Cromatógrafo de gases Nexis™ GC-2060 con detector FID-U (2.0); comparación con el modelo anterior; modos FID: Regular Mode (con botellas de gas), Hy/Air Mode (sin botellas), Hy/Ox Mode (sin botellas); microrreactores opcionales Jetanizer™ y Polyarc™; código de documento C184-E060

Technical Data

Parameter Value Condición
Minimum Detectable Quantity (MDQ) 1.0 pg C/s p.1 ↗ GC-2060, N2/He makeup gas
Minimum Detectable Quantity (MDQ) 1.2 pg C/s p.1 ↗ Previous Model
Nozzle Materials Metal p.1 ↗ GC-2060
Nozzle Materials Metal & Quartz p.1 ↗ Previous Model
Makeup Gas Types He / N2 / Air / O2 p.1 ↗ GC-2060, cylinder-free OK
Makeup Gas Types He / N2 p.1 ↗ Previous Model
Gas supply configuration Air Compressor (+ Air Purifier) p.1 ↗ Hy/Air Mode, cylinder-free
Gas supply configuration PEAK Intura H2/O2 Generator p.1 ↗ Hy/Ox Mode, cylinder-free
Gas supply configuration Gas Cylinders (≥ 2 types) p.1 ↗ Regular Mode, with gas cylinders

Operating Limits

  • MDQ applicability: For N2/He makeup gas; results may vary depending on sample and analytical conditions p.1 ↗