Table 3. Reported size and percentage of adapter dimer present within an RNA-Seq NGS library analyzed on the Agilent 5200 Fragment Analyzer system and the Agilent 4200 TapeStation system.
Expected Adapter Dimer
Fragment Analyzer HS Small Fragment Kit — Fragment Size (bp)
Fragment Analyzer HS Small Fragment Kit — % (Conc.)
Fragment Analyzer HS NGS Kit — Fragment Size (bp)
Fragment Analyzer HS NGS Kit — % (Conc.)
TapeStation HS D1000 ScreenTape Assay — Fragment Size (bp)
TapeStation HS D1000 ScreenTape Assay — % Integrated Area
TapeStation HS D5000 ScreenTape Assay — Fragment Size (bp)
TapeStation HS D5000 ScreenTape Assay — % Integrated Area
5%
151
4.6%
154
4.9%
161
4.3%
132
5.8%
1%
150
0.9%
154
1.0%
157
0.7%
Not detected
Not detected
0.5%
151
0.4%
154
0.5%
159
0.4%
Not detected
Not detected
0.2%
151
0.2%
154
0.2%
170
0.1%
Not detected
Not detected
0.1%
Not detected
Not detected
154
0.1%
166
0.1%
Not detected
Not detected
Table 2. Reported size and percentage of adapter dimer present within an NGS library analyzed on the Agilent 5200 Fragment Analyzer system and the Agilent 4200 TapeStation system.
Expected Adapter Dimer
Fragment Analyzer HS NGS Kit Fragment Size (bp)
Fragment Analyzer HS NGS Kit % (Conc.)
TapeStation HS D5000 ScreenTape Assay Fragment Size (bp)
TapeStation HS D5000 ScreenTape Assay % Integrated Area
5%
153
4.2%
124
4.7%
1%
154
0.9%
138
1.3%
0.5%
153
0.5%
138
0.8%
0.1%
154
0.1%
Not detected
Not detected
Engineering Facts
11 facts · every claim linked to a page
Selection
Choose a sequencing platform-aware dimer threshold: random flow cells tolerate up to 5% dimer, while patterned flow cells require ≤0.5%, so the QC assay must resolve at least that low.
Determines the required detection sensitivity of the QC instrument before purchase.
Source
“Patterned flow cells, such as the Illumina HiSeq X and NovaSeq 6000, are more sensitive and recommend a threshold of only 0.5% dimer4.”
Selection
For short-read libraries, the HS NGS kit on the Fragment Analyzer covers 100–6,000 bp and the HS D5000 ScreenTape covers 100–5,000 bp, so both fit typical adapter-ligated library sizes.
Confirms the assay sizing range matches the expected library distribution.
Source
“The HS NGS kit offers accurate sizing from 100 to 6,000 bp, while the HS D5000 ScreenTape assay ranges from 100 to 5,000 bp.”
Performance
The Fragment Analyzer HS NGS kit detected the 150 bp dimer down to 0.1% and reported 0.1% accurately, meeting the 0.5% patterned-flow-cell requirement with margin.
Verifies the instrument can resolve below the strictest platform threshold.
Source
“The adapter dimer fragment was detected at all variations tested, from 5% (Figure 2A) to 0.1% (Figure 2D).”
Performance
The TapeStation HS D5000 assay failed to detect the 0.1% dimer sample, so it cannot be used where a 0.5% threshold must be verified.
Prevents selecting an assay that cannot confirm compliance with patterned-flow-cell limits.
Source
“0.1% 154 0.1% Not detected Not detected”
Selection
For RNA-Seq libraries, the HS D1000 ScreenTape resolved the 130 bp contaminant, the 150 bp spike, and the library, while the HS D5000 only saw the 5% spike.
Guides assay choice when small fragments must be separated from the library smear.
Source
“The TapeStation HS D1000 assay fully resolved the 130 bp dimer contamination, the 150 bp fragment, and the library”
Performance
The HS D5000 assay's limit of detection is 5 pg/µL, so 0.1% and 0.5% spikes fell below it and could not be resolved.
Sets the minimum sample concentration needed for reliable low-level dimer detection.
Source
“The concentration of the fragment in the 0.1% and 0.5% samples was below the assay’s limit of detection of 5 pg/uL”
Operation
On the Fragment Analyzer, set minimum peak height to 10 RFU and peak width to 40 seconds to ensure full integration of the smear.
Defines the analysis settings required for accurate percent-area quantification.
Source
“The minimum peak height was adjusted to 10 RFU, and the peak width to 40 seconds.”
Operation
On the TapeStation, manually move the peak start and end points to include the entire smear before calculating percent integrated area.
Prevents under- or over-integration that would misreport dimer percentage.
Source
“The peak width of the library was manually adjusted by moving the start and end points to include the entire smear”
Performance
The Fragment Analyzer sized the 150 bp dimer at 153–154 bp with <3% error, within the kit's 5% specification.
Confirms sizing accuracy is adequate to distinguish dimer from insert.
Source
“a size of 153 or 154 bp in all sample wells and a percent error of less than 3%, within the kit specification of 5%”
Operation
Perform two 0.8x SPRI bead cleanups to eliminate inherent dimers before analysis.
Establishes the baseline cleanup protocol needed before QC measurement.
Source
“0.8x SPRI bead cleanup was performed two times to eliminate any inherent dimers.”
Cumplimiento
The system is for research use only and must not be used in diagnostic procedures.
Restricts deployment to non-diagnostic laboratories.
Source
“For Research Use Only. Not for use in diagnostic procedures.”