Data Sheet Model BF 260

Technical Reference Summary

Covered Models & Document Scope

BINDER BF 260 Incubateurs standard à convection forcée (BF Avantgarde.Line), volume intérieur 257 L. Deux variantes standard : BF260-230V (Art.-No. 9010-0319, 230 V 1~ 50/60 Hz, fiche CEE 7/7) et BF260UL-120V (Art.-No. 9010-0320, 120 V 1~ 60 Hz, fiche NEMA 5-15). Le document répertorie également des accessoires et services en option avec leurs propres numéros d'article.

Technical Data

Parameter Value Condition
Temperature range +7 °C above ambient temperature to 100 °C p.2 ↗ BF260-230V / BF260UL-120V
Temperature uniformity at 37°C 0.2 ± K p.2 ↗ BF260-230V / BF260UL-120V
Temperature uniformity at 100°C 1.2 ± K p.2 ↗ BF260-230V / BF260UL-120V
Temperature fluctuation at 37°C 0.1 ± K p.2 ↗ BF260-230V / BF260UL-120V
Temperature fluctuation at 100°C 0.3 ± K p.2 ↗ BF260-230V / BF260UL-120V
Heating up time to 100°C 30 min p.2 ↗ BF260-230V / BF260UL-120V
Heating up time to 37°C 7 min p.2 ↗ BF260-230V / BF260UL-120V
Recovery time after door was opened for 30 s at 37°C 3 min p.2 ↗ BF260-230V / BF260UL-120V
Rated Voltage 230 V p.2 ↗ BF260-230V
Rated Voltage 120 V p.2 ↗ BF260UL-120V
Power frequency 50/60 Hz p.2 ↗ BF260-230V
Power frequency 60 Hz p.2 ↗ BF260UL-120V
Nominal power 0,9 kW p.2 ↗ BF260-230V
Nominal power 1 kW p.2 ↗ BF260UL-120V
Unit fuse 8,0 A p.2 ↗ BF260-230V
Unit fuse 12,5 A p.2 ↗ BF260UL-120V
Phase (Nominal voltage) 1~ p.2 ↗ BF260-230V / BF260UL-120V
Interior volume 257 L p.2 ↗ BF260-230V / BF260UL-120V
Net weight of the unit (empty) 85 kg p.2 ↗ BF260-230V / BF260UL-120V
Load per rack 40 kg p.2 ↗ BF260-230V / BF260UL-120V
Permitted load 270 kg p.2 ↗ BF260-230V / BF260UL-120V
Wall clearance back 160 mm p.2 ↗ BF260-230V / BF260UL-120V
Wall clearance sidewise 100 mm p.2 ↗ BF260-230V / BF260UL-120V
Width net 810 mm p.2 ↗ BF260-230V / BF260UL-120V
Height net 845 mm p.2 ↗ BF260-230V / BF260UL-120V

Operating Limits

  • Temperature range (min): +7 °C above ambient temperature p.2 ↗
  • Temperature range (max): 100 °C p.2 ↗
  • Permitted load: 270 kg p.2 ↗
  • Load per rack: 40 kg p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-2 p.1 ↗ BF Series: Standard-Incubator with forced convection (BF Avantgarde.Line)
3-5 p.1 ↗ 260 Model / size (interior volume 257 L)
suffix p.1 ↗ UL UL variant for 120 V / 60 Hz mains (NEMA 5-15 plug); absent = standard 230 V version
voltage field p.1 ↗ 230V Mains version 230 V 1~ 50/60 Hz, CEE 7/7 plug
voltage field p.1 ↗ 120V Mains version 120 V 1~ 60 Hz, NEMA 5-15 plug

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
BF260-230V p.1 ↗ 9010-0319 230 V 1~ 50/60 Hz, 8,0 A, CEE 7/7 plug, Standard Standard model version
BF260UL-120V p.1 ↗ 9010-0320 120 V 1~ 60 Hz, 12,5 A, NEMA 5-15 plug, Standard UL standard model version