Induztr
Résumé technique Visionneuse de documents Modèles associés

Data Sheet Model ED 23

Technical Reference Summary

Covered Models & Document Scope

Cámara de secado y calentamiento BINDER ED 23 con convección natural, volumen interior de 20 L. Dos variantes estándar: ED023-230V (230 V 1~ 50/60 Hz, enchufe CEE 7/7, Art.-No. 9010-0190) y ED023UL-120V (120 V 1~ 60 Hz, enchufe NEMA 5-20, Art.-No. 9010-0192).

Technical Data

Parameter Value Condition
Temperature range +5 °C above ambient temperature to 300 °C p.2 ↗ ED023-230V / ED023UL-120V
Temperature uniformity at 150°C 2.8 ± K p.2 ↗ ED023-230V / ED023UL-120V
Temperature fluctuation at 150°C 0.5 ± K p.2 ↗ ED023-230V / ED023UL-120V
Heating up time to 150°C 28 min p.2 ↗ ED023-230V / ED023UL-120V
Recovery time after door was opened for 30 s at 150°C 5 min p.2 ↗ ED023-230V / ED023UL-120V
Air change rate at 150°C 13 x/h p.2 ↗ ED023-230V / ED023UL-120V
Rated voltage 230 V p.2 ↗ ED023-230V
Rated voltage 120 V p.2 ↗ ED023UL-120V
Power frequency 50/60 Hz p.2 ↗ ED023-230V
Power frequency 60 Hz p.2 ↗ ED023UL-120V
Nominal power 0,8 kW p.2 ↗ ED023-230V / ED023UL-120V
Unit fuse 10,0 A p.2 ↗ ED023-230V
Unit fuse 12,5 A p.2 ↗ ED023UL-120V
Phase (nominal voltage) 1~ p.2 ↗ ED023-230V / ED023UL-120V
Interior volume 20 L p.2 ↗ ED023-230V / ED023UL-120V
Net weight of the unit (empty) 27 kg p.2 ↗ ED023-230V / ED023UL-120V
Load per rack 12 kg p.2 ↗ ED023-230V / ED023UL-120V
Permitted load 25 kg p.2 ↗ ED023-230V / ED023UL-120V
Wall clearance back 100 mm p.2 ↗ ED023-230V / ED023UL-120V
Wall clearance sidewise 50 mm p.2 ↗ ED023-230V / ED023UL-120V
Width net 435 mm p.2 ↗ ED023-230V / ED023UL-120V
Height net 495 mm p.2 ↗ ED023-230V / ED023UL-120V
Depth net 520 mm p.2 ↗ ED023-230V / ED023UL-120V
Interior width 222 mm p.3 ↗ ED023-230V / ED023UL-120V
Interior height 330 mm p.3 ↗ ED023-230V / ED023UL-120V

Operating Limits

  • Temperature range: +5 °C above ambient temperature to 300 °C p.2 ↗
  • Wall clearance back: 100 mm p.2 ↗
  • Wall clearance sidewise: 50 mm p.2 ↗
  • Permitted load: 25 kg p.2 ↗
  • Load per rack: 12 kg p.2 ↗

Type Code Decoder

Position Token Meaning
1-5 p.1 ↗ ED023 Model series ED (drying/heating chamber with natural convection), size 23
suffix p.1 ↗ UL UL variant for 120 V / 60 Hz mains (NEMA 5-20 plug); absent = standard 230 V / 50/60 Hz variant (CEE 7/7 plug)
voltage field p.1 ↗ 230V Rated voltage 230 V, 1~, 50/60 Hz, unit fuse 10,0 A
voltage field p.1 ↗ 120V Rated voltage 120 V, 1~, 60 Hz, unit fuse 12,5 A

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
ED023-230V p.1 ↗ 9010-0190 230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard Standard model version
ED023UL-120V p.1 ↗ 9010-0192 120 V 1~ 60 Hz, NEMA 5-20 plug, Standard UL model version

Modèles liés