Protein Sizing and Quantification with the Agilent ProteoAnalyzer System
Technical Reference Summary
Covered Models & Document Scope
Sistema Agilent ProteoAnalyzer con el kit Agilent Protein Broad Range P240; descripción técnica que cubre el dimensionamiento y la cuantificación de proteínas (ejemplos con BSA y CAII), muestras reducidas, rango de dimensionamiento de 10–240 kDa, rango de concentración de 2–2,000 ng/µL.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Samples analyzed simultaneously per run | up to 12 samples p.1 ↗ | |
| Run time | approximately 30 min p.1 ↗ | |
| Sizing range | 10 to 240 kDa p.1 ↗ | |
| Detection concentration range | 2 to 2,000 ng/µL p.1 ↗ | |
| Typical sizing accuracy (LM only) | < 15 % p.3 ↗ | BSA, CAII |
| Typical sizing accuracy (LM and UM) | < 10 % p.3 ↗ | BSA, CAII |
| Average size, BSA | 67.99 to 69.84 kDa p.4 ↗ | LM only, Instruments 1–3 |
| Average size, CAII | 27.17 to 27.87 kDa p.5 ↗ | LM only, 2 to 2,000 ng/µL |
| Lower marker size | 6 kDa p.2 ↗ | |
| Upper marker precision benefit | higher precision for samples smaller than 200 kDa p.2 ↗ | UM added |
| Sample buffer | 1x PBS (30 mM Tris-HCl, 26 mM NaH2PO4, 41 mM Na2HPO4, 79 mM NaCl, pH 8.5) p.2 ↗ | |
| CAII extinction coefficient | 0.575 p.2 ↗ | Nanodrop |
Operating Limits
- Sizing range: 10 to 240 kDa p.1 ↗
- Detection concentration range: 2 to 2,000 ng/µL p.1 ↗
- Maximum samples per run: 12 samples p.1 ↗
- Run time: approximately 30 min p.1 ↗