Data Sheet Model BF 260
Technical Reference Summary
Covered Models & Document Scope
BINDER BF 260 Incubateurs standard à convection forcée (BF Avantgarde.Line), volume intérieur 257 L. Deux variantes standard : BF260-230V (Art.-No. 9010-0319, 230 V 1~ 50/60 Hz, fiche CEE 7/7) et BF260UL-120V (Art.-No. 9010-0320, 120 V 1~ 60 Hz, fiche NEMA 5-15). Le document répertorie également des accessoires et services en option avec leurs propres numéros d'article.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Temperature range | +7 °C above ambient temperature to 100 °C p.2 ↗ | BF260-230V / BF260UL-120V |
| Temperature uniformity at 37°C | 0.2 ± K p.2 ↗ | BF260-230V / BF260UL-120V |
| Temperature uniformity at 100°C | 1.2 ± K p.2 ↗ | BF260-230V / BF260UL-120V |
| Temperature fluctuation at 37°C | 0.1 ± K p.2 ↗ | BF260-230V / BF260UL-120V |
| Temperature fluctuation at 100°C | 0.3 ± K p.2 ↗ | BF260-230V / BF260UL-120V |
| Heating up time to 100°C | 30 min p.2 ↗ | BF260-230V / BF260UL-120V |
| Heating up time to 37°C | 7 min p.2 ↗ | BF260-230V / BF260UL-120V |
| Recovery time after door was opened for 30 s at 37°C | 3 min p.2 ↗ | BF260-230V / BF260UL-120V |
| Rated Voltage | 230 V p.2 ↗ | BF260-230V |
| Rated Voltage | 120 V p.2 ↗ | BF260UL-120V |
| Power frequency | 50/60 Hz p.2 ↗ | BF260-230V |
| Power frequency | 60 Hz p.2 ↗ | BF260UL-120V |
| Nominal power | 0,9 kW p.2 ↗ | BF260-230V |
| Nominal power | 1 kW p.2 ↗ | BF260UL-120V |
| Unit fuse | 8,0 A p.2 ↗ | BF260-230V |
| Unit fuse | 12,5 A p.2 ↗ | BF260UL-120V |
| Phase (Nominal voltage) | 1~ p.2 ↗ | BF260-230V / BF260UL-120V |
| Interior volume | 257 L p.2 ↗ | BF260-230V / BF260UL-120V |
| Net weight of the unit (empty) | 85 kg p.2 ↗ | BF260-230V / BF260UL-120V |
| Load per rack | 40 kg p.2 ↗ | BF260-230V / BF260UL-120V |
| Permitted load | 270 kg p.2 ↗ | BF260-230V / BF260UL-120V |
| Wall clearance back | 160 mm p.2 ↗ | BF260-230V / BF260UL-120V |
| Wall clearance sidewise | 100 mm p.2 ↗ | BF260-230V / BF260UL-120V |
| Width net | 810 mm p.2 ↗ | BF260-230V / BF260UL-120V |
| Height net | 845 mm p.2 ↗ | BF260-230V / BF260UL-120V |
Operating Limits
- Temperature range (min): +7 °C above ambient temperature p.2 ↗
- Temperature range (max): 100 °C p.2 ↗
- Permitted load: 270 kg p.2 ↗
- Load per rack: 40 kg p.2 ↗
- Wall clearance back: 160 mm p.2 ↗
- Wall clearance sidewise: 100 mm p.2 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-2 p.1 ↗ | BF | Series: Standard-Incubator with forced convection (BF Avantgarde.Line) |
| 3-5 p.1 ↗ | 260 | Model / size (interior volume 257 L) |
| suffix p.1 ↗ | UL | UL variant for 120 V / 60 Hz mains (NEMA 5-15 plug); absent = standard 230 V version |
| voltage field p.1 ↗ | 230V | Mains version 230 V 1~ 50/60 Hz, CEE 7/7 plug |
| voltage field p.1 ↗ | 120V | Mains version 120 V 1~ 60 Hz, NEMA 5-15 plug |
Ordering Data
| Ordering Code | Part No. | Variant / Code | S'applique à | Note |
|---|---|---|---|---|
| BF260-230V p.1 ↗ | 9010-0319 | 230 V 1~ 50/60 Hz, 8,0 A, CEE 7/7 plug, Standard | Standard model version | |
| BF260UL-120V p.1 ↗ | 9010-0320 | 120 V 1~ 60 Hz, 12,5 A, NEMA 5-15 plug, Standard | UL standard model version |