Induztr
Résumé technique Visionneuse de documents Modèles associés

Data Sheet Model CB S 170

Technical Reference Summary

Covered Models & Document Scope

Incubateur CO2 modèle CB-S 170 avec stérilisation à air chaud, volume intérieur 170 L. Variantes : CBS170-230V (standard, 200...230 V, fiche CEE 7/7, Art.-No. 9040-0219), CBS170-230V-L (charnière à gauche, 200...230 V, Art.-No. 9040-0220), CBS170UL-120V (standard, 100...120 V, fiche NEMA 5-20, Art.-No. 9040-0221), CBS170UL-120V-L (charnière à gauche, 100...120 V, Art.-No. 9040-0222).

Technical Data

Parameter Value Condition
Interior volume 170 L p.3 ↗ all variants
Rated voltage 200...230 V p.3 ↗ CBS170-230V / CBS170-230V-L
Rated voltage 100...120 V p.3 ↗ CBS170UL-120V / CBS170UL-120V-L
Power frequency 50/60 Hz p.3 ↗ all variants
Phase (Nominal voltage) 1~ p.3 ↗ all variants
Nominal power 1,3 kW p.3 ↗ all variants
Unit fuse 10,0 A p.3 ↗ CBS170-230V / CBS170-230V-L
Unit fuse 16,0 A p.3 ↗ CBS170UL-120V / CBS170UL-120V-L
Temperature range to 50 °C p.3 ↗ all variants
Temperature uniformity at 37°C 0.3 ± K p.3 ↗ all variants
Temperature fluctuation at 37°C 0.1 ± K p.3 ↗ all variants
Recovery time 6 min p.3 ↗ all variants
Humidity range 90 ...95 % RH p.3 ↗ all variants
CO₂ range 0 ...20 Vol.-% CO₂ p.3 ↗ all variants
CO₂ measuring technology IR p.3 ↗ all variants
CO₂ recovery time 5 min p.3 ↗ all variants
Net weight of the unit (empty) 96 kg p.3 ↗ all variants
Load per rack 10 kg p.3 ↗ all variants
Permitted load 30 kg p.3 ↗ all variants
Wall clearance back 100 mm p.3 ↗ all variants
Wall clearance sidewise 50 mm p.3 ↗ all variants
Width net 680 mm p.4 ↗ all variants
Height net 870 mm p.4 ↗ all variants
Depth net 715 mm p.4 ↗ all variants
Sound-pressure level 41 dB(A) p.4 ↗ all variants

Operating Limits

  • max temperature: 50 °C p.3 ↗
  • humidity range: 90 ...95 % RH p.3 ↗
  • CO₂ range: 0 ...20 Vol.-% CO₂ p.3 ↗
  • permitted load: 30 kg p.3 ↗
  • load per rack: 10 kg p.3 ↗
  • wall clearance back: 100 mm p.3 ↗

Type Code Decoder

Position Token Meaning
1-3 p.1 ↗ CBS Series: CO2 incubator with hot air sterilization
4-6 p.1 ↗ 170 Interior volume size: 170 L
suffix p.1 ↗ UL UL version for 100...120 V supply with NEMA 5-20 plug
suffix p.1 ↗ -230V Version for 200...230 V supply with CEE 7/7 plug
suffix p.1 ↗ -120V Version for 100...120 V supply with NEMA 5-20 plug
suffix p.1 ↗ -L Option: with left-hand hinge

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
CBS170-230V p.1 ↗ 9040-0219 200...230 V 1~ 50/60 Hz, 10,0 A, CEE 7/7 plug, standard Standard model
CBS170-230V-L p.1 ↗ 9040-0220 200...230 V 1~ 50/60 Hz, 10,0 A, CEE 7/7 plug, with left-hand hinge Left-hand hinge
CBS170UL-120V p.1 ↗ 9040-0221 100...120 V 1~ 50/60 Hz, 16,0 A, NEMA 5-20 plug, standard UL version
CBS170UL-120V-L p.2 ↗ 9040-0222 100...120 V 1~ 50/60 Hz, 16,0 A, NEMA 5-20 plug, with left-hand hinge UL version, left-hand hinge

Modèles liés