A 30-Color Immunophenotyping Panel

Sommaire

  1. Overview p. 2
    Introduces the 30-color immunophenotyping panel purpose and context
  2. Procedure p. 3
    Step-by-step cell pellet centrifugation and staining protocol
  3. pp. 4-5
    Presents results and discussion of panel performance
  4. p. 6
    Gating strategy for defining cell subpopulations
  5. p. 7
    Frequencies of main subpopulations across samples
  6. p. 8
    Summarizes findings and panel utility
  7. p. 3 table, 30 rows

Table 1. Antibodies used in 30-color immunophenotyping panel on the Agilent NovoCyte Penteon Flow Cytometer, ordered by channels. The laser and filter used for detection for each antibody can be found in the table, as well as concentration and product details for each antibody.

LaserFilterSpecificityFluorochromeCloneVendorPart NumberDilutionAmount Added to 3 × 106 Cells (50 μL)
349445/45Live/Dead Fixable BlueBUV395InvitrogenL231051/10000.05
525/45MHC-IIBUV496M5/114.15.2Becton Dickinson7502811/8000.0625
586/20CD3BUV56317A2Becton Dickinson7413191/4000.125
615/20gdTBUV615GL3Becton Dickinson7511831/2000.25
667/30CD88BUV66120/70Becton Dickinson7500801/4000.125
725/40CD43BUV737S7Becton Dickinson6128401/2000.25
780/60NKp46BUV80529A1.4Becton Dickinson7420661/1000.5
405445/45XCR1BV421ZETBioLegend1482161/4000.125
525/45CD103BV480M290Becton Dickinson5661181/2000.25
586/20TCRbBV570H57-597BioLegend1092311/2000.25
615/20Siglec Hsb600eBio440cThermo Fisher Scientific63-0333-821/2000.25
667/30CD11bBV650M1/70BioLegend1012591/4000.125
725/40CD8bsb70253-6.7Thermo Fisher Scientific67-0081-821/4000.125
780/60CD69BV786H1.2F3Becton Dickinson5646831/2000.25
488525/45SiglecFBB515E50-2440Becton Dickinson5645141/4000.125
586/20CD45Spark 57430F11BioLegend1031841/2000.25
615/20CD4BB630-PGK1.5Becton DickinsonPrototype1/4000.125
667/30CD90.2PerCP30H12BioLegend1053221/2000.25
695/40Ly6GBB7001A8Becton Dickinson5664351/8000.0625
725/40CD62LBB755MEL-14Becton DickinsonPrototype1/4000.125
561586/20B220SYG570RA36B2BioLegend1032861/4000.125
615/20CD172aAF594P84BioLegend1440201/4000.125
667/30CD25PE Cy5PC61BioLegend1020101/2000.25
695/40CD11cPE Cy5.5N418Thermo Fisher Scientific35-0114-821/4000.125
725/40CD19PE-AF7001D3AAT-bioquestATB101941O01/4000.125
780/60NK1.1PE-Cy7PK136BioLegend1087141/1000.5
637667/30CD64AF647X54-5/7.1BioLegend1393221/2000.25
695/40IgDSPARK 68511-26c.2aBioLegend4057501/10000.05
725/40CD44APC-R700IM7Becton Dickinson5654801/4000.125
780/60Ly6CAPC-Cy7HK1.4BioLegend1280261/8000.0625

Engineering Facts

12 facts · every claim linked to a page
Selection

The instrument must be specified with all five laser wavelengths to run this 30-color panel.

Laser and detector count defines whether the full 30-color panel can be acquired in a single tube.

Source
“the Novocyte Penteon is equipped with five lasers (UV, 349 nm; violet, 405 nm; blue, 488 nm; yellow-green, 561 nm; and red, 637 nm) and 30 fluorescence detectors”
Selection

Panel design must account for spectral overlap and the specific instrument configuration, not just marker expression.

Fluorophore choice is constrained by the detector/filter layout, so the panel must be validated against the actual instrument configuration.

Source
“designed through diligent fluorophore selection considering expression markers, fluorescence intensity, spectral overlap, and Agilent NovoCyte Penteon configuration”
Operation

Tissue dissociation uses a two-pass enzymatic protocol with a defined 30-minute incubation at 37 °C.

Reproducible sample prep depends on holding the specified temperature and timing between dissociation passes.

Source
“Wait 30 minutes at 37 °C. Run the "m_lung-01_02" program again on OctoGentleMacs (Miltenyi).”
Operation

EDTA is added after dissociation and held for 2 minutes before filtering.

The EDTA quench step must be timed and dosed exactly to stop enzymatic digestion consistently.

Source
“Add 30 µL of EDTA (0.5 M) and wait 2 minutes.”
Operation

All centrifugation steps are specified at 400 g and 4 °C for 7 minutes.

The centrifuge must reliably hold 400 g at 4 °C to preserve cell viability and pellet recovery.

Source
“Centrifuge for 7 minutes at 4 °C, 400 g.”
Operation

Cells are stained for 30 minutes at 37 °C in the dark, which requires a temperature-controlled dark incubation capability.

Staining temperature and light protection directly affect panel performance and must be controllable.

Source
“Incubate for 30 minutes at 37 °C in the dark.”
Operation

Viability dye is used at a 1/1,000 dilution in PBS with a 15-minute room-temperature dark incubation.

Dilution accuracy and ambient temperature control are needed for consistent live/dead discrimination.

Source
“Resuspend the cells in 100 µL of Live Dead Fixable blue (diluate at 1/1,000 in PBS) for 15 minutes at room temperature (RT) in the dark.”
Performance

The panel is designed to use the maximum detector count available on a single cytometer.

Confirms the configuration ceiling and that all 30 detectors are utilized, so no spare channels are assumed.

Source
“Design of a 30-color immunophenotyping panel maximizes throughput on a 30-detector instrument—to date the maximum number of detectors on a single flow cytometer.”
Operation

The antibody cocktail volume is fixed at 50 µL per 3 million cells.

Staining volume and cell number must be matched to keep the panel reproducible.

Source
“To generate the panel, 50 µL of the antibody cocktail was added to 3 million cells.”
Performance

Absolute counts in events/µL are computed by the software during acquisition.

Volumetric absolute counting must be supported by the acquisition software if concentration data is required.

Source
“The absolute count (on event/µL) of each sample was also determined by NovoExpress Software during acquisition.”
Operation

A single gating strategy is applied across all three organs to allow direct comparison.

Consistent gating across samples is required for valid cross-organ comparison in analysis.

Source
“The same gating strategy was used across all three organs to easily compare the data.”
Selection

The system enables multi-parameter acquisition from a single tube, conserving sample material.

Single-tube acquisition reduces sample consumption, which matters when biological material is limited.

Source
“With five lasers and 30 detectors, it is possible to extract more information from a single tube, saving precious biological material.”