Quick Reference Guide: Gas Chromatography Detectors

Technical Reference Summary

Covered Models & Document Scope

Descripción general de los detectores de GC de Agilent: FID, NPD (N), NPD (P), µECD, TCD, FPD (S), FPD (P), NCD, SCD. Cubre la descripción, el rango dinámico lineal, la destructividad y los analitos típicos de cada detector.

Technical Data

Parameter Value Condición
Linear dynamic range >107 (±10%) p.1 ↗ FID
Destructive Yes p.1 ↗ FID
Linear dynamic range 105 (±10%) p.1 ↗ NPD (N)
Destructive No p.1 ↗ NPD (N)
Linear dynamic range >5 × 104 p.1 ↗ µECD
Linear dynamic range >104 p.2 ↗ NCD
Destructive Yes p.2 ↗ NCD
Linear dynamic range >104 p.2 ↗ SCD
Destructive Yes p.2 ↗ SCD
Relative sensitivity 10–15 g g p.2 ↗ NPD (P), 1 ppt
Relative sensitivity 10–12 g g p.2 ↗ NPD (N), 1 ppb
Relative sensitivity 10–9 g g p.2 ↗ µECD, 1 ppm
Relative sensitivity 10–6 g g p.2 ↗ FID, 0.1%
Relative sensitivity 10–3 g g p.2 ↗ TCD, %

Operating Limits

  • Detection limits and ranges depend on compound structure and analytical conditions: See GC data sheets for actual specifications p.2 ↗