Induztr
Résumé technique Visionneuse de documents Modèles associés

Data Sheet Model BD 115

Technical Reference Summary

Covered Models & Document Scope

Incubateur standard BINDER BD 115 (BD Avantgarde.Line), volume intérieur 112 L, convection naturelle. Deux variantes : BD115-230V (Art.-No. 9010-0325, 230 V 1~ 50/60 Hz, fiche CEE 7/7) et BD115UL-120V (Art.-No. 9010-0326, 120 V 1~ 60 Hz, fiche NEMA 5-15).

Technical Data

Parameter Value Condition
Temperature range +5 °C above ambient temperature to 100 °C °C p.2 ↗ BD115-230V / BD115UL-120V
Temperature uniformity at 37°C 0.3 ± K p.2 ↗ BD115-230V / BD115UL-120V
Temperature uniformity at 100°C 1.0 ± K p.2 ↗ BD115-230V / BD115UL-120V
Temperature fluctuation at 37°C 0.1 ± K p.2 ↗ BD115-230V / BD115UL-120V
Temperature fluctuation at 100°C 0.3 ± K p.2 ↗ BD115-230V / BD115UL-120V
Heating up time to 100°C 90 min p.2 ↗ BD115-230V / BD115UL-120V
Heating up time to 37°C 55 min p.2 ↗ BD115-230V / BD115UL-120V
Recovery time after door was opened for 30 s at 37°C 15 min p.2 ↗ BD115-230V / BD115UL-120V
Rated voltage 230 V p.2 ↗ BD115-230V
Rated voltage 120 V p.2 ↗ BD115UL-120V
Power frequency 50/60 Hz p.2 ↗ BD115-230V
Power frequency 60 Hz p.2 ↗ BD115UL-120V
Nominal power 0,35 kW p.2 ↗ BD115-230V / BD115UL-120V
Unit fuse 6,3 A p.2 ↗ BD115-230V
Unit fuse 12,5 A p.2 ↗ BD115UL-120V
Phase (nominal voltage) 1~ p.2 ↗ BD115-230V / BD115UL-120V
Interior volume 112 L p.2 ↗ BD115-230V / BD115UL-120V
Net weight of the unit (empty) 54 kg p.2 ↗ BD115-230V / BD115UL-120V
Load per rack 30 kg p.2 ↗ BD115-230V / BD115UL-120V
Permitted load 150 kg p.2 ↗ BD115-230V / BD115UL-120V
Wall clearance back 160 mm p.2 ↗ BD115-230V / BD115UL-120V
Wall clearance sidewise 100 mm p.2 ↗ BD115-230V / BD115UL-120V
Width net 710 mm p.2 ↗ BD115-230V / BD115UL-120V
Height net 705 mm p.2 ↗ BD115-230V / BD115UL-120V
Depth net 605 mm p.2 ↗ BD115-230V / BD115UL-120V

Operating Limits

  • Temperature range (min): +5 °C above ambient temperature p.2 ↗
  • Temperature range (max): 100 °C p.2 ↗
  • Permitted load: 150 kg p.2 ↗
  • Load per rack: 30 kg p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-2 p.1 ↗ BD Series: Standard-Incubator with natural convection (BD Avantgarde.Line)
3-5 p.1 ↗ 115 Model/size: interior volume 112 L
suffix p.1 ↗ UL UL variant for 120 V / 60 Hz mains (NEMA 5-15 plug)
suffix p.1 ↗ 230V 230 V 1~ 50/60 Hz variant (CEE 7/7 plug)
suffix p.1 ↗ 120V 120 V 1~ 60 Hz variant (NEMA 5-15 plug)

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
BD115-230V p.1 ↗ 9010-0325 230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard Standard version
BD115UL-120V p.1 ↗ 9010-0326 120 V 1~ 60 Hz, NEMA 5-15 plug, Standard UL version

Modèles liés