Data Sheet Model ED 260

Technical Reference Summary

Covered Models & Document Scope

Cámara de secado y calefacción BINDER ED 260 con convección natural, 255 L de volumen interior. Dos variantes estándar: ED260-230V (Art.-No. 9010-0339, 230 V 1~ 50/60 Hz, enchufe CEE 7/7) y ED260UL-240V (Art.-No. 9010-0340, 240 V 1~ 60 Hz, enchufe NEMA 6-20). El documento también enumera accesorios y servicios opcionales.

Technical Data

Parameter Value Condition
Temperature range +5 °C above ambient temperature to 300 °C p.2 ↗ ED260-230V / ED260UL-240V
Temperature uniformity at 150°C 1.8 ± K p.2 ↗ ED260-230V / ED260UL-240V
Temperature fluctuation at 150°C 0.8 ± K p.2 ↗ ED260-230V / ED260UL-240V
Heating up time to 150°C 50 min p.2 ↗ ED260-230V / ED260UL-240V
Recovery time after door was opened for 30 s at 150°C 20 min p.2 ↗ ED260-230V / ED260UL-240V
Air change rate at 100°C 3 x/h p.2 ↗ ED260-230V / ED260UL-240V
Rated voltage 230 V p.2 ↗ ED260-230V
Rated voltage 240 V p.2 ↗ ED260UL-240V
Power frequency 50/60 Hz p.2 ↗ ED260-230V
Power frequency 60 Hz p.2 ↗ ED260UL-240V
Nominal power 2,25 kW p.2 ↗ ED260-230V
Nominal power 2,45 kW p.2 ↗ ED260UL-240V
Unit fuse 12,5 A p.2 ↗ ED260-230V
Unit fuse 16,0 A p.2 ↗ ED260UL-240V
Phase (nominal voltage) 1~ p.2 ↗ ED260-230V / ED260UL-240V
Interior volume 255 L p.2 ↗ ED260-230V / ED260UL-240V
Net weight of the unit (empty) 85 kg p.2 ↗ ED260-230V / ED260UL-240V
Load per rack 40 kg p.2 ↗ ED260-230V / ED260UL-240V
Permitted load 270 kg p.2 ↗ ED260-230V / ED260UL-240V
Wall clearance back 160 mm p.2 ↗ ED260-230V / ED260UL-240V
Wall clearance sidewise 100 mm p.2 ↗ ED260-230V / ED260UL-240V
Width net 810 mm p.2 ↗ ED260-230V / ED260UL-240V
Height net 940 mm p.2 ↗ ED260-230V / ED260UL-240V
Depth net 760 mm p.2 ↗ ED260-230V / ED260UL-240V
Energy consumption at 150°C 355 Wh/h p.3 ↗ ED260-230V / ED260UL-240V

Operating Limits

  • Temperature range: +5 °C above ambient temperature to 300 °C p.2 ↗
  • Permitted load: 270 kg p.2 ↗
  • Load per rack: 40 kg p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-2 p.1 ↗ ED Series: drying and heating chamber with natural convection
3-5 p.1 ↗ 260 Model / size (interior volume class, 255 L)
suffix p.1 ↗ UL UL variant for 240 V / 60 Hz with NEMA 6-20 plug (absent = standard 230 V 50/60 Hz with CEE 7/7 plug)
suffix p.1 ↗ 230V Rated voltage 230 V, 1~, 50/60 Hz
suffix p.1 ↗ 240V Rated voltage 240 V, 1~, 60 Hz

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
ED260-230V p.1 ↗ 9010-0339 230 V 1~ 50/60 Hz, 12,5 A, CEE 7/7 plug, Standard Standard model version
ED260UL-240V p.1 ↗ 9010-0340 240 V 1~ 60 Hz, 16,0 A, NEMA 6-20 plug, Standard UL model version