Performance Evaluation of a Reactor Sampling Device for Online LC
Technical Reference Summary
Covered Models & Document Scope
Evaluación de un dispositivo de muestreo de reactor para LC en línea utilizando un Agilent 1260 Infinity II Prime Online LC System (G3167AA) con Snapdragon Chemistry SRS y Mettler Toledo EasySampler 1210; abarca instrumentación, parámetros del método y ajustes de software/hardware de SRS.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Analytical flow rate | 1.0 mL/min p.2 ↗ | |
| Solvent composition | Isocratic 15% B p.2 ↗ | |
| Column temperature | 45 °C p.2 ↗ | |
| Feed speed | 80% of analytical flow rate p.2 ↗ | |
| Injection volume | 5 µL p.2 ↗ | |
| Needle wash | 3 s, Water: ACN 1:1 + 0.1% FA (S1) p.2 ↗ | |
| Diode array detector wavelength | 273 ± 4 nm p.2 ↗ | reference: off, 20 Hz data rate |
| Draw speed | 100 µL/min p.2 ↗ | Setting 2 |
| Wait time | 3.6 s p.2 ↗ | Setting 2 |
| Dilution | 1:150, 1:200, 1:400, 1:600, 1:700 p.2 ↗ | |
| Mixing time | 10 s p.2 ↗ | |
| Cleaning volume | 1.6 mL p.2 ↗ | 2 Wash repeats |
| Sampling interval | 10 min p.2 ↗ | |
| Inlet pressure | 30 psi p.2 ↗ | on inlet of instrument |
| Regulator pressure | 15 psi p.2 ↗ | on regulator on front of instrument |
| Flow meter | 0.1 liter/min p.2 ↗ | |
| Column dimensions | 3.0 × 50 mm p.2 ↗ | ZORBAX RRHD Eclipse Plus C18, 1.8 µm (p/n 959757-302) |
| Detector flow cell path length | 10 mm p.2 ↗ | InfinityLab Max-Light Cartridge Cell (G4212-60008) |
Operating Limits
- Inlet pressure of instrument: 30 psi p.2 ↗
- Regulator pressure on front of instrument: 15 psi p.2 ↗
- Flow meter setting: 0.1 liter/min p.2 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 5067-6680 p.2 ↗ | 5067-6680 | Reactor valve pod part number |
| G4212-60008 p.2 ↗ | G4212-60008 | InfinityLab Max-Light Cartridge Cell (10 mm) |
| 959757-302 p.2 ↗ | 959757-302 | ZORBAX RRHD Eclipse Plus C18 column part number |