Analysis of Perfluoroalkyl Substances in Hair by Agilent Bond Elut ENV Solid Phase Extraction and LC/Q-TOF
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que cubre un método de cribado de biomonitorización para 20 compuestos PFAS en cabello humano utilizando cartuchos de SPE Agilent Bond Elut ENV (200 mg, 6 mL), un LC Agilent 1290 Infinity II con columna de retardo Agilent C18 4.6 × 30 mm, y un Agilent 6546 Q-TOF operado en modo de iones negativos; aplicado a 11 muestras de cabello de voluntarios.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Sample amount | 100 mg p.3 ↗ | hair sample |
| Reconstitution volume | 500 µL p.3 ↗ | water/methanol |
| Cartridge sorbent mass | 200 mg p.2 ↗ | Bond Elut ENV SPE |
| Cartridge volume | 6 mL p.2 ↗ | Bond Elut ENV SPE |
| Delay column dimensions | 4.6 × 30 mm p.3 ↗ | Agilent C18 delay column |
| Ion mode | negative p.3 ↗ | 6546 Q-TOF |
| Reference mass 1 | 112.9855 m/z p.3 ↗ | Q-TOF |
| Reference mass 2 | 980.0163 m/z p.3 ↗ | Q-TOF |
| QC Low concentration | 0.7 ng/g p.4 ↗ | method validation |
| QC High concentration | 1.5 ng/g p.4 ↗ | method validation |
| Number of PFAS compounds | 20 p.2 ↗ | method scope |
| Number of hair samples | 11 p.2 ↗ | proof-of-concept |
| Retention time PFBA | 2.64 min p.3 ↗ | PFBA |
| Retention time PFOS | 6.68 min p.3 ↗ | PFOS |
| Retention time 11Cl-PF3OUdS | 8.21 min p.3 ↗ | 11Cl-PF3OUdS |
| Exact mass [M-H]– PFBA | 212.9792 m/z p.3 ↗ | PFBA |
| Exact mass [M-H]– PFOS | 498.9302 m/z p.3 ↗ | PFOS |
| Exact mass [M-H]– 11Cl-PF3OUdS | 630.8892 m/z p.3 ↗ | 11Cl-PF3OUdS |
| Recovery PFOS QC Low intraday | 112 % p.4 ↗ | QC Low 0.7 ng/g |
| Recovery PFOS QC High intraday | 112 % p.4 ↗ | QC High 1.5 ng/g |
| Recovery PFHxS QC Low intraday | 81 % p.4 ↗ | QC Low 0.7 ng/g |
| Recovery PFBA QC Low intraday | 104 % p.4 ↗ | QC Low 0.7 ng/g |
Operating Limits
- Required detection sensitivity in hair: order of ng/g p.2 ↗
- Hair drying temperature: room temperature p.3 ↗