Clean air. No toxic, corrosive, explosive or particulate forming gases
Ultimate pressure (Torr)
1.0 x 102 Torr (1.3 x 102 mbar)
Maximum inlet pressure
1.0 atmosphere (0 psig)
Maximum continuous inlet pressure
100 Torr
Maximum outlet pressure
1.1 atmosphere (1.5 psig)
Inlet connection
NW25
Exhaust connection
Female 1/4" National Pipe Thread (NW16 adapter provided)
Gas ballast
Female 1/4" National Pipe Thread (40 Micron sintered filter provided)
Ambient operating temperature
5 °C to 40 °C (41 °F to 104 °F)
Storage temperature
–20 °C to 60 °C (–4 °F to 140 °F)
Motor rating
0.67 HP (0.5 kW)
Operating voltage
50-60 Hz:200-240 VAC Single-phase / 100-120 VAC Single Phase* ±10%
Motor line current
200 VAC; 5 A 240 VAC; 4.2 A
Motor thermal protection
Automatic
Operating speed
35-65 Hz Factory setting: 62 Hz, 1800 RPM
Cooling system
Air-cooled
Weight
Pump only: 26 kg (57 lbs) Shipping weight: 33.6 kg (74 lbs)
Noise level (per ISO 11201)
Variable with frequency, 55-68 dB(A)
Vibration level at inlet (per ISO 10816-1)
Variable with frequency
Main fuse
Type H; 10 A, 250 V (Littlefuse 0216010.HxP or equivalent)
Installation and storage
Installation category 2, indoor usage and storage
Pollution degree
2
Altitude
2000 m
*Inverter with P/N of S4897001 ONLY
Table 5 J2 - Serial Port
PIN #
SIGNAL NAME
1
+ 5 V out
2
TX (RS232)
3
RX (RS232)
4
Spare
5
GND
6
A+ (RS485)
7
Spare
8
B- (RS485)
9
Spare
Table 10 Maintenance and Tooling Kits
Description
Contents
Applicable TriScroll Model
Part Number
Major Maintenance Kit
All bearings, bearing seals, bearing lubricant, O-rings, and tip seals required to rebuild TriScroll 300 series pumps.
All TriScroll 300 Series models
PTSS0300MK
Maintenance Tool Kit
All fixtures and tools required to perform any maintenance on TriScroll 300 Series pumps.
All TriScroll 300 Series models
PTSS0300TK
Tip Seal Tool Kit
All tools required to change the tip seals on the TriScroll 300 Series pumps.
All TriScroll 300 Series models
PTSTSTKIT
Replacement Tip Seal Set
Replacement tip seals and static O-rings for TriScroll 300 Series pumps.
All TriScroll 300 Series models
PTSS0300TS
NOTE: After tip seal replacement, the TriScroll 300 Inverter pump may require up to 24 hours run time to achieve full rotational speed.
Table 4 P1 Interlock
PIN #
SIGNAL NAME
IN/OUT
1
VPI relay contact 1
Out
2
0-10 V remote control
In
3
No Connection
4
Start (-) - Connect to Pin 5
In
5
Groundout
Out
6
VPI relay contact 2
Out
7
No Connection
8
Start (+) - Connect to Pin 9
In
9
24 V
Out
Pump operation is interlock dependent. The provided mating connector must be plugged into P1 to start the pump.
Table 9 Other Related Manuals
Title
Applicable TriScroll Model
Part Number
Tip Seal Replacement Manual
All TriScroll 300 Series models
699904280
Pump Module Replacement Manual
All TriScroll 300 Series models
699904285
Major Maintenance Manual
All TriScroll 300 Series models
699904260
Table 11 Factory Service Options
Factory Service Options
Part Number
Advance Exchange TriScroll 300 Pump Module Only
EXPTS0300SC
Service/Rebuild TriScroll 300 Inverter Pump
PTS0300KMA
RETURN THE COMPLETED REQUEST FOR RETURN FORM TO YOUR NEAREST LOCATION:
NORTH AMERICA:
PACIFIC RIM:
EUROPE:
Fax: 1 781 860 9252
please visit our website for individual office information http://www.DJLOHQW.com
Fax: 00 39 011 9979 330
Toll Free: 800 882 7426, Option 3
Fax Free: 00 800 345 345 00
vpl-ra@agilent.com
Toll Free: 00 800 234 234 00
vpt-customercare@agilent.com
Environmental requirements for transport and storage
Parameter
Requirement
Temperature
–20 °C to 60 °C (–4 °F to 140 °F)
Relative humidity
0 to 95% (non-condensing)
Engineering Facts
13 facts · every claim linked to a page
Selection
The pump is only suitable for air or inert gases and must not be used for toxic, corrosive, explosive, or particulate-forming gases.
Media compatibility determines whether this pump can be selected for a given process gas.
Source
“This pump is suitable for pumping air or inert gases. The pump is not intended to pump toxic, corrosive, explosive, or particulate-forming gases.”
Installation
The pump must be connected using a high-voltage IEC-320 power cord rated for at least 10 A.
Using an undersized or incorrect cord creates a fire/electrical hazard and may void compliance.
Source
“The pump must be connected to the power supply using a high voltage IEC-320 type power cord of at least 10 A capacity.”
Installation
If an extension cord is used, it must be at least 16-gauge wire and no longer than 25 ft (7.6 m).
An undersized or too-long extension cord causes voltage drop, loss of power, and overheating.
Source
“Agilent recommends using only extension cords with a minimum of 16-gauge wire and a maximum length of 25’ (7.6 m).”
Installation
A circuit breaker of the proper capacity must be installed near the pump and marked with the on/off symbol.
Proper breaker sizing and location are required for safe disconnection and short-circuit protection.
Source
“Protect against short circuits by installing a circuit breaker of the proper capacity.”
Operation
The pump must be run at atmosphere periodically to flush out accumulated dust and debris because it has no oil to cleanse itself.
Regular atmospheric flushing prevents debris accumulation and maintains pump performance.
Source
“Run the pump periodically at atmosphere for a minute or two to flush out the pump.”
Operation
The pump requires about one hour of running with the inlet sealed to reach ultimate vacuum.
Operators should not expect ultimate pressure immediately after startup.
Source
“In order to reach ultimate vacuum, the pump must be left running for about one hour with the inlet sealed.”
Installation
A fast-acting, normally closed isolation valve is strongly recommended and must open at least 350 ms after startup and close within 250 ms after shutdown.
Proper valve timing prevents pump debris from backstreaming into the vacuum chamber.
Source
“The opening of this valve must occur 350 ms after pump startup.”
Operation
The gas ballast port should not be blocked; dry nitrogen at about 5 lpm can be used if atmospheric air ingress is undesirable.
Blocking the gas ballast can cause condensate accumulation and degrade pump performance.
Source
“Blocking of the gas ballast port is not recommended.”
Installation
The bearing purge port must be supplied with dry nitrogen at 2 psig or less and must be kept below 5 psig.
Exceeding the pressure limit can damage the bearing seals and cause leakage.
Source
“This gas supply should be maintained at 2 psig or less and must be kept below 5 psig.”
Maintenance
After tip seal replacement, the pump may require up to 24 hours of run time to achieve full rotational speed.
Maintenance planning must account for this break-in period before full performance is restored.
Source
“After tip seal replacement, the TriScroll 300 Inverter pump may require up to 24 hours run time to achieve full rotational speed.”
Performance
The pump is power-limited to 750 W; if more power is needed at the set frequency, the drive reduces output frequency (autotuning).
Understanding autotuning explains why the pump may not reach the target frequency under high load.
Source
“If the pump requires more than 750 Watts to run at the set frequency, the pump will be power limited and the output frequency will be reduced.”
Operation
Remote speed control uses a 0–10 V analog input: above 9 V gives full speed, below 1 V gives minimum speed (35 Hz).
Correct analog signal levels are needed for proper remote speed control.
Source
“Vin > 9 V: Driving frequency will be the full speed (factory set @62 Hz)”
Safety
The pump surface can exceed 55 °C (131 °F) during operation; allow cooling before maintenance.
Burn hazard requires cool-down before servicing.
Source
“If you have to perform maintenance on the pump after a considerable time in operation, allow the pump to cool as the temperature of the outer surface may be in excess of 55 °C (131 °F).”