Introduces cannabinoid quantification purpose and document structure
Procedurep. 3
Describes preparation steps for plant material samples
p. 4
Lists cited literature and reference sources
p. 2table, 11 rows
Table 1. Target cannabinoids commonly found in Cannabis sativa L. plants and products.
Target Compounds
Cannabidivarin (CBDV)
Tetrahydrocannabivarin (THCV)
(−)-trans-Δ9-Tetrahydrocannabinol (THC)
Cannabidiol (CBD)
Cannabigerol (CBG)
Δ9-Tetrahydrocannabinolic acid (THCA-A)
Cannabidiolic acid (CBDA)
Cannabinol (CBN)
Cannabigerolic acid (CBGA)
Cannabichromene (CBC)
(–)-Δ8-THC
Engineering Facts
12 facts · every claim linked to a page
Conformité
Hemp must contain no more than 0.3% THC by dry weight to be legally compliant, which sets the required detection range and calibration for the assay.
Defines the regulatory threshold the analytical method must reliably quantify for compliance testing.
Source
“with a dry weight concentration of tetrahydrocannabinols not greater than 0.3 % (wt/wt)”
Conformité
Total THC must be calculated as 0.877 × [THCA] + [THC], so the method must quantify both THC and THCA to report compliant total THC.
Determines that both analytes must be in the calibration scope and that the reporting calculation is fixed.
Source
“Quantify tetrahydrocannabinolic acid (THCA) to provide a total THC content (0.877 × [THCA] + [THC]) where required”
Performance
The method separates 11 targeted cannabinoids in under 10 minutes, so a gradient LC with the specified column and modifiers is required to meet throughput.
Sets the performance expectation for run time and confirms gradient capability is needed.
Source
“to separate the targeted cannabinoids in under 10 minutes”
Selection
The workflow is built on the Agilent 1220 Infinity II LC Gradient System VL (G4290C), so the instrument configuration must be gradient-capable.
Identifies the exact system model and part number required for method transfer.
Source
“Agilent 1220 Infinity II LC Gradient System VL”
Operation
Plant samples are extracted at 200 mg in 20 mL methanol (100-fold), then diluted 20-fold for a total 2,000-fold dilution before injection.
Defines the sample prep dilution scheme that keeps concentrations within the calibration range.
Source
“20 fold dilution for a total of 2,000 fold”
Operation
Extracts must be centrifuged at 5,000 rpm for five minutes before the dilution step to remove particulates.
Specifies the centrifugation condition needed to protect the column and ensure clean injections.
Source
“Centrifuge at 5,000 rpm for five minutes.”
Operation
Samples must be filtered through 4 mm, 0.45 μm regenerated cellulose syringe filters (p/n 5190-5107) before analysis.
Specifies the exact filter type and part number to prevent column fouling and ensure reproducibility.
Source
“Filter with 4 mm, 0.45 μm regenerated cellulose (RC) syringe filters (p/n 5190-5107).”
Operation
Oil and concentrate samples use ethanol for the initial 100-fold dilution and methanol for the subsequent 10-fold dilution, giving a final 1,000-fold factor.
Defines the two-solvent dilution protocol and final dilution factor for concentrated matrices.
Source
“final dilution factor = 1,000”
Operation
The second dilution of oil samples must be done in an amber glass autosampler vial, indicating light-sensitive handling is required.
Specifies vial type and volume to protect analytes from light degradation during autosampler residence.
Source
“transferring a 100 µL aliquot into an amber glass 2 mL autosampler vial”
Operation
Higher dilution factors may be necessary depending on the product matrix, so the dilution scheme must be adjustable for potency.
Signals that the fixed dilution may need scaling for high-potency concentrates to stay in range.
Source
“higher dilution factors may be required depending on the product”
Conformité
The products are intended only for cannabis quality control and safety testing where permitted under state or country law.
Establishes the legal-use restriction that must be confirmed before deployment.
Source
“intended to be used for cannabis quality control and safety testing in laboratories where such use is permitted under state/country law”
Selection
A ready-to-run eMethod (G5277#020) for the hemp potency test on the 1220VL LC/VWD system is available, simplifying method implementation.
Provides a validated method package reference to accelerate installation and qualification.
Source
“eMethod G5277#020, Hemp Potency test with the 1220VL LC/VWD system”