The TPS-mini cannot be mounted by its base, so a separate support or flange connection must be designed.
Prevents a mounting design that relies on bolting the base to a frame.
Source
“Le TPS-mini ne peut être fixé par sa base.”
| Parameter | Specification |
|---|---|
| Pumping speed (l/s) (With inlet screen) | |
| With CF 4.5” or ISO 63: | |
| N2 | 48 l/s |
| He | 44 l/s |
| H2 | 32 l/s |
| With KF 40 NW: | |
| N2 | 37 l/s |
| He | 29 l/s |
| H2 | 23 l/s |
| Base pressure* | 5 x 10-8 mbar (3.75 x 10-8 Torr) |
| Inlet flange | CF 4.5" O.D.; ISO63; KF 40 NW |
| Rotational speed | 70000 RPM |
| Start-up time | < 2 minutes |
| Operating position | any |
| Operating ambient temperature | + 5 °C to + 35 °C |
| Bakeout temperature | 120 °C at inlet flange max. (CF flange) |
| 80 °C at inlet flange max. (ISO flange) | |
| Noise level | 56.4 dB (A) at 1 meter |
| Lubricant | permanent lubrication |
| Power supply (Line type): | |
| Input voltage: | 100 – 240 Vac |
| Input frequency: | 50 – 60 Hz |
| Max input power: | 220 VA |
| Stand-by power: | 10 W |
| Protection fuse | 2.5 A |
| Compliance with: | EN 55011 (Class A Group 1) |
| EN 61326-1 | |
| EN 61010-1 | |
| EN 1012-2 | |
| EN 12100 | |
| Indoor Use Only | |
| Max Altitude 2000m | |
| Installation category | II |
| Pollution degree | 2 |
| Power cable | With European, USA or UK plug 2 meters long |
*Base pressure
| N. | Read/Write | Data Type | Description | Admitted Values |
|---|---|---|---|---|
| 205 | R | N | Pump status | Stop = 0 Waiting intlk = 1 Starting = 2 Auto-tuning = 3 Braking = 4 Normal = 5 Fail = 6 |
| 206 | R | N | Error code | Bit description: see the following figure |
| 211 | R | N | Controller Heatsink Temperature (°C) | 0 to 110 |
| 216 | R | N | Controller Air Temperature (°C). | 0 to 70 |
| 224 | R | A | Pressure reading | Format = X.X E XX |
| 226 | R | N | Rotation Frequency (rpm) | |
| 300 | R | N | Cycle time in minutes (zeroed by the reset command) | 0 to 999999 |
| 301 | R | N | Cycle number (zeroed by the reset command) | 0 to 9999 |
| 302 | R | N | Pump life in hours (zeroed by the reset command) | 0 to 999999 |
| 310 to 347 | Reserved to Agilent service | |||
| 400 | R | A | CRC EPROM (QE) | QE8XXXX (where “XXXX” are variable) |
| 401 | R | A | Boot CRC (BL) | BL2XXXX (where “XXXX” are variable) |
| 402 | R | A | CRC Param. (PA) | PA8XXXX (where “XXXX” are variable) |
| 404 | R | A | CRC Parameter structure | "XXXX" |
| 406 | R | A | Program Listing Code | "XXXXXXXXXX" |
| 407 | R | A | Parameter Listing Code | "XXXXXXXXXX" |
| Characteristic | Value |
|---|---|
| Serial communication (kit) | RS232 cable with a 9-pin D type male connector and a 9-pin D type female connector, and software (optional) |
| Storage temperature | - 20 °C to + 70 °C |
| Weight kg (lbs) | KF: 7.9 kg (17.4 lbs) ISO: 7.5 kg (16.5 lbs) CFF: 8.8 kg (19.4 lbs) |
*According to standard DIN 28 428, the base pressure is that measured in a leak-free test dome, 48 hours after the completion of test dome bake-out, with a Turbopump fitted with a ConFlat flange.
| PIN N. | SIGNAL NAME | INPUT/OUTPUT |
|---|---|---|
| 14 | PROGRAMMABLE ANALOG SIGNAL (+) | OUT |
| 15 | GROUND | |
| PROGRAMMABLE ANALOG SIGNAL (-) | OUT |
| N. | Read/Write | Data Type | Description | Admitted Values |
|---|---|---|---|---|
| 000 | R/W | L | Start/Stop (in remote/ mode the window is a read only) | Start = 1 Stop = 0 |
| 001 | R/W | L | Low Speed Activation (in remote/ mode the window is a read only) | No = 0 Yes = 1 (default = 0) |
| 008 | R/W | L | Remote or Serial configuration | Remote = 1 Serial = 0 (default = 1) |
| 100 | R/W | L | Soft Start (write only in Stop condition) | YES = 1 NO = 0 |
| 101 | R/W | N | R1 Set Point type | 0 = Frequency 1 = Power 2 = Time 3 = Normal 4 =Pressure (default = 0) |
| 102 | R/W | N | R1 Set Point valve (expressed in Hz, W or s) | (default = 867) |
| 103 | R/W | N | Set Point delay: time between the pump start and the set point check (seconds) | 0 to 999999 (default = 0) |
| 104 | R/W | L | Set Point signal activation type: the signal can be "high level active" or "low level active" | 0 = high level active 1 = low level active (default = 0) |
| 105 | R/W | N | Set point hysteresis (in % of value) | 0 to 100 (default = 2) |
| Description | Part number |
|---|---|
| Mains cable NEMA plug [3m] | 9699958 |
| Mains cable EU plug [3m] | 9699957 |
| Serial cable and T-Plus Software | 9699883 |
| DIY Full Range Gauge Kit / 74 FS KF 40 | 9699190 |
| DIY Full Range Gauge Kit / 74 FS ISO 63 | 9699192 |
| DIY Full Range Gauge Kit / 74 FS CFF 4.5” | 9699193 |
| FRG-700 3 m cable | 9699960 |
| FRG-700 5 m cable | 9699961 |
| Inlet screen DN40 | 9699309 |
| Inlet screen DN63 | 9699300 |
| FLANGE | FIXING DEVICE | N. | FIXING TORQUE |
|---|---|---|---|
| ISO 63 | M10 clamps | 4 | 22 Nm |
| M8 claws | 4 | 11 Nm |
| CONTROL NAME | DESCRIPTION |
|---|---|
| J1 IN-OUT | Input-output connector to remote control the TPS-mini |
| J2 SERIAL | Serial input-output connector to control the TPS-mini via an RS 232 or RS 485 connection |
| STATUS | Blue LED to show the system operating condition |
| P1 VENT | Connector to control the optional vent valve |
| J5 GAUGE | Connector to control external EyeSys Gauge |
| ON-OFF red switch with integrated light (not shown in the picture) | It is located on the other side of the system (with respect to the illustrated control panel). It is the power supply switch. The integrated red light shows the power supply status: it lights when the power supply is applied to the TPS-mini |
| PIN N. | SIGNAL NAME | INPUT/OUTPUT |
|---|---|---|
| 1 | START/STOP (+) | IN |
| 2 | START/STOP (-) | IN |
| 3 | INTERLOCK (+) | IN |
| 4 | INTERLOCK (-) | IN |
| 5 | SPEED SETTING (+) | IN |
| 6 | SPEED SETTING (-) | IN |
| 7 | SOFT START (+) | IN |
| 8 | SOFT START (-) | IN |
| 9 | +24 Vdc Imax = 30 mA | OUT |
| 10 | NORMAL OUTPUT (relay) Vmax = 125V Imax = 200 mA | OUT |
| 11 | PROGRAMMABLE SET POINT | OUT |
| 12 | NORMAL OUTPUT | OUT |
| 13 | FAULT OUTPUT | OUT |
The TPS-mini cannot be mounted by its base, so a separate support or flange connection must be designed.
Prevents a mounting design that relies on bolting the base to a frame.
“Le TPS-mini ne peut être fixé par sa base.”
The inlet flange must be connected to a fixed counter-flange able to withstand 50 Nm torque around its axis.
Defines the mechanical strength required of the chamber-side flange and support.
“Fix the TPS-mini in a stable position connecting the inlet flange of the turbopump to a fixed counter-flange capable of withstanding a torque of 50 Nm around its axis.”
ISO 63 inlet flanges require 4 M10 clamps at 22 Nm or 4 M8 claws at 11 Nm.
Gives the exact clamp count and tightening torque for leak-free ISO mounting.
“ISO 63 M10 clamps 4 22 Nm M8 claws 4 11 Nm”
A minimum free space of 20 cm around the system is required for proper air circulation.
Sets the clearance needed in the vacuum system enclosure for cooling.
“Assure a free space all around the system at minimum of 20 cm to allow proper air circulation.”
The mains cable must be three-wire (Ph+N+Earth) with at least 0.75 mm2 (AWG18) conductor section.
Specifies the minimum supply cable gauge and earthing requirement.
“The wire section has to be at least 0.75 mm2 (AWG18).”
During bakeout the inlet flange must not exceed 120 °C with ConFlat flanges or 80 °C with ISO flanges.
Limits bakeout temperature depending on flange type to avoid pump damage.
“While heating the vacuum chamber, the temperature of the inlet flange must not exceed 120 °C with ConFlat flanging and 80 °C with ISO flanging.”
The pump must never be run with the inlet flange open to atmosphere or unblanked.
Prevents rotor overspeed and contamination damage during commissioning.
“Never use the turbopump when the inlet flange is not connected to the vacuum chamber or is not blanked.”
The blue STATUS LED flashing at about 400 ms indicates ramp, braking, stop or waiting-for-interlock; about 200 ms indicates an error.
Allows field diagnosis of operating state from the LED without a controller.
“slowly flashing (period of about 400 ms): the system is in ramp, or in braking, or in Stop, or in “Waiting for interlock” status;”
For emergency stop, remove the supply cable from the mains plug or move the external switch to OFF.
Defines the emergency shutdown method to be included in the site safety procedure.
“To immediately stop the TPS-mini in an emergency condition it is necessary to remove the supply cable from the mains plug or moving the external switch to OFF position.”
Before any work on the system, disconnect mains, vent the pump, wait for rotor stop and for surface temperature below 50 °C.
Sets the safe lockout and cool-down conditions for service work.
“wait until the rotor has stopped turning and wait until the surface temperature of the pump falls below 50 °C.”
The TPS-mini requires no maintenance and any work must be done by authorized personnel.
Confirms no scheduled servicing is needed but restricts interventions to qualified staff.
“The TPS-mini does not require any maintenance. Any work performed on the system must be carried out by authorized personnel.”
The pump is designed for high throughput of N2, Ar and lighter gases; heavier gases require contacting Agilent support.
Determines whether the pump suits the process gas before selection.
“The pump is designed to pump high throughput of N2, Ar and lighter gas. Should you need to pump gases heavier than Ar, please contact Agilent technical support for informations.”
The system is not suitable for pumping liquids or solid particles.
Excludes applications with condensable liquids or particulate carryover.
“It is not suitable for pumping liquids or solid particles.”
The shelf life is 12 months from shipping date; if exceeded, the pump must be returned to the factory.
Affects stock rotation and acceptance of pumps stored long before installation.
“the shelf life of a turbomolecular pump is 12 months from the shipping date.”