Glycerophospholipids Analysis by Comprehensive HPLC Coupled with a Triple Quadrupole Mass Spectrometer
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación n.º L462 (LAAN-A-LC-E239) que describe un método integral de HPLC-ESI-MS/MS para el análisis de glicerofosfolípidos; abarca las condiciones analíticas para la separación en 1D en fase normal y en 2D en fase inversa con detección Shimadzu LCMS-8050. No se definen modelos de producto, tamaños ni códigos de pedido.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| 1D column | Nucleosil SIL (150 mm L. × 1.0 mm I.D., 3 µm) p.1 ↗ | 1D separation |
| 1D mobile phase A | Isooctane / Acetone / Ethyl Acetate / Acetic acid = 40/20/20/0.03 v/v/v/v p.1 ↗ | 1D separation |
| 1D mobile phase B | Isooctane / 2-propanol / Water / Acetic acid / 28 % Ammonia aq.sol. = 40/51/9/0.03/0.03 v/v/v/v/v p.1 ↗ | 1D separation |
| 1D flow rate | 0.02 mL/min p.1 ↗ | 1D separation |
| 1D time program | B Conc. 30 % (0 min) → 40 % (25 min) → 100 % (40 min) → 100 % (55 min) → 30 % (55.1 min) → STOP (70 min) p.1 ↗ | 1D separation |
| Column temperature | 40 °C p.1 ↗ | 1D separation |
| Injection volume | 5 µL p.1 ↗ | |
| Loop volume | 20 µL p.1 ↗ | |
| 2D column | Phenomenex Kinetex C18 (50 mm L. × 4.6 mm I.D., 2.6 µm) p.1 ↗ | 2D separation |
| 2D mobile phase A | Methanol / Water / Acetic acid / 28 % Ammonia aq.sol. = 90/10/0.05/0.05 v/v/v/v p.1 ↗ | 2D separation |
| 2D mobile phase B | 2-propanol / Acetic acid / 28 % Ammonium hydroxide = 100/0.05/0.05 V/V/V p.1 ↗ | 2D separation |
| 2D flow rate | 3.5 (50 % split to MS) mL/min p.1 ↗ | 2D separation |
| 2D time program | B Conc. 10 % (0 min) → 50 % (0.75 min) → 10 % (0.76 min) → STOP (1 min) p.1 ↗ | 2D separation |
| Detector | Shimadzu LCMS-8050 (ESI positive, MRM mode) p.1 ↗ | |
| MRM transition 16:0-16:0 PC | 734.6 > 184.1 m/z p.2 ↗ | |
| MRM transition 18:0-18:1 PC | 788.6 > 184.1 m/z p.2 ↗ | |
| MRM transition 18:0p-18:1 PC | 772.6 > 184.1 m/z p.2 ↗ | |
| Total retention time 16:0-16:0 PC | 0.0072 min p.2 ↗ | |
| Total retention time 18:0-18:1 PC | 0.013 min p.2 ↗ | |
| Total retention time 18:0p-18:1 PC | 0.013 min p.2 ↗ | |
| Correlation coefficient (R) 16:0-16:0 PC | 0.999799 p.2 ↗ | |
| Correlation coefficient (R) 18:0-18:1 PC | 0.999947 p.2 ↗ | |
| Correlation coefficient (R) 18:0p-18:1 PC | 0.999656 p.2 ↗ |
Operating Limits
- Column temperature: 40 °C p.1 ↗
- 1D flow rate: 0.02 mL/min p.1 ↗
- 2D flow rate: 3.5 mL/min (50 % split to MS) p.1 ↗
- Injection volume: 5 µL p.1 ↗
- Loop volume: 20 µL p.1 ↗