Induztr
Résumé technique Visionneuse de documents Modèles associés

Data Sheet Model ED 115

Technical Reference Summary

Covered Models & Document Scope

Cámara de secado y calefacción modelo ED 115 con convección natural. Dos variantes estándar: ED115-230V (Art.-No. 9010-0335, 230 V 1~ 50/60 Hz, enchufe CEE 7/7) y ED115UL-120V (Art.-No. 9010-0336, 120 V 1~ 60 Hz, enchufe NEMA 5-15).

Technical Data

Parameter Value Condition
Temperature range +5 °C above ambient temperature to 300 °C p.2 ↗ ED115-230V / ED115UL-120V
Temperature uniformity at 150°C 1.5 ± K p.2 ↗ ED115-230V / ED115UL-120V
Temperature fluctuation at 150°C 0.4 ± K p.2 ↗ ED115-230V / ED115UL-120V
Heating up time to 150°C 45 min p.2 ↗ ED115-230V / ED115UL-120V
Recovery time after door was opened for 30 s at 150°C 16 min p.2 ↗ ED115-230V / ED115UL-120V
Air change rate at 100°C 6 x/h p.2 ↗ ED115-230V / ED115UL-120V
Rated voltage 230 V p.2 ↗ ED115-230V
Rated voltage 120 V p.2 ↗ ED115UL-120V
Power frequency 50/60 Hz p.2 ↗ ED115-230V
Power frequency 60 Hz p.2 ↗ ED115UL-120V
Nominal power 1,25 kW p.2 ↗ ED115-230V
Nominal power 1,35 kW p.2 ↗ ED115UL-120V
Unit fuse 6,3 A p.2 ↗ ED115-230V
Unit fuse 12,5 A p.2 ↗ ED115UL-120V
Phase (nominal voltage) 1~ p.2 ↗ ED115-230V / ED115UL-120V
Interior volume 114 L p.2 ↗ ED115-230V / ED115UL-120V
Net weight of the unit (empty) 54 kg p.2 ↗ ED115-230V / ED115UL-120V
Load per rack 30 kg p.2 ↗ ED115-230V / ED115UL-120V
Permitted load 150 kg p.2 ↗ ED115-230V / ED115UL-120V
Wall clearance back 160 mm p.2 ↗ ED115-230V / ED115UL-120V
Wall clearance sidewise 100 mm p.2 ↗ ED115-230V / ED115UL-120V
Width net 710 mm p.2 ↗ ED115-230V / ED115UL-120V
Height net 705 mm p.2 ↗ ED115-230V / ED115UL-120V
Depth net 605 mm p.2 ↗ ED115-230V / ED115UL-120V
Energy consumption at 150°C 245 Wh/h p.3 ↗ ED115-230V / ED115UL-120V

Operating Limits

  • Temperature range: +5 °C above ambient temperature to 300 °C p.2 ↗
  • Max permitted load: 150 kg p.2 ↗
  • Max load per rack: 30 kg p.2 ↗
  • Min wall clearance back: 160 mm p.2 ↗
  • Min wall clearance sidewise: 100 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-2 p.1 ↗ ED Series: drying and heating chamber with natural convection
3-5 p.1 ↗ 115 Model size (interior volume class, 114 L)
suffix p.1 ↗ UL UL variant for 120 V / 60 Hz mains (NEMA 5-15 plug)
suffix p.1 ↗ 230V Standard variant for 230 V 1~ 50/60 Hz mains (CEE 7/7 plug)
suffix p.1 ↗ 120V Variant for 120 V 1~ 60 Hz mains (NEMA 5-15 plug)

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
ED115-230V p.1 ↗ 9010-0335 230 V 1~ 50/60 Hz, 6,3 A, CEE 7/7 plug, Standard Standard model version
ED115UL-120V p.1 ↗ 9010-0336 120 V 1~ 60 Hz, 12,5 A, NEMA 5-15 plug, Standard UL model version

Modèles liés