| Number of controllable units |
8
p.3 ↗ |
SCL-40 |
| Number of controllable units |
8 (Using option: 12)
p.3 ↗ |
CBM-40 |
| Number of controllable units |
4 (excluding built-in solvent delivery unit)
p.3 ↗ |
CBM-40lite |
| Event input/output |
Input: 1, Output: 2
p.3 ↗ |
SCL-40 |
| Analog signal input |
Up to two channels (option)
p.3 ↗ |
SCL-40 |
| Analog signal input |
One channel (option)
p.3 ↗ |
CBM-40 |
| Communication interface |
Ethernet
p.3 ↗ |
SCL-40 |
| Reservoir tray |
Included as standard
p.3 ↗ |
SCL-40 |
| Dimensions and weight |
W260 × D500 × H140 mm, 6 kg
p.3 ↗ |
SCL-40 |
| Dimensions and weight |
W260 × D500 × H72 mm, 5 kg
p.3 ↗ |
CBM-40 |
| Operating temperature range |
4 to 35 °C
p.3 ↗ |
SCL-40 / CBM-40 |
| Power supply |
100 to 240 V AC, 50 VA, 50/60 Hz
p.3 ↗ |
SCL-40 / CBM-40 |
| Pumping method |
Micro-volume double plunger pump
p.3 ↗ |
LC-40B X4 |
| Allowable maximum pressure |
130 MPa
p.3 ↗ |
LC-40B X4 |
| Flow rate settings range |
0.0001 to 2.0000 mL/min
p.3 ↗ |
LC-40B X4 |
| Flow rate accuracy |
±1 % (at 1 mL/min, 80 MPa, with water, stable within a temperature range of 20 to 30 °C)
p.3 ↗ |
LC-40B X4 |
| Gradient mode |
Two-solvent high-pressure gradient
p.3 ↗ |
LC-40B X4 |
| Range of gradient mixing ratio |
0 to 100 % (0.1 % step)
p.3 ↗ |
LC-40B X4 |
| Available pH range |
1 to 14
p.3 ↗ |
LC-40B X4 |
| Dimensions and weight |
W260 × D500 × H210 mm (excluding protrusions), 24 kg
p.3 ↗ |
LC-40B X4 |
| Power supply |
100 to 240 V AC, 50/60 Hz, 150 VA
p.3 ↗ |
LC-40B X4 |
| Number of selectable solvents |
Up to 4 solvents per pump (maximum 2 pumps)
p.3 ↗ |
SSU-40 |
| Number of degassed solvent lines |
5
p.3 ↗ |
SSU-40 |
| Degassed flow line capacity |
400 μL/1 line
p.3 ↗ |
SSU-40 |
| Dimensions and weight |
W260 × D500 × H72 mm, 5 kg
p.3 ↗ |