Induztr
Résumé technique Visionneuse de documents Modèles associés

Data Sheet Model B 28

Technical Reference Summary

Covered Models & Document Scope

Incubadoras estándar BINDER Model B 28 con ajuste mecánico, volumen interior 28 L. Dos variantes de tensión: B028-230V (Art.-No. 9010-0002, 230 V 1~ 50/60 Hz, enchufe CEE 7/7) y B028-120V (Art.-No. 9010-0067, 120 V 1~ 60 Hz, enchufe NEMA 5-15).

Technical Data

Parameter Value Condition
Temperature range 30...70 °C p.2 ↗ B028-230V / B028-120V
Temperature uniformity at 37°C 1.5 ± K p.2 ↗ B028-230V / B028-120V
Temperature fluctuation at 37°C 1.0 ± K p.2 ↗ B028-230V / B028-120V
Heating up time to 37°C 95 min p.2 ↗ B028-230V / B028-120V
Rated voltage 230 V p.2 ↗ B028-230V
Rated voltage 120 V p.2 ↗ B028-120V
Power frequency 50/60 Hz p.2 ↗ B028-230V
Power frequency 60 Hz p.2 ↗ B028-120V
Nominal power 0,25 kW p.2 ↗ B028-230V / B028-120V
Unit fuse 6,3 A p.2 ↗ B028-230V / B028-120V
Phase (nominal voltage) 1~ p.2 ↗ B028-230V / B028-120V
Interior volume 28 L p.2 ↗ B028-230V / B028-120V
Net weight of the unit (empty) 23 kg p.2 ↗ B028-230V / B028-120V
Load per rack 10 kg p.2 ↗ B028-230V / B028-120V
Permitted load 25 kg p.2 ↗ B028-230V / B028-120V
Wall clearance back 160 mm p.2 ↗ B028-230V / B028-120V
Wall clearance sidewise 100 mm p.2 ↗ B028-230V / B028-120V
Width net 580 mm p.2 ↗ B028-230V / B028-120V
Height net 405 mm p.2 ↗ B028-230V / B028-120V
Depth net 425 mm p.2 ↗ B028-230V / B028-120V
Interior width 400 mm p.2 ↗ B028-230V / B028-120V
Interior height 280 mm p.2 ↗ B028-230V / B028-120V
Interior depth 250 mm p.2 ↗ B028-230V / B028-120V
Unit doors 1 p.3 ↗ B028-230V / B028-120V
Number of shelves (std./max.) 2/4 p.3 ↗ B028-230V / B028-120V

Operating Limits

  • Temperature range: 30...70 °C p.2 ↗
  • Permitted load: 25 kg p.2 ↗
  • Load per rack: 10 kg p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗
  • Max number of shelves: 4 p.3 ↗

Type Code Decoder

Position Token Meaning
1-4 p.1 ↗ B028 Model series and size (B series incubator, 28 L interior volume)
6-8 p.1 ↗ 230 / 120 Rated voltage variant: 230 V or 120 V

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
B028-230V p.1 ↗ 9010-0002 230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard Standard version
B028-120V p.1 ↗ 9010-0067 120 V 1~ 60 Hz, NEMA 5-15 plug, Standard Standard version