Agilent SureCycler 8800 Setup and User's Guide

Sommaire

  1. Overview pp. 1-5
    Title page, legal notices, guide overview, and table of contents
  2. Overview pp. 6-7
    Introduction to the home screen and PCR program editing
  3. Overview pp. 8-9
    Prerequisites and preparatory information before installation
  4. Spec pp. 10-11
    Provided materials, fluids, and reagents required for operation
  5. Spec p. 12
    Electrical and environmental ratings of the instrument
  6. Spec p. 13
    Description of the status indicator LED behavior
  7. p. 14
    Technical support contact information and procedures
  8. Procedure pp. 15-17
    Step-by-step instrument installation and power connection
  9. Procedure p. 18
    Procedure for exchanging the thermal block
  10. Procedure pp. 19-20
    Guidance on preparing and loading samples
  11. Spec p. 21
    Recommended Agilent plasticware and consumables
  12. Procedure pp. 22-23
    Starting up the software and initial setup steps
  13. Procedure pp. 24-28
    Managing and editing user accounts and permissions
  14. Procedure pp. 29-31
    Configuring network settings and remote access options
  15. Procedure pp. 32-34
    Configuring cycler name and general instrument settings
  16. Procedure pp. 35-42
    Creating, editing, and tips for PCR programs
  17. Procedure pp. 43-50
    Detailed description of program steps and commands
  18. Procedure pp. 51-52
    Accessing and using preloaded PCR programs
  19. Procedure pp. 53-54
    Editing and copying PCR program steps
  20. Procedure pp. 55-56
    Running PCR programs and incubation protocols
  21. Procedure p. 57
    Accessing and viewing run reports
  22. Procedure pp. 58-60
    Troubleshooting common issues such as condensation and blank screen
  23. Procedure pp. 61-64
    Instrument maintenance and software update procedures
  24. p. 10 table, 7 rows
  25. p. 21 table, 6 rows
  26. p. 51 table, 8 rows
  27. p. 13 table, 4 rows
  28. p. 14 table, 21 rows
  29. p. 22 table, 6 rows
  30. p. 23 table, 2 rows
  31. p. 24 table, 3 rows

Table 1 Materials Provided

PartQuantity
SureCycler 8800 instrument base1
Thermal block unit (96-well or 384-well)*1
Power cord1
USB external drive**1
Functional test certificate1
Quick Start installation poster1
CD with Setup and User’s Guide PDF1

* The thermal block units are sold separately as part numbers G8810A (96-well block) and G8820A (384-well block). **The USB external drive is provided for transferring files to and from the cycler as needed. We recommend using this drive for all file transfers on the cycler.

Table

Agilent Part NumberDescription
40133396-well plate, 25-pack
410188384-well plates, 50-pack
410186Adhesive sealing films, 100-pack
410187Compression mats, 10-pack
410082Strip tubes, 200 l, 80 strips of 12
410086Strip tube caps, 80 strips of 12

Table 5 Description of the preloaded PCR programs for the 96-well SureCycler 8800

PCR Program File NamePCR Enzyme or KitDenaturationAnnealingElongation
Herculase II Fusion 1kb target.pcrHerculase II Fusion DNA Polymerase95°C for 20 seconds55°C for 20 seconds72°C for 30 seconds
Paq5000 HS MM 1kb target.pcrPaq5000 Hotstart Master Mix95°C for 20 seconds55°C for 20 seconds72°C for 30 seconds
PfuTurbo HS MM 1kb target.pcrPfuTurbo Hotstart Master Mix95°C for 30 seconds55°C for 30 seconds72°C for 1 minute
PfuUltra II HS MM 1kb target.pcrPfuUltra II Hotstart Master Mix95°C for 20 seconds55°C for 20 seconds72°C for 15 seconds
QuikChange II 4 kb target.pcrQuikChange II Site-Directed Mutagenesis Kit95°C for 30 seconds55°C for 1 minute68°C for 4 minutes
QuikChange Lightning 4kb target.pcrQuikChange Lightning Site-Directed Mutagenesis Kit95°C for 20 seconds60°C for 10 seconds68°C for 2 minutes
QuikChange Lighting Multi 4 kb target.pcrQuikChange Lightning Multi Site-Directed Mutagenesis Kit95°C for 20 seconds55°C for 30 seconds65°C for 2 minutes
QuikChange Multi 4kb target.pcrQuikChange Multi Site-Directed Mutagenesis Kit95°C for 1 minute55°C for 1 minute65°C for 8 minutes

Table 2 Appearance of the status indicator LED

AppearanceInstrument Status
OffThe cycler is idle.
Blinking greenThe cycler is running a program or incubation.
Solid greenThe cycler is paused.
Blinking redThe cycler has detected an error. Check the display for an error message containing further details.

SureCycler 8800 Technical Support

ContactRegionDetails
EmailUSA and Canadatechservices@agilent.com
EmailEuropetech_europe@agilent.com
EmailJapanemail_japan@agilent.com
EmailAll other countriestechservices@agilent.com
World Wide Webwww.genomics.agilent.com
TelephoneUSA and Canada(800) 227-9770
TelephoneEurope
TelephoneJapan0120-477-111
TelephoneAll other countriesGo to www.agilent.com/genomics/contactus.
TelephoneAustria01 25125 6800
TelephoneBenelux02 404 92 22
TelephoneDenmark45 70 13 00 30
TelephoneFinland010 802 220
TelephoneFrance0810 446 446
TelephoneGermany0800 603 1000
TelephoneItaly800 012575
TelephoneNetherlands020 547 2600
TelephoneSpain901 11 68 90
TelephoneSweden08 506 4 8960
TelephoneSwitzerland0848 8035 60
TelephoneUK/Ireland0845 712 5292

Table 3 Home Screen Button Descriptions

ButtonDescription
New ProgramUse this button to create a new PCR program.
ProgramsOpens the file directory to browse stored PCR programs. Use this button to edit or run programs.
PCR WizardUse this button to create a new PCR program using a guided Wizard.
Network ControlUse this button to connect to multiple instruments via a network.
IncubateStarts a single temperature incubation for a predefined period or for an indefinite hold.
ReportsOpens the report directory allowing you to browse for stored report files.

Table 3 Home Screen Button Descriptions (continued)

ButtonDescription
LoginOpens the Login screen.
SettingsUse this button to adjust cycler settings, upgrade software and manage user accounts.

Table 4 Access levels

Access LevelFunctions allowed
Guest• Access rights to the Guest folder • Creating, copying, editing and executing programs stored in the Guest folder
User• Access rights to the Guest and personal folders • Creating, copying, editing and executing programs stored in the Guest and personal folders
Administrator• Access rights to all Guest and user folders • Setup and maintenance of user accounts • Setup and maintenance of cyclers • Archiving/maintenance of all reports

Engineering Facts

14 facts · every claim linked to a page
Safety

The instrument must be unplugged before cleaning major spills, servicing electrical/internal components, or swapping the thermal block.

Defines mandatory lockout step before maintenance or block exchange to avoid electric shock.

Source
“Unplug the instrument before you clean any major liquid spills and before you service any of the electrical or internal components or swap out the thermal block.”
Safety

The instrument must never be run from an ungrounded outlet.

Requires a grounded AC outlet at the installation site.

Source
“Do not operate the instrument from a power outlet that has no ground connection.”
Safety

The thermal block, heated lid interior and vessels exceed 50°C and must not be touched; the lid stays closed until 30°C or lower.

Burn hazard governs user handling and lab safety briefing.

Source
“Do not touch the thermal block, inner side of heated lid and reaction vessels. These areas quickly attain temperatures of greater than 50°C.”
Safety

Only labware rated to at least 120°C may be used in the cycler.

Sets the temperature rating required when selecting consumables.

Source
“Do not use any materials (plates, sealings, foils, mats) which are not sufficiently temperature-stable (up to 120°C).”
Installation

At least 10 cm clearance must be kept around the instrument so ventilation slots stay free.

Determines bench layout and spacing to neighboring equipment.

Source
“The ventilation slots of the device must remain free to vent at all times. Leave at least 10 cm of space around the instrument.”
Installation

Ambient must be held at 20–30°C and 20–80% non-condensing humidity.

Defines the room HVAC envelope required for correct operation.

Source
“Keep the ambient temperature between 20°C and 30°C with humidity levels between 20% and 80% non-condensing.”
Selection

The electrical supply rating is 100–240 V, 50/60 Hz, 11 A, indoor use only.

Used to size the branch circuit and confirm mains compatibility.

Source
“Electrical supply 100-240 V, 50/60 Hz, 11 A”
Installation

The unit is ESD sensitive; discharges above 8000 V can disrupt the USB ports, so a grounded wrist strap is needed in high-static areas.

Requires ESD precautions during handling and installation.

Source
“Electrostatic discharges greater than 8000 volts may interfere with the normal operation of the USB ports on the instrument.”
Installation

The thermal block only locks into place when placed on the base in the correct orientation, pressed down on both sides.

Prevents improper seating that causes missing temperature display or failed program start.

Source
“The thermal block unit will only lock into place if it is placed on the instrument base in the proper orientation.”
Installation

The instrument must be connected to a grounded AC outlet via the rear power connector.

Confirms the required power connection point and grounding.

Source
“The instrument must be connected to a grounded AC outlet.”
Operation

Two interchangeable blocks exist: 96-well and 384-well.

Drives the block/format choice at purchase and the exchange procedure.

Source
“Two interchangeable thermal block units are available with the SureCycler 8800: the 96-well block and the 384-well block.”
Maintenance

Power must be off and the supply disconnected before removing the thermal block.

Mandatory step sequence for safe block exchange.

Source
“Power off the instrument and disconnect the power supply.”
Operation

Only the listed Agilent plates, strip tubes, sealing films and compression mats are compatible.

Restricts consumable selection to validated part numbers.

Source
“The following PCR plates, reaction tubes and sealing films are compatible with the SureCycler 8800:”
Operation

A compression mat must be used during the run to keep the sealing film adhered to the plate.

Prevents sample volume loss from sealed microplates.

Source
“Ensure a compression mat is used during the PCR run to keep the sealing film adhered to the plate.”