Performance Characteristics of the Agilent 1260 Infinity II LC System for Routine Analysis
Technical Reference Summary
Covered Models & Document Scope
Descripción técnica del sistema Agilent 1260 Infinity II LC (sistema cuaternario): bomba cuaternaria G7111B, muestreador de viales G7129A con compartimento de columna integrado G7130A, detector de longitud de onda variable G7114B; ejemplos de aplicación con columnas InfinityLab Poroshell 120 EC-C18
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Maximum operating pressure | 600 bar p.1 ↗ | 1260 Infinity II quaternary pump |
| Flow rate range | up to 5 mL/min p.1 ↗ | 2.1, 3, and 4.6 mm id columns |
| Column compartment temperature | up to 80 °C p.2 ↗ | integrated column compartment, two LC columns |
| Sample cooler temperature | down to 4 °C p.2 ↗ | vialsampler |
| Detector noise | <±2.5 μAU p.2 ↗ | VWD |
| Baseline drift | <1 × 10–4 AU/h p.2 ↗ | VWD |
| Maximum data rate | 120 Hz p.2 ↗ | VWD |
| Injection loop volume | 100 µL p.2 ↗ | vialsampler G7129A |
| Metering device volume | 100 µL p.2 ↗ | vialsampler G7129A |
| Detector flow cell path length | 10 mm p.2 ↗ | VWD standard cell |
| Injection volume | 5 µL p.3 ↗ | isocratic sample, Poroshell 120 EC-C18 |
| Flow rate | 1.0 mL/min p.3 ↗ | isocratic sample, Poroshell 120 EC-C18 |
| Stop time | 15 min p.3 ↗ | isocratic sample, Poroshell 120 EC-C18 |
| Detection wavelength | 254 nm p.3 ↗ | UV, 40 Hz |
| Post time | 1 min p.3 ↗ | RRLC checkout sample |
Operating Limits
- Maximum operating pressure: 600 bar p.1 ↗
- Maximum column compartment temperature: 80 °C p.2 ↗
- Minimum sample cooler temperature: 4 °C p.2 ↗
- Maximum detector data rate: 120 Hz p.2 ↗