Induztr
Resumen técnico Visor de documentos Modelos relacionados

VHS-4, VHS-6 and VHS-250 Diffusion Pumps

Contenido

  1. Quality p. 3
    Manufacturer warranty terms and conditions
  2. Overview p. 5
    Table of contents listing document sections
  3. Overview p. 7
    List of figures with page references
  4. Overview p. 9
    List of tables with page references
  5. Overview p. 11
    Document purpose, audience and conventions
  6. Warning pp. 12-18
    Safety instructions, symbols and hazard warnings
  7. Procedure pp. 19-32
    Pump description, specifications and installation steps
  8. Procedure pp. 33-34
    Operating instructions and startup/shutdown procedures
  9. Procedure pp. 35-48
    Routine maintenance, cleaning and service procedures
  10. Procedure pp. 49-52
    Fault symptoms, causes and corrective actions
  11. Ordering pp. 53-58
    Spare parts lists, part numbers and ordering information
  12. Services p. 59
    Return authorization form and instructions
  13. pp. 60-64
    Contact details for sales and technical support
  14. p. 19 table, 5 rows
  15. p. 50 table, 12 rows
  16. p. 23 table, 24 rows
  17. p. 25 table, 24 rows
  18. p. 27 table, 24 rows
  19. p. 53 table, 18 rows
  20. p. 56 table, 3 rows
  21. p. 20 table, 16 rows
  22. p. 21 table, 12 rows

Table 1-1 Pump Operational Specifications

SpecificationVHS-4VHS-6VHS-250
Power Rating1450 Watts2200 Watts2200 Watts
Optimum Operating Range (Torr)1x10−3 to < 5x10−9 (1.3 x 10-3 to < 6.6 x 10-9 mbars)1x10−3 to < 5x10−9 (1.3 x 10-3 to < 6.6 x 10-9 mbars)7x10−4 to < 5x10−9 (9 x 10-4 to < 6 x 10-9 mbars)
Maximum Pumping Speed (l/s)Air – 1200 with std cold cap; 950 with extended cold cap Helium –1500 with std cold cap; 1200 with extended cold capAir – 2400 with std cold cap; 1600 with extended cold cap Helium –3000 with std cold cap; 2000 with extended cold capAir – 3700 with std cold cap; Helium –4600
Maximum Throughput (Torr-l/s)1.2 (1.6 mbar-l/s) in operating range 2.5 (3.2 mbar-l/s) @0.01 Torr2.4 (3.2 mbar-l/s) in operating range 3.5 (4.5 mbar-l/s) @0.01 Torr2.6 (3.5 mbar-l/s) in operating range 3.5 (4.5 mbar-l/s) @0.01 Torr
Maximum ForepressureNo Load – 0.65 Torr (.87 mbars) Full Load – 0.55 Torr (.73 mbars)No Load – 0.65 Torr (.87 mbars) Full Load – 0.55 Torr (.73 mbars)No Load – 0.65 Torr (.87 mbars) Full Load – 0.55 Torr (.73 mbars)

Table 4-1 Troubleshooting

SymptomProbable CauseRepair Action
Poor system pressureLeak in system — virtual or real.Locate and repair.
Poor system pressureHigh process gas load.Measure gas load, and eliminate the cause.
Poor system pressureSystem dirty.Clean system to reduce outgassing.
Poor ultimate pressureContaminated pump fluid.Examine, clean pump, and replace fluid.
Poor ultimate pressureLow heat input.Check voltage. Check for continuity, burned out element, poor thermal contact.
Poor ultimate pressureInadequate cooling water flow.Check water pressure. Check tubing for obstructions or backpressure.
Poor ultimate pressureExcessive or too cold cooling water.Check temperature, adjust flow.
Poor ultimate pressureHigh forepressure.Check for leak in foreline, poor mechanical pump performance, breakdown of mechanical pump fluid.
Poor ultimate pressureWater in Quick Cool Coil.Check and remove water.
Slow pumpdown (prolonged cycle after checking gas load conditions)Low heat input.Check heaters.
Slow pumpdown (prolonged cycle after checking gas load conditions)Low oil level.Add oil.
Slow pumpdown (prolonged cycle after checking gas load conditions)Malfunctioning pump assembly Improperly located jets. Damaged jet system.Check and repair or replace.

Table 1-2 VHS-4 Dimensions

ParameterASA inASA mmISO inISO mmConflat inConflat mm
Height18.0545818.0545818.05458
Centerline to Centerline10.6227010.6227010.62270
Flange Face to Flange Face5.031285.031285.03128
EIectrlc8I Box Height8.632198.632198.63219
Sight Glass, Degrees from Foreline90° clockwise90° clockwise90° clockwise
Electrical Box, Degrees from Foreline30° counterclockwise30° counterclockwise30° counterclockwise
Inlet Flange, Nominal Size4” ASAISO -160K8” CFF
Inlet Flange, OD9.002297.091807.96202
Inlet Flange, ID5.911505.781475.91150
Inlet Flange, Thickness0.50130.47120.9324
Inlet Flange, Bolt Circle7.50191N/AN/A7.12181
Inlet Flange, Number of Holes8N/A20
Inlet Range, Hole Size0.6918N/AN/A.338
O-ring Groove ID6.06154N/AN/AN/AN/A
O-ring Groove Width0.184N/AN/AN/AN/A
Foreline Flange, Nominal SizeKF40KF-402¾ CFF
OD2.16552.16552.7369
ID1.38351.38351.3835
Thickness0.2050.2051.3835
Bolt CircleN/AN/AN/AN/A2.3159
Number of HolesN/AN/A6
Hole SizeN/AN/AN/AN/A0.277
O-ring Groove IDN/AN/AN/AN/AN/AN/A
O-ring Groove WidthN/AN/AN/AN/AN/AN/A

Table 1-3 VHS-6 Dimensions

ASA inASA mmISO inISO mmConflat inConflat mm
Height21.9255721.9255721.92557
Centerline to Centerline13.3834013.3834013.38340
Flange Face to Flange Face8.622198.622198.62219
EIectrlc8I Box Height9.122329.122329.12232
Sight Glass, Degrees from Foreline90° clockwise90° clockwise90° clockwise
Electrical Box, Degrees from Foreline30° counterclockwise30° counterclockwise30° counterclockwise
Inlet Flange, Nominal Size6” ASAISO -200K10” CFF
Inlet Flange, OD11.002799.452409.97253
Inlet Flange, ID7.882007.721967.88200
Inlet Flange, Thickness0.75190.47120.9624
Inlet Flange, Bolt Circle9.50241N/AN/A9.12232
Inlet Flange, Number of Holes8N/A24
Inlet Range, Hole Size0.8121N/AN/A.338
O-ring Groove ID8.20208N/AN/AN/AN/A
O-ring Groove Width0.184N/AN/AN/AN/A
Foreline Flange, Nominal Size1½" ASAKF-5033/8" CFF
OD5.001272.05753.3886
ID1.95501.95501.9550
Thickness0.63160.2460.6216
Bolt Circle3.8899N/AN/A2.8572
Number of Holes4N/AN/A8
Hole Size0.6216N/AN/A0.338
O-ring Groove ID2.2256N/AN/AN/AN/A
O-ring Groove Width0.308N/AN/AN/AN/A

Значения приведены дословно, включая опечатки оригинала (EIectrlc8I, Inlet Range).

Table 1-4 VHS-250 Dimensions

ISOinmm
Height22.79579
Centerline to Centerline13.38340
Flange Face to Flange Face9.56243
Electrical Box Height9.12232
Sight Glass, Degrees from Foreline90° clockwise
Electrical Box, Degrees from Foreline30° counterclockwise
Inlet Flange, Nominal SizeISO - 250F
Inlet Flange, OD13.19335
Inlet Flange, ID10.75273
Inlet Flange, Thickness0.6316
Inlet Flange, Bolt Circle12.205310
Inlet Flange, Number of Holes12
Inlet Range, Hole Size0.4411
O-ring Groove ID10.95278
O-ring Groove Width0.144
Foreline Flange, Nominal Size1½ ASA
OD5.00127
ID1.8848
Thickness0.6216
Bolt Circle3.8899
Number of Holes4
Hole Size0.6216
O-ring Groove ID2.2256
O-ring Groove Width0.308

Table 5-1 VHS-4 Models

PumpVoltagePart NumberFlange Type
Standard cold cap120 V86460301ASA
Standard cold cap120 VL6256301ISO
Standard cold cap120 VL6188301Conflat
Extended cold cap120 V86460311ASA
Extended cold cap120 VL6256311ISO
Extended cold cap120 VL6188311Conflat
Standard cold cap208 V86460306ASA
Standard cold cap208 VL6256306ISO
Standard cold cap208 VL6188306Conflat
Extended cold cap208 V86460316ASA
Extended cold cap208 VL6256316ISO
Extended cold cap208 VL6188316Conflat
Standard cold cap240 V86460302ASA
Standard cold cap240 VL6256302ISO
Standard cold cap240 VL6188302Conflat
Extended cold cap240 V86460312ASA
Extended cold cap240 VL6256312ISO
Extended cold cap240 VL6188312Conflat

Table 5-5 VHS-250 Models

PumpVoltagePart NumberFlange Type
Standard cold cap120 VK0543301ISO
Standard cold cap208 VK0543306ISO
Standard cold cap240 VK0543302ISO

Table 1-1 Pump Operational Specifications (Continued)

SpecificationValue
Backstreaming Rate at Pump Inlet5x10−4 mg/cm2/minute (with std cold cap)
Electrical Requirements120, 208, 240; 50/60 Hz; single phase
Warm-up Time10 minutes
Cooldown Time (using Quick Cool Coil)10 minutes
Fluid ChargeVHS-4: 300 cc; VHS-6 and VHS-250: 500 cc
Cooling Water RequirementsMaximum inlet temperature – 60/80 °F (15 - 26 °C); Maximum outlet temperature at foreline – 120 °F (48.9 °C); Flow rate – VHS-4: 0.15 gpm; VHS-6 and VHS-250: 0.25 gpm
Pressure drop across coilsVHS-4: 4 psi; VHS-6 and VHS-250: 15 psi
Backing Pump SizeVHS-4: ≥ 10 cfm for maximum throughput; VHS-6 and VHS-250: 17 cfm for maximum throughput
Recommended Jet Assembly4-stage, self-aligning, stainless steel
Foreline BaffleStacked half moon
Cold CapNickel-plated copper
Water Connections1/8" FPT
Thermal SwitchesManual reset at 300 °F (149 °C)
EnvironmentalMaximum ambient temperature 113 °F (45 °C)
InstallationIndoor use, Installation Category 2, Pollution Degree 2
Altitude6562’ (2000 m)

Table 1-1 Pump Operational Specifications (Continued)

SpecificationVHS-4VHS-6 and VHS-250
Heater circuit resistance (ohms)
Heater circuit resistance (ohms) — 120 V9.46.2
Heater circuit resistance (ohms) — 208 V28.318.7
Heater circuit resistance (ohms) — 240 V37.724.9
Heater power1450 watts2200 watts
Clearance (for heater removal)6" minimum6" minimum
Materials — Body, Flanges, Foreline, Bafflestainless steelstainless steel
Materials — Jet Assemblystainless steelstainless steel
Materials — Body Cooling Coilscoppercopper
Materials — Quick Cooling Coilstainless steelstainless steel
Materials — Cold Capnickel-plated coppernickel-plated copper
Shipping Weight55 pounds75 pounds

Engineering Facts

14 facts · every claim linked to a page
Selection

Pump power rating differs by model, so the electrical supply and heater circuit must be sized per model.

Determines heater power supply, wiring and breaker sizing at selection.

Source
“VHS-4: 1450 Watts”
Installation

A minimum clearance is required below the pump to allow heater removal.

Affects mechanical layout and serviceability of the heater.

Source
“Clearance (for heater removal)”
Installation

The pump body must be mounted vertical and plumb, and the inlet mating flange horizontal within ±1°.

Prevents uneven fluid level and boiler plate overheating.

Source
“Ensure that the pump inlet mating flange on the system is horizontal within ±1°.”
Installation

Cooling water must be supplied continuously at model-specific flow and 60 to 80 °F.

Defines cooling water utility requirements and prevents overheating.

Source
“Connect the inlet water fitting (near the inlet flange at the top of the pump) to a continuously running water supply at 0.15 gpm (VHS-4) or .25 gpm (VHS-6 and 250) and at a temperature of 60 to 80 °F.”
Safety

The thermal switch must be wired in series with the heater to cut power on overtemperature.

Critical interlock wiring to protect personnel and pump from overtemperature.

Source
“Connect the thermal switch in series with the heater.”
Operation

The pump must be rough-pumped below 0.5 Torr before the heater is energized.

Operating above this pressure risks explosion and pump damage.

Source
“The pressure must be reduced to less than 0.5 Torr (.66 mbars).”
Operation

In humid environments heaters must be started at reduced voltage to expel absorbed moisture.

Prevents heater damage on startup in humid conditions.

Source
“start pumps at a low line voltage,e.g., 50% of line voltage, for a period or time to slowly expel any moisture.”
Operation

Cooling water must keep flowing until the body cools to about 130 °F before venting.

Prevents fluid degradation and unsafe venting of a hot pump.

Source
“Allow cooling water to flow through the diffusion pump until the pump body temperature, located just above the boiler plate, has cooled to a temperature of approximately 130 °F.”
Maintenance

New O-rings must be used whenever fill plugs or the sight glass are replaced.

Prevents leaks at fill/drain and sight glass during servicing.

Source
“Always use new O-rings when replacing fill plugs or the sight glass.”
Maintenance

The heater clamp bolt must be tightened to a specific torque during heater replacement.

Ensures proper thermal contact and prevents heater damage.

Source
“Tighten the heater clamp bolt to a torque of 250 inch-pounds.”
Maintenance

The boiler stud should be coated with anti-seize compound before heater installation.

Prevents seizing and eases future heater servicing.

Source
“Coat the boiler stud with an anti-seize compound such as FEL-PRO C5A or common milk of magnesia.”
Maintenance

Fluid charge is model-specific and must be replenished to the FULL/COLD mark.

Correct fluid charge is required for proper pumping and thermal switch protection.

Source
“300 cc for the VHS-4”
Performance

If gas load exceeds maximum throughput, the diffusion pump stops functioning and only the backing pump acts.

Guides throughput sizing and diagnosis of poor performance.

Source
“If the gas load (Q) exceeds the maximum throughput capability of the diffusion pump, the diffusion pump will not function and the pumping action will essentially be performed by the mechanical backing pump.”
Safety

Hydrocarbon-based pumping oils must not be used due to explosion risk.

Fluid selection directly affects explosion safety of the system.

Source
“Do not use hydrocarbon-based pumping oils.”