Analysis of Aldehydes in Beer by Agilent PAL 3 Autosampler and 5977C GC/MSD
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que cubre el análisis de cuatro aldehídos (hexanal, furfural, fenilacetaldehído, trans-2-nonenal) en cerveza utilizando un autosampler Agilent PAL3 RTC con herramienta SPME y 8890/5977C GC/MSD; no se indican variantes de modelo/tamaño de producto ni códigos de pedido.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Flavor threshold, hexanal | 350 μg/L p.2 ↗ | |
| Flavor threshold, furfural | 15.157 mg/L p.2 ↗ | |
| Flavor threshold, phenylacetaldehyde | 105 μg/L p.2 ↗ | |
| Flavor threshold, trans-2-nonenal | 0.1 μg/L p.2 ↗ | |
| Calibration range, hexanal | 0.05 to 10 µg/L p.6 ↗ | |
| Calibration range, furfural | 5 to 1,000 µg/L p.6 ↗ | |
| Calibration range, phenylacetaldehyde | 0.1 to 50 µg/L p.6 ↗ | |
| Calibration range, trans-2-nonenal | 0.025 to 5 µg/L p.6 ↗ | |
| Coefficient of determination R², hexanal | 0.999 p.6 ↗ | |
| Coefficient of determination R², furfural | 0.998 p.6 ↗ | |
| Coefficient of determination R², phenylacetaldehyde | 0.996 p.6 ↗ | |
| Coefficient of determination R², trans-2-nonenal | 0.998 p.6 ↗ | |
| Repeatability %RSD (n = 3), hexanal in beer | 1.3 to 4.9 % p.7 ↗ | Brands 1–4 |
| Repeatability %RSD (n = 3), furfural in beer | 1.3 to 2.8 % p.7 ↗ | Brands 1–4 |
| Repeatability %RSD (n = 3), phenylacetaldehyde in beer | 0.9 to 3.7 % p.7 ↗ | Brands 1–4 |
| Repeatability %RSD (n = 3), trans-2-nonenal in beer | 1.7 to 2.8 % p.7 ↗ | Brands 1–4 |
| Method detection limit, hexanal | 0.0009 µg/L p.10 ↗ | 0.05 µg/L level |
| Method detection limit, furfural | 0.52 µg/L p.10 ↗ | 5 µg/L level |
| Method detection limit, phenylacetaldehyde | 0.015 µg/L p.10 ↗ | 0.1 µg/L level |
| Method detection limit, trans-2-nonenal | 0.003 µg/L p.10 ↗ | 0.025 µg/L level |
Operating Limits
- Calibration range, hexanal: 0.05 to 10 µg/L p.6 ↗
- Calibration range, furfural: 5 to 1,000 µg/L p.6 ↗
- Calibration range, phenylacetaldehyde: 0.1 to 50 µg/L p.6 ↗
- Calibration range, trans-2-nonenal: 0.025 to 5 µg/L p.6 ↗