Impurity Profiling of GLP-1 Receptor Agonists by HILIC-MS
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que describe un método genérico de HILIC-MS para el perfilado de principios activos agonistas del receptor de GLP-1 y de impurezas relacionadas con el producto (Semaglutida, Exenatida, Tirzepatida) utilizando un sistema Agilent 1290 Infinity III Bio LC System con Agilent InfinityLab LC/MSD iQ (G6160A).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Flow rate | 0.4 mL/min p.3 ↗ | |
| Detector (DAD) | Agilent 1290 Infinity III Diode Array Detector p.3 ↗ | |
| Flow cell | Agilent Max-Light Cartridge Cell LSS, 10 mm p.3 ↗ | |
| Detector (MS) | Agilent InfinityLab LC/MSD iQ (G6160A) p.3 ↗ | |
| Observed mass, Semaglutide | 4,113.7 Da p.4 ↗ | |
| Observed mass, Tirzepatide | 4,814.0 Da p.4 ↗ | |
| Observed mass, Exenatide | 4,186.9 Da p.4 ↗ | |
| Observed mass, Exenatide impurity 1 | 4,049.5 Da p.8 ↗ | |
| Observed mass, Exenatide impurity 2 | 4,244.1 Da p.8 ↗ | |
| Observed mass, Exenatide main peak | 4,187.6 Da p.8 ↗ | |
| Impurity level, Semaglutide impurity 1 | 0.4 % p.6 ↗ | UV 280 nm |
| Low-abundant impurity range, Semaglutide | 0.2 to 0.4 % p.6 ↗ |