Grid Voltage: +600 VDC max
Emission Current (total): 67 mA DC
Filament Voltage: 7 V
Measurement Specifications
Sensitivity:
• 25/Torr (UHV-24), nominal
• 20/Torr (UHV-24p), nominal Emission Current:
• 4 mA for widest measurement range 5x10-10 Torr
to 1x10-4 Torr
• 10 mA for pressures lower than 5x10-10 Torr ≤ 0.1 mA for
pressure over 1x10-4 Torr
Materials
Filament: Tungsten or yttria-coated iridium filaments
Filament Supports: SST
Collector - tungsten:
• 0.007" diameter for UHV-24
• 0.002" for UHV-24p
Grid: SST
Feedthrough: SST, Alumina, Nickel Alloy
Technical Information
Раздел
Страница
Description of the UHV-24/UHV-24p Ionization Gauge
16
Operation
16
Installation
18
Technical Specifications
19
Application Notes
21
Degas
23
Bakeout
23
Maintenance: Filament Replacement
24
Tungsten Filament Replacement
24
Yttria-coated iridium filament Replacement
25
Gas Correction Factors
26
Accessories and spare parts
29
Table 3 Accessories and spare parts
Part Number
Description
X3004-64403
Filament replacement kit (yttria-iridium)
9710018
UHV-24 Filament Replacement kit, Tungsten
Engineering Facts
14 facts · every claim linked to a page
Selection
The UHV-24 covers 1 mTorr to 2x10-10 Torr while the UHV-24p extends the low end to 5x10-11 Torr, so choose the p-version for measurements below 2x10-10 Torr.
Defines the usable measurement range and where accuracy degrades, which drives gauge model selection.
Source
“provides reliable pressure measurement from 1 mTorr down to 2x10-10 Torr, with reduced performance at pressures lower than 2x10-10 Torr.”
Selection
The UHV-24p uses a thinner collector that lowers the x-ray limit to 5x10-12 Torr, enabling measurement down to 5x10-11 Torr.
The x-ray limit is the hard floor of UHV measurement, so it determines whether the gauge can reach the target base pressure.
Source
“utilizes an extremely thin collector which lowers its x-ray limit to 5x10-12 Torr.”
Installation
Mount the nude gauge in a relatively open space to reduce wall outgassing caused by localized filament heating.
“Install the gauge in a relatively open space to reduce wall outgassing due to localized heating by the filament.”
Installation
When mounted horizontally, place the filaments to the side of the grid so small vertical movements do not change the grid-to-filament spacing.
Orientation affects sensitivity stability and measurement repeatability.
Source
“position the filaments to the side of the grid rather than below or above it so that a slight vertical movement of the grid or filament does not change their relative spacing.”
Safety
The gauge controller and vacuum system must be separately grounded to a common ground, and the ground must support at least 10 A.
Improper grounding can expose personnel to 160-900 V and cause lethal shock.
Source
“Be certain that your UHV-24/UHV-24p Ion Gauge Controller and vacuum system are separately grounded to a common ground.”
Safety
Do not run a ground wire between the vacuum chamber and the controller chassis because large continuous currents could flow through it.
Prevents hazardous ground-loop currents and equipment damage during installation.
Source
“Do not place a ground wire between the vacuum chamber and the controller chassis; large continuous currents could flow through it.”
Safety
The gauge is not intrinsically safe; do not turn it on when combustible gases or mixtures are present because the filament and high-voltage discharge can ignite them.
Critical for process safety when pumping or backfilling flammable gas mixtures.
Source
“As with all Ionization gauges, this device is not intrinsically safe.”
Maintenance
Do not use silicone oil or silicone grease on vacuum equipment; use powder-free butyl or polycarbonate gloves and avoid Alconox on aluminum.
Silicone contamination ruins UHV performance and Alconox damages aluminum parts.
Source
“Do not use silicone oil or silicone grease.”
Operation
A +450 °C bakeout is the safe maximum for repeated or extended bakeout of the nude gauge; do not exceed it or apply thermal shock.
Exceeding the bakeout limit or thermal shock can destroy the gauge during chamber bake.
Source
“A temperature of +450 °C is a safe maximum for repeated or extended bakeout of the nude gauge.”
Operation
Degassing is generally not needed unless the goal is to reach pressures below 1x10-8 Torr, and a ~25 minute e-beam degas is sufficient.
Avoids unnecessary filament wear and chamber heating when lower base pressure is not required.
Source
“Gauge degassing is not generally needed unless the goal is to reach pressures below 1x10-8 Torr.”
Maintenance
Handle the gauge only by its ceramic base and wear nylon gloves, because fingerprint oil extends pumpdown time.
Prevents wire damage and contamination that degrade achievable vacuum.
Source
“Handle the gauge only by its ceramic base to prevent damage to the delicate wires.”
Maintenance
Set the long filament support bracket so the filament sits 0.050 to 0.070 inch from the grid structure.
Correct filament-to-grid spacing is required for proper sensitivity after replacement.
Source
“Insert the long filament support bracket into its collar so that the filament is located 0.050 to 0.070" from the grid structure.”
Performance
Long cables (100 ft or more) cause leakage currents and noise pickup; PTFE is the recommended insulator and ferrite beads may help.
Cable length and routing directly limit the lowest measurable ion current and reading accuracy.
Source
“Leakage currents are extremely difficult to avoid when using cables of 100 ft or longer.”
Performance
A grounded conductive wall near the tube can raise sensitivity by as much as 30%, typical of gauges mounted in tubulations.
Mounting geometry changes calibration, so installation position must be considered for accurate readings.
Source
“A grounded conductive wall near the tube raises the sensitivity by as much as 30%.”