5977C Series MSD Operating Manual

Sommaire

  1. Overview pp. 11-26
    Overview of the 5977C MSD, hardware description, and safety warnings
  2. Overview pp. 27-34
    Intended use and regulatory information for the instrument
  3. Procedure pp. 35-54
    Column conditioning and installation procedures for 8890 GC
  4. Procedure pp. 55-76
    Column and gasket replacement procedures for Intuvo 9000 GC
  5. Procedure pp. 77-114
    EI operation, DS control, eModule readout, and venting procedures
  6. Procedure pp. 115-138
    CI autotune, reagent gas control, and PCI/NCI operation procedures
  7. Procedure pp. 139-202
    Analyzer maintenance, HES and source removal, cleaning, and assembly
  8. Procedure pp. 203-234
    Switching between EI and CI sources, cleaning, and filament replacement
  9. p. 121 table, 6 rows
  10. p. 36 table, 8 rows
  11. p. 173 table, 8 rows
  12. p. 174 table, 11 rows
  13. p. 189 table, 5 rows
  14. p. 190 table, 14 rows
  15. p. 16 table, 10 rows
  16. p. 128 table, 7 rows
  17. p. 155 table, 3 rows
  18. p. 156 table, 15 rows

Table 12 Default Tune Control Limits, used by CI autotune only

Reagent gasMethaneMethaneIsobutaneIsobutaneAmmoniaAmmonia
Ion polarityPositiveNegativePositiveNegativePositiveNegative
Abundance target1x1061x106N/A1x106N/A1x106
Peakwidth target0.60.6N/A0.6N/A0.6
Maximum repeller44N/A4N/A4
Maximum emission current, µA24050N/A50N/A50
Max electron energy, eV240240N/A240N/A240

N/A: Not available. There are no PFDTD ions formed in PCI with any reagent gas but methane, hence, CI autotune is not available with these configurations.

Table 5 Gas flows

FeatureGas flows
High vacuum pumpDiffusionTurbo
Optimal He column flow mL/min11 to 2
Maximum recommended gas flow mL/min*1.54
Maximum gas flow, mL/min†26.5
Max column id0.53 mm (30 m)0.53 mm (30 m)
CI capabilityNoYes
CI reagent flows, mL/minNA1 to 2
JetClean option H2 flowNA0.4 mL/min

*Total gas flow into the MSD: column flow plus reagent gas flow (if applicable) plus JetClean H2 flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. †Expect degradation of spectral performance and sensitivity.

Table 19 Parts list for the EI SS or Inert source (Figure 56)

Item numberItem descriptionPart number (SS)Part number (Inert)
1Gold plated set screwG1999-20022G1999-20022
2Gold plated screwG3870-20021G3870-20021
3Interface socketG1099-20136G1099-20136
4Source bodyG1099-20130G2589-20043
5Drawout cylinderG1072-20008G1072-20008
6Drawout plate05971-20134G2589-20100
74-turn filamentG7005-60061G7005-60061
8Spring washer3050-13743050-1374

Table 19 Parts list for the EI SS or Inert source (Figure 56) (continued)

Item numberItem descriptionPart number (SS)Part number (Inert)
8Flat washer3050-09823050-0982
9Lens insulatorG3170-20530G3170-20530
10Entrance lensG3170-20126G3170-20126
11Ion focus lens05971-2014305971-20143
12Repeller insulatorG1099-20133G1099-20133
13RepellerG3870-60172G3870-60173
14Flat washer3050-06273050-0627
15Belleville spring washer3050-13013050-1301
16Repeller nut0535-00710535-0071
17Source heater block assemblyG3870-60180G3870-60179
18Repeller block insertG3870-20135G3870-20135

Table 23 Parts list for the EI standard or EI Inert source (Figure 61)

Item numberItem descriptionPart number (SSL)Part number (Inert)
1Gold plated set screwG1999-20022G1999-20022
2Gold plated screwG3870-20021G3870-20021
3Interface socketG1099-20136G1099-20136
4Source bodyG1099-20130G2589-20043
5Drawout cylinderG1072-20008G1072-20008

Table 23 Parts list for the EI standard or EI Inert source (Figure 61) (continued)

Item numberItem descriptionPart number (SSL)Part number (Inert)
6Drawout plate05971-20134G2589-20100
74-turn filamentG7005-60061G7005-60061
8Spring washer3050-13743050-1374
8Flat washer3050-09823050-0982
9Lens insulatorG3170-20530G3170-20530
10Entrance lensG3170-20126G3170-20126
11Ion focus lens05971-2014305971-20143
12Repeller insulatorG1099-20133G1099-20133
13RepellerG3870-60172G3870-60173
14Flat washer3050-06273050-0627
15Belleville spring washer3050-13013050-1301
16Repeller nut0535-00710535-0071
17Source heater block assemblyG3870-60180G3870-60179
18Repeller block insertG3870-20135G3870-20135

Table 3 5977C Series MSD features

FeatureDiffusionTurbo
High vacuum pumpDiffusionTurbo
Optimal He column flow mL/min 11 to 2
Maximum recommended gas flow mL/min*1.54
Maximum gas flow, mL/min†26.5
Max column id0.25 mm (30 m)0.53 mm (30 m)
CI capability‡NoYes
Inert ion sources availableYesYes
GC compatibility9000 and 88909000 and 8890
Foreline pumps availablePfeiffer DuoPfeiffer Duo, Pfeiffer MVP-070-3x, Agilent IDP-3 24V
DIP** capability (3rd party)YesYes

* Total gas flow into the MSD: column flow plus reagent gas flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. † Expect degradation of spectral performance and sensitivity. ‡ Turbo pump models are field upgradeable to CI. ** Direct insertion probe.

Table 14 Reagent gases

Reagent gas/modeReagent ion massesPFDTD Calibrant ionsFlow adj ions: Ratio
EI/PCI/NCI MSD Turbo pump Recommended flow: 20% PCI 40% NCI
Methane/PCI17, 29, 41*41, 267, 59928/27: 1.5 – 5.0
Methane/NCI17, 35, 235†185, 351, 449N/A
Isobutane/PCI39, 43, 57N/A57/43: 5.0 – 30.0
Isobutane/NCI17, 35, 235185, 351, 449N/A
Ammonia/PCI18, 35, 52N/A35/18: 0.1 – 1.0
Ammonia/NCI17, 35, 235185, 351, 517N/A

* There are no PFDTD ions formed with any reagent gas but methane. Tune with methane and use the same parameters for the other gas. † There are no negative reagent gas ions formed. To pretune in negative mode, use background ions: 17 (OH-), 35 (Cl-), and 235 (ReO3-). These ions cannot be used for reagent gas flow adjustment. Set the flow to 40% for NCI and adjust as necessary to get acceptable results for your application.

Table 17 Parts list for EI HES (Figure 48)

Item numberItem descriptionPart number (XTR)
1Source finger gripG7002-20008
2Filament blockG7002-20019
3Extractor lens (5)*, with 3 mm openingG7004-20061

Table 17 Parts list for EI HES (Figure 48) (continued)

Item numberItem descriptionPart number (XTR)
4Ceramic insulator for extractorG7002-20064
5Entrance lens assembly, Extended, HES (1)*G7004-20065
6Ion focus lens (2)*G7004-20068
7Lens insulator/holderG7002-20074
8M2 x 0.4 screw x 12 mm long gold plated screwG7002-20083
9Source bodyG7002-20084
10Post extractor lens 2 (3)*G7004-20090
11Post extractor lens 1 (4)*G7004-20004
12M2 x 6 mm gold plated screwG7002-20109
13Locking ring lens insulatorG7002-20126
14High efficiency dual filamentG7002-60001
15Ring heater/sensor assemblyG7002-60043
16Source mount 1.5 mmG7002-60053
17Repeller assemblyG7002-67057
Not shownHES assemblyG7004-67056

* The number in parenthesis is the number engraved on the lens

Engineering Facts

14 facts · every claim linked to a page
Selection

Diffusion-pump models cannot be upgraded to CI, so CI capability must be chosen at purchase.

Determines whether the CI option can be added later or must be ordered with the instrument.

Source
“Turbo pump models are field upgradeable to CI.”
Selection

The diffusion-pump model accepts only up to 1.5 mL/min total gas flow versus 4 mL/min for turbo models.

Limits column flow and reagent gas choices when selecting the pump configuration.

Source
“Maximum recommended gas flow mL/min*”
Selection

The diffusion-pump model is limited to 0.25 mm ID columns, while turbo models accept 0.53 mm ID columns.

Constrains column selection and method transfer between MSD models.

Source
“Max column id”
Installation

Column installation procedures differ from earlier MSDs, so using another instrument's procedure can damage the column or MSD.

Prevents installers from reusing older 5975/5977 procedures that may not fit the 5977C interface.

Source
“Using the procedure from another instrument may not work, and may damage the column or the MSD.”
Installation

Carrier gas must be at least 99.9995% pure to protect the column and MSD.

Sets the minimum gas purity specification for the installation.

Source
“Always use carrier gas that is at least 99.9995% pure.”
Installation

New ferrules may loosen after thermal cycling, so tightness must be rechecked or self-tightening nuts used.

Requires a leak check after the first heating cycles to avoid air leaks into the MSD.

Source
“Because of thermal expansion, new ferrules may loosen after heating and cooling a few times.”
Safety

Hydrogen flow must be shut off before MSD power is removed, otherwise hydrogen can accumulate and explode.

Defines the mandatory shutdown sequence when hydrogen is used as carrier or JetClean gas.

Source
“the hydrogen gas flow must be off before turning off the MSD power.”
Safety

The GC oven and GC/MSD interface must be cool before carrier gas is shut off to avoid column damage.

Sets the correct order of operations during venting and shutdown.

Source
“Ensure the GC oven and the GC/MSD interface are cool before turning off carrier gas flow to prevent damage to the column.”
Operation

The tune file must match the installed ion source type.

Prevents wrong tuning and poor data when EI, XTR, HES, or CI sources are swapped.

Source
“The tune file must match the type of ion source in the analyzer.”
Operation

In CI mode an air leak is detected by a visible m/z 32 peak, and operating with a leak rapidly contaminates the source.

Gives the acceptance criterion for leak checking before CI operation.

Source
“A peak there indicates an air leak. Repair the leak before proceeding.”
Operation

In CI mode the m/z 19 peak must be less than 50% of the m/z 17 peak before proceeding.

Defines the water background criterion for a valid methane reagent gas setup.

Source
“The peak at m/z 19 (protonated water) is less than 50% of the peak at m/z 17.”
Operation

EI performance must be verified before switching to CI, and CI must be set up in PCI first even for NCI work.

Establishes the required sequence for CI commissioning and troubleshooting.

Source
“Always verify MSD performance in EI before switching to CI operation.”
Maintenance

An antistatic wrist strap must fit snugly to be effective.

Ensures ESD protection is actually effective when handling boards and the analyzer.

Source
“To be effective, an antistatic wrist strap must fit snugly (not tight).”
Maintenance

Excess stress when removing the filament can crack the filament ceramic, and a defective filament must not be operated.

Warns service engineers about the handling limit during HES disassembly.

Source
“Excessive stress can crack the filament ceramic.”