Nota de aplicación/flujo de trabajo para el perfilado y la cuantificación de ácidos siálicos en glicoproteínas biofarmacéuticas utilizando el kit AdvanceBio Sialic Acid profiling and quantitation (GKK-407) y el kit AdvanceBio Total Sialic Acid quantitation; formato de placa de 96 pocillos; método RP-UHPLC/FLD de 10 minutos con detección MS opcional; demostrado en rituximab, etanercept, NISTmAb y cetuximab.

Technical Data

Parameter Value Condition
Sample volume 20 µL p.3 ↗ glycoprotein or sialic acid standard
Sialic acid release 2 hours at 80 °C p.3 ↗ acid hydrolysis
DMB labeling 3 hours at 50 °C p.3 ↗
LC system Agilent 1290 Infinity II LC System p.3 ↗
Autosampler Agilent Infinity multisampler (G7167B) p.2 ↗
Mass spectrometer Agilent 6545XT AdvanceBio LC/Q-TOF p.3 ↗
Ion source Dual AJS ESI p.3 ↗ 6545XT AdvanceBio LC/Q-TOF
Column Agilent InfinityLab PoroShell 120EC-C18 p.3 ↗
Mobile phase A 100 % p.3 ↗ 0.00 min, isocratic elution
Mobile phase B acetonitrile p.3 ↗
Flow rate 0.4 mL/min p.3 ↗ 0.00–10.00 min
Run time 10 min p.2 ↗ rapid method
Gas temperature 350 °C p.3 ↗ 6545XT AdvanceBio LC/Q-TOF
Sheath gas temperature 400 °C p.3 ↗ 6545XT AdvanceBio LC/Q-TOF
Sheath gas flow 12 L/min p.3 ↗ 6545XT AdvanceBio LC/Q-TOF
Nozzle voltage 1,800 V p.3 ↗ 6545XT AdvanceBio LC/Q-TOF
Fragmentor voltage 120 V p.3 ↗ 6545XT AdvanceBio LC/Q-TOF
Detection FLD or MS p.3 ↗
Software Agilent OpenLab CDS 2.3 p.2 ↗
Kit order number GKK-407 p.6 ↗ AdvanceBio Sialic Acid profiling and quantitation kit

Operating Limits

  • Sialic acid release temperature: 80 °C p.3 ↗
  • DMB labeling temperature: 50 °C p.3 ↗
  • LC run time: 10 min p.2 ↗

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
AdvanceBio Sialic Acid profiling and quantitation kit p.6 ↗ GKK-407 Kit used in the workflow