work surface should be at least as wide as the hood and aligned flush with the back of the fume hood. p.15 ↗ Provides correct spacing under the air foil for proper bypass airflow. Whole series

Faults & Troubleshooting

Código / Síntoma Possible Cause Remedy Note
Remote blower and lights won’t operate p.41 ↗ Wires not connected at junction boxes or switches. Check connection of switches. Check connection to control box on top of unit. Chapter 8
Remote blower and lights won’t operate p.41 ↗ Circuit breakers tripped in building electrical supply. Reset circuit breakers. Chapter 8
Remote blower won’t operate, but lights work p.41 ↗ Blower wiring is disconnected. Inspect blower wiring and switch. Chapter 8
Remote blower won’t operate, but lights work p.41 ↗ Belt broken. Replace belt. Chapter 8
Remote blower won’t operate, but lights work p.41 ↗ Blower motor is defective. Replace blower motor. Chapter 8
Fume hood blower operates but lights will not operate p.41 ↗ Lamp not installed correctly. Inspect lamp installation. Chapter 8
Fume hood blower operates but lights will not operate p.41 ↗ Lamp is defective. Replace lamp. Chapter 8
Fume hood blower operates but lights will not operate p.41 ↗ Lamp circuit breaker in building is tripped. Reset the lamp circuit breaker. Chapter 8
Fume hood blower operates but lights will not operate p.42 ↗ Lamp wiring is disconnected. Inspect lamp wiring. Chapter 8
Fume hood blower operates but lights will not operate p.42 ↗ Defective lamp ballasts. Replace lamp ballasts. Chapter 8
Contaminants outside of fume hood p.42 ↗ Improper user techniques for the fume hood. See “Certifying the Hood” Chapter 3 and “Safety Precautions” Chapter 4 sections in the manual. (Ref. Appendix E) Chapter 8
Contaminants outside of fume hood p.42 ↗ Restriction of the baffle air slots or blockage of the exhaust outlet. Remove baffles to ensure that all air slots, and the exhaust outlet are unobstructed. Chapter 8
Contaminants outside of fume hood p.42 ↗ External factors are disrupting the fume hood airflow patterns or acting as a source of contamination. See “Location Requirements” Chapter 2, “Certifying the Hood” Chapter 3, and “Safety Precautions” Chapter 4 sections of this manual. (Ref. Appendix E) Chapter 8
Contaminants outside of fume hood p.42 ↗ Fume hood has improper face velocity. Have fume hood re-certified and check remote blower exhaust system. Hood should have average face velocity of 60-100 fpm. Chapter 8
Vertical sash no longer operates smoothly p.42 ↗ Cable is frayed or plastic protection is damaged. Inspect cable and replace cable if worn or damaged immediately; otherwise injury could result. Chapter 8
Vertical sash no longer operates smoothly p.42 ↗ Pulley bearing is damaged. Replace pulley, bearing or add grease. Chapter 8
Vertical sash no longer operates smoothly p.42 ↗ Cable or chain has slipped off the pulleys or sprockets. Re-install, cable or chain must be replaced immediately if damaged. Chapter 8
Vertical sash no longer operates smoothly p.42 ↗ Weight has broken pulleys or sprockets. Replace pulleys or sprockets. Chapter 8
Combination A-Style sash no longer operates smoothly p.42 ↗ Horizontal glass panels have come off the tracks. Re-install horizontal glass on tracks. Chapter 8
Combination A-Style sash no longer operates smoothly p.43 ↗ Vertical sash frame is distorted. Place horizontal glass symmetrically and pull sash down to air foil. Straighten damaged frame. Chapter 8
Combination A-Style sash no longer operates smoothly p.43 ↗ Cable is frayed or has slipped off the pulleys. Re-install, cable must be replaced immediately if damaged. Chapter 8
Electrical duplex outlets no longer have power p.43 ↗ Wires not connected or faulty duplex. Check wire connection or replace duplex. Chapter 8
Electrical duplex outlets no longer have power p.43 ↗ Circuit breakers tripped in building electrical supply. Reset circuit breakers. Chapter 8
Service valves no longer operate p.43 ↗ Faulty building supply. Inspect building supply shut off valves and appropriate pressures below the recommended 40 PSI. Chapter 8
Service valves no longer operate p.43 ↗ Valve no longer operates. Replace valve and check for leaks. Chapter 8
Service valves no longer operate p.43 ↗ Supply line or outlet line has leaks. Inspect line for leaks and fix any leaking plumbing connections. Chapter 8