Solid state laser with on-board thermal-electric cooling and guaranteed thermal stability and beam quality
Laser Power
100 mW
Laser Beam Profile
10 x 60 µm elliptical beam
Optical Alignment Procedure
Fixed, no operator alignment required
Fluorescence Detection
Silicon Photomultiplier (SiPM) with high photon detection efficiency; Individual photodetector for each channel
FSC/SSC Sensitivity
FSC: 0.4 µm; SSC: 0.1 µm
Fluorescence Resolution
<3% CV for CEN
Optical Filters
User exchangeable, “Smart” filter automatically read by the system
Cell Size
0.1 - 50 µm
Fluidics
Flow Cell
170 x 290 µm rectangular quartz flow cell
Sample Acquisition Rate
100,000 events/second
Sample Delivery
Positive-displacement syringe pump enabling direct absolute count without the need for reference counting beads.
Volumetric Absolute Count Precision
< 5%
1 Сноска в тексте страницы не раскрыта.
Table 16 NovoCyte Workstation and Monitor Specifications
Specification
NovoCyte Workstation
Small form factor (SFF). Intel Core i7 CPU or above, 16G RAM or above, 512G sold state drive or above.
Monitor
High resolution flat panel with resolution of 1,920 x 1,080 or above
Computer Operating System
Microsoft Windows 10 Professional (64 bit) or above with Microsoft Office and NovoExpress software pre-installed
Table 10 Unscheduled Maintenance and When to Perform
Unscheduled Maintenance
When to Perform
Debubble
• The CV of the data is larger than normal. • QC test failed. • No events showed up after sample collection started for certain period of time. • Instrument was not used for over one week.
Clean the External Surface of the Instrument
Keep the external surface of the instrument clean.
Decontamination
When suspecting contamination in the system.
Prime Fluidics System before Long-Term Storage
No plan to run the instrument for more than two weeks.
Prime Fluidics System after Long-Term Storage
Instrument has not been used for more than two weeks.
Purge Fluidics System before Shipment
The instrument needs to be shipped.
Calibrate the NovoCyte Fluidics Station II or NovoCyte Fluidics Cart
Abnormal warning about the liquid level in the reagent containers is observed.
Clean the Optical Filters
The optical filters get dirty.
Clean Fluidics System
• The CV of the data is larger than normal. • QC test failed. • Excessive debris in display.
Administrative creation of individual user accounts and user groups with privilege control. Configurable roles for individual users. User operation time tracking.
Quality Control
Instrument Daily QC
Automatic daily quality control with Agilent Flow Cytometer QC Particles. Levi-Jennings Plot.
Production QC
Each instrument is verified for assembly integrity and performance to specifications.
Sampling
Manual Sample Loading
12 X 75 mm tube, 1.5 mL or 2.0 mL tube with Agilent tube holder.
Automatic Sample Loading with NovoSampler Q
Agilent Flow Cytometer 40 tube rack for 12 x 75 mm tube, 24-well, 48-well, 96-well (flat, V-, U-bottom), 384-well microtiter plates, Agilent Flow Cytometer 24 tube cooling rack for 12 x 75 mm tube, Agilent Flow Cytometer 96-well (flat, U-, V-bottom) plate cooling box.
Operating Conditions
Instrument Dimension (W × D × H)
24.4 × 18.1 × 18.8 in (62.0 × 46 ×48.0 cm) 33.5 × 18.1 × 18.8 in (85.0 × 46.0 ×48.0 cm) with NovoSampler Q
Instrument Weight
104.0 lb (47.0 kg) 133.0 lb (60.0 kg) with NovoSampler Q
No fluidic tubing disassembly or re-connection required.
Fluid Level Sensing
Active sensing using weight sensors with automated warnings when any fluid level is out of specified range.
Fluid Container Capacity
3 L sheath, 3 L waste, 500 mL cleaning, 500 mL rinse. Optional large container for sheath (15 L) and waste (15 L) with Fluidics Cart.
Carryover
< 0.1%
Sample Injection Probe (SIP) Rinse
Automated flying collar wash of inner and outer SIP surface after each sampling.
Fluidics System Monitoring
In-line pressure sensor monitors the pressure in real time. Automated system recovery when pressure is out of range due to clogging.
Fluidics Maintenance
Automated startup and shutdown with fluidic system cleaning. Automated user executable cleaning, debubble, rinse, unclog, priming and decontamination. Automated execution of user definable maintenance sequence.
Data Acquisition Software
NovoExpress
Parameters
Height and Area for FSC, SSC and all Fluorescent Channels, Width off FSC and Time.
Dynamic Range
24 bits; 7.2 decades logarithmic scale
Fluorescence Photodetector Gain Control
0-1000 User adjustable, optimized default gain setting for each individual channel.
Compensation
Full inter-beam matrix, during or post acquisition.
File Export
FCS 3.0, FCS 3.1; CSV; Batch PDF report
Data Report
Automatic report. Customizable report. Batch PDF report.
Usage Monitor
Comprehensive Transaction Log and System Log.
1 Сноска в тексте страницы не приведена.
Engineering Facts
14 facts · every claim linked to a page
Safety
The instrument must be operated only by personnel trained by Agilent or an authorized representative, so plan operator training before commissioning.
Determines who may legally and safely run the system after installation.
Source
“Agilent flow cytometers and associated accessories are limited to be operated by professionals trained by Agilent Technologies or an authorized representative.”
Installation
The unit weighs over 23 kg, so plan a two-person lift or mechanical aid during site installation.
Affects rigging, manpower and floor/site planning for installation.
Source
“This product is large and the weight is more than 51 lb (23 kg).”
Installation
Moving the instrument after installation is discouraged and requires Agilent professionals to re-install it, so fix the final location before commissioning.
Drives layout decisions and avoids costly re-installation and data-quality risk.
Source
“moving the instrument is not recommended. If you need to move the instrument, be sure to contact Agilent Technologies professionals.”
Safety
The instrument is a Class I laser product with lasers fully enclosed, so no special laser work area is needed except during service.
Simplifies lab laser-safety classification and room requirements at selection.
Source
“Agilent flow cytometer is a Class I laser product according to IEC/EN 60825-1:2014.”
Safety
Optical covers and laser shielding must never be removed by the user, so service access must be planned through Agilent only.
Defines user-permitted service boundaries and laser safety rules on site.
Source
“Modification or removal of the optical covers or laser shielding could result in exposure to hazardous laser radiation.”
Installation
Reagent containers are color-coded and position-coded; wrong placement risks damaging pressure sensors, so verify labeling during setup.
Prevents fluidics station damage and misconnection during installation.
Source
“Please ensure that the container position in the NovoCyte Fluidics Station II is correct”
Installation
Fluidics tubing is color-coded by function (blue rinse, green sheath, yellow clean, red waste), so connect ports accordingly.
Ensures correct plumbing hookup and avoids cross-contamination at install.
Source
“NovoRinse is the rinsing solution and is connected by the blue tubing.”
Operation
Optical configuration should be verified before each experiment and can be changed by swapping filters, requiring a QC test to take effect.
Affects daily workflow and panel-to-configuration matching decisions.
Source
“It is recommended for users to view and verify the current optical configuration before each experiment.”
Operation
If a QC channel is Acceptable or Failed, at least 400 µL must remain in the tube and the test repeated with automatic gain re-adjustment.
Defines minimum QC sample volume and repeat procedure for valid QC.
Source
“Ensure there is at least 400 μL sample remaining in the sample tube.”
Operation
Threshold values are sample-type dependent, with recommended FSC-H ranges given per cell type and microsphere size.
Guides initial threshold setup to avoid debris and signal loss.
Source
“setting the FSC-H threshold as 100,000 ensures the correct collection of the leukocyte signals”
Operation
For 0.1 µm microspheres the SSC gain must be increased to achieve resolution, so low-size particle work needs gain adjustment.
Sets expectations for small-particle detection performance and setup.
Source
“To detect microsphere with 0.1 µm diameter, the photodetector gain of SSC needs to be increased.”
Operation
Default auto-compensation acquisition uses 10,000 events or 20 µL, slow flow, and FSC-H threshold above 100,000.
Defines default compensation run settings engineers must know for setup.
Source
“Stop condition: total 10,000 events or 20μL sample volume.”
Operation
Compensation samples should be prepared at 1–2 × 10^6 /mL for correct automatic compensation.
Affects sample prep protocol and validity of compensation matrix.
Source
“it is recommended to prepare the compensation sample concentration to be 1 x 10^6 ~ 2 x 10^6 /mL.”
Maintenance
The sample injection probe moves downward after confirming, so fingers must be kept clear during probe cleaning.
Pinch/injury hazard during routine probe maintenance.
Source
“Sample injection probe will move downward after clicking the OK button.”