Development of a Rapid 2-AB Sample Preparation Workflow for N-Glycans
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que compara los flujos de trabajo Agilent GlykoPrep con 2-AB y Agilent AdvanceBio Gly-X N-Glycan Prep con 2-AB Express para la preparación de muestras de N-glicanos y el análisis por HILIC-UHPLC; no se cubren modelos de instrumentos ni códigos de pedido.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Denature time (GlykoPrep with 2-AB) | 5 min p.2 ↗ | GlykoPrep with 2-AB |
| Deglycosylation time (GlykoPrep with 2-AB) | 15 to 60 min p.2 ↗ | GlykoPrep with 2-AB |
| Dry down time (GlykoPrep with 2-AB) | 30 to 60 min p.2 ↗ | GlykoPrep with 2-AB |
| Labeling time (GlykoPrep with 2-AB) | 60 min p.2 ↗ | GlykoPrep with 2-AB |
| Cleanup time (GlykoPrep with 2-AB) | 15 to 30 min p.2 ↗ | GlykoPrep with 2-AB, solid phase, vacuum driven |
| Total workflow time (GlykoPrep with 2-AB) | ~3.5 hours p.2 ↗ | GlykoPrep with 2-AB |
| Denature time (Gly-X with 2-AB Express) | 3 min p.2 ↗ | AdvanceBio Gly-X with 2-AB Express |
| Deglycosylation time (Gly-X with 2-AB Express) | 5 min p.2 ↗ | AdvanceBio Gly-X with 2-AB Express, in solution |
| Labeling time (Gly-X with 2-AB Express) | 60 min p.2 ↗ | AdvanceBio Gly-X with 2-AB Express, on matrix |
| Cleanup time (Gly-X with 2-AB Express) | 10 to 15 min p.2 ↗ | AdvanceBio Gly-X with 2-AB Express, on matrix |
| Total workflow time (Gly-X with 2-AB Express) | ~2 hours p.2 ↗ | AdvanceBio Gly-X with 2-AB Express |
| Fluorescence detection excitation/emission | 360/428 nm p.3 ↗ | HILIC-UHPLC analysis |
Operating Limits
- Typical time to UHPLC analysis readiness (Gly-X with 2-AB Express): 2 hours p.2 ↗
- Total workflow time (GlykoPrep with 2-AB): ~3.5 hours p.2 ↗