Induztr
Résumé technique Visionneuse de documents Modèles associés

Data Sheet Model ED 56

Technical Reference Summary

Covered Models & Document Scope

Cámara de secado y calefacción modelo ED 56 con convección natural, volumen interior de 57 L. Dos variantes estándar: ED056-230V (230 V 1~ 50/60 Hz, enchufe CEE 7/7, Art.-No. 9010-0333) y ED056UL-120V (120 V 1~ 60 Hz, enchufe NEMA 5-15, Art.-No. 9010-0334).

Technical Data

Parameter Value Condition
Temperature range +5 °C above ambient temperature to 300 °C p.2 ↗ ED056-230V / ED056UL-120V
Temperature uniformity at 150°C 2.3 ± K p.2 ↗ ED056-230V / ED056UL-120V
Temperature fluctuation at 150°C 0.4 ± K p.2 ↗ ED056-230V / ED056UL-120V
Heating up time to 150°C 50 min p.2 ↗ ED056-230V / ED056UL-120V
Recovery time after door was opened for 30 s at 150°C 19 min p.2 ↗ ED056-230V / ED056UL-120V
Air change rate at 100°C 8 x/h p.2 ↗ ED056-230V / ED056UL-120V
Rated voltage 230 V p.2 ↗ ED056-230V
Rated voltage 120 V p.2 ↗ ED056UL-120V
Power frequency 50/60 Hz p.2 ↗ ED056-230V
Power frequency 60 Hz p.2 ↗ ED056UL-120V
Nominal power 1,05 kW p.2 ↗ ED056-230V
Nominal power 1,15 kW p.2 ↗ ED056UL-120V
Unit fuse 6,3 A p.2 ↗ ED056-230V
Unit fuse 12,5 A p.2 ↗ ED056UL-120V
Phase (nominal voltage) 1~ p.2 ↗ ED056-230V / ED056UL-120V
Interior volume 57 L p.2 ↗ ED056-230V / ED056UL-120V
Net weight of the unit (empty) 39 kg p.2 ↗ ED056-230V / ED056UL-120V
Load per rack 30 kg p.2 ↗ ED056-230V / ED056UL-120V
Permitted load 70 kg p.2 ↗ ED056-230V / ED056UL-120V
Wall clearance back 160 mm p.2 ↗ ED056-230V / ED056UL-120V
Wall clearance sidewise 100 mm p.2 ↗ ED056-230V / ED056UL-120V
Width net 560 mm p.2 ↗ ED056-230V / ED056UL-120V
Height net 625 mm p.2 ↗ ED056-230V / ED056UL-120V
Depth net 565 mm p.2 ↗ ED056-230V / ED056UL-120V
Interior width 360 mm p.3 ↗ ED056-230V / ED056UL-120V

Operating Limits

  • Temperature range: +5 °C above ambient temperature to 300 °C p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗
  • Permitted load: 70 kg p.2 ↗
  • Load per rack: 30 kg p.2 ↗

Type Code Decoder

Position Token Meaning
1-2 p.1 ↗ ED Series: drying and heating chamber with natural convection
3-5 p.1 ↗ 056 Model size (56 / 57 L interior volume)
6-8 p.1 ↗ 230 / UL-120 Voltage variant: 230 V standard or UL 120 V version
suffix p.1 ↗ V Voltage designation

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
ED056-230V p.1 ↗ 9010-0333 230 V 1~ 50/60 Hz, 6,3 A, CEE 7/7 plug, Standard Standard model version
ED056UL-120V p.1 ↗ 9010-0334 120 V 1~ 60 Hz, 12,5 A, NEMA 5-15 plug, Standard UL model version

Modèles liés