Induztr
Résumé technique Visionneuse de documents Modèles associés

Improved GPC with the Agilent 1290 Infinity II LC System

Sommaire

  1. Overview p. 2
    Introduces GPC application needs and the Agilent 1290 Infinity II LC solution
  2. Procedure pp. 3-4
    Describes experimental setup, method, and chromatographic results for GPC
  3. Procedure p. 5
    Demonstrates identical molecular weight results and comparison data
  4. p. 6
    Summarizes benefits of the improved GPC solution
  5. p. 2 table, 6 rows
  6. p. 5 table, 3 rows

Method

ParameterValue
SolventTHF, isocratic, channel B
Flow Rates0.06, 0.3, 0.6, 1.0 mL/min
Stop Time22, 11, or 7 minutes
Column Temperature35 °C, 2 columns in series connected with 75 µm capillary (105 mm length)
Injection Volume20 µL, 4 µL
RIDOptical unit temperature: 35 °C Data rate: 18 Hz Signal polarity: positive

Table 1. Comparison of solvent consumption per sample and sample throughput of the method on a smaller id column (two 250 × 4.6 mm at 0.3, 0.6, and 1.0 mL/min) compared to the standard method*.

Flow Rate (mL/min)Run Time (min)Solvent Consumption Per Sample (%)*Sample Throughput Factor*
0.32230.01
0.61130.02
1.0731.83

* In comparison to the standard method, 1 mL/min, 22 minute run time, see Figure 1.

Engineering Facts

12 facts · every claim linked to a page
Selection

Narrow-bore GPC columns down to 2.1 mm id can be used with the Micro RID, enabling micro GPC with much lower solvent consumption.

Determines whether the detector is suitable for narrow-bore columns and micro GPC methods.

Source
“The Micro RID enables the use of GPC columns with narrow ids down to 2.1 mm, while maintaining highest resolution.”
Selection

The detector cell volume is 2 µL for the Micro RID versus 8 µL for the standard RID, which matters for resolution with narrow columns.

Cell dead volume is a key selection criterion for maintaining resolution with low-id columns.

Source
“The difference between the Micro RID and the standard RID is their cell volume of 2 and 8 µL, respectively.”
Selection

Modern UHPLC pumps are required to achieve sufficient flow-rate precision for low flow rates used with narrow-id GPC columns.

Guides pump selection when configuring a system for reduced flow rates.

Source
“To achieve enough precision for the flow rate, modern UHPLC pumps are required.”
Selection

Detectors with very low cell dead volumes are required to maintain resolution when using narrow-id columns.

Sets a detector specification requirement for narrow-bore GPC setups.

Source
“To maintain resolution, detectors with very low cell dead volumes are required.”
Installation

Two columns in series were connected with a 75 µm capillary of 105 mm length in the experimental setup.

Specifies the connection capillary dimensions needed when installing columns in series.

Source
“35 °C, 2 columns in series connected with 75 µm capillary (105 mm length)”
Operation

The method uses THF isocratic on channel B at flow rates of 0.06, 0.3, 0.6, and 1.0 mL/min.

Defines the mobile phase and channel assignment for method setup.

Source
“THF, isocratic, channel B”
Operation

The RID optical unit must be thermostatted at 35 °C with a data rate of 18 Hz and positive signal polarity.

Specifies detector operating conditions required for reproducible results.

Source
“Optical unit temperature: 35 °C”
Operation

At 0.6 mL/min on a 4.6 mm id column, throughput doubles with identical resolution for low molecular weight components.

Supports choosing a flow rate that balances throughput and resolution.

Source
“The increase to 0.6 mL/min exactly doubled the possible throughput, with identical resolution for the low molecular weight components.”
Operation

Increasing flow to 1 mL/min triples throughput but slightly compromises resolution of higher molecular weight compounds.

Warns that maximum throughput settings may degrade resolution for high-MW polymers.

Source
“The resolution of the low molecular weight components was still good, but slightly compromised in terms of higher molecular weight compounds”
Performance

Solvent consumption per sample is reduced to approximately 30% compared to the standard 7.5 mm id method.

Quantifies the solvent savings achievable with narrow-bore columns.

Source
“The solvent consumption per sample was reduced to approximately 30% for all cases”
Performance

For a 2.1 mm id column the calculated flow rate is 60 µL/min, requiring high flow accuracy to maintain retention times.

Defines the very low flow rate needed for micro GPC and the pump accuracy requirement.

Source
“The downwards calculation of the applied flow rate from the previously shown 4.6 mm id column to the 2.1 mm id column resulted in a flow rate of 60 µL/min.”
Performance

The lower dispersion volume of the 2 µL Micro RID cell provides enhanced resolution for low molecular weight oligomers.

Explains the resolution benefit of the micro cell for oligomer analysis.

Source
“Another advantage is the lower dispersion volume of the Micro RID cell (2 µL) compared to the standard RID (8 µL), which provides enhanced resolution for lower molecular weight oligomers”