An Integrated Workflow for Automated Calculation of
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que describe un flujo de trabajo automatizado integrado para la determinación del DAR de ADC utilizando el Agilent AssayMAP Bravo, Agilent MassHunter Walkup y el software Agilent MassHunter BioConfirm DAR Calculator con un sistema LC/MS TOF de masa exacta; abarca ADC intacto y reducido, condiciones de muestra glicosilada y deglicosilada.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| LC/MS system | Accurate-Mass TOF LC/MS p.3 ↗ | |
| Ion source | Dual Agilent Jet Stream p.3 ↗ | Intact ADC |
| Ion source | Dual Agilent Jet Stream p.3 ↗ | Reduced ADC |
| Gas temperature | 325 °C p.3 ↗ | Intact ADC |
| Gas flow | 13 L/min p.3 ↗ | Intact ADC |
| Gas flow | 13 L/min p.3 ↗ | Reduced ADC |
| Nebulizer | 60 psig p.3 ↗ | Intact ADC |
| Sheath gas temperature | 350 °C p.3 ↗ | Intact ADC |
| Sheath gas temperature | 350 °C p.3 ↗ | Reduced ADC |
| Sheath gas flow | 12 L/min p.3 ↗ | Intact ADC |
| Sheath gas flow | 12 L/min p.3 ↗ | Reduced ADC |
| VCap | 4,000 V p.3 ↗ | Intact ADC |
| VCap | 4,000 V p.3 ↗ | Reduced ADC |
| Nozzle voltage | 2,000 V p.3 ↗ | Intact ADC |
| Nozzle voltage | 2,000 V p.3 ↗ | Reduced ADC |
| Fragmentor | 400 V p.3 ↗ | Intact ADC |
| Skimmer | 65 V p.3 ↗ | Intact ADC |
| Mass range | 800–8,000 m/z p.3 ↗ | Intact ADC |
| Scan rate | 1.0 spectra/sec p.3 ↗ | Intact ADC |
| Scan rate | 1.0 spectra/sec p.3 ↗ | Reduced ADC |
Operating Limits
- Autosampler thermostat temperature: 5 °C p.3 ↗
- Sample incubation temperature: 50 °C for 10 minutes p.2 ↗
- ADC storage temperature: –80 °C p.2 ↗