Data Sheet Model FP 56

Technical Reference Summary

Covered Models & Document Scope

Cámara de secado y calefacción modelo FP 56 con convección forzada y funciones de programa, volumen interior 60 L. Dos variantes estándar: FP056-230V (Art.-No. 9010-0363, 230 V 1~ 50/60 Hz, enchufe CEE 7/7) y FP056UL-120V (Art.-No. 9010-0367, 120 V 1~ 60 Hz, enchufe NEMA 5-20).

Technical Data

Parameter Value Condition
Temperature range +10 °C above ambient temperature to 300 °C p.2 ↗ FP056-230V / FP056UL-120V
Temperature uniformity at 150°C 1.4 ± K p.2 ↗ FP056-230V / FP056UL-120V
Temperature fluctuation at 150°C 0.4 ± K p.2 ↗ FP056-230V / FP056UL-120V
Heating up time to 150°C 15 min p.2 ↗ FP056-230V / FP056UL-120V
Recovery time after door was opened for 30 s at 150°C 4 min p.2 ↗ FP056-230V / FP056UL-120V
Air change rate at 100°C 80 x/h p.2 ↗ FP056-230V / FP056UL-120V
Rated voltage 230 V p.2 ↗ FP056-230V
Rated voltage 120 V p.2 ↗ FP056UL-120V
Power frequency 50/60 Hz p.2 ↗ FP056-230V
Power frequency 60 Hz p.2 ↗ FP056UL-120V
Nominal power 1,2 kW p.2 ↗ FP056-230V / FP056UL-120V
Unit fuse 10,0 A p.2 ↗ FP056-230V
Unit fuse 16,0 A p.2 ↗ FP056UL-120V
Phase (nominal voltage) 1~ p.2 ↗ FP056-230V / FP056UL-120V
Interior volume 60 L p.2 ↗ FP056-230V / FP056UL-120V
Net weight of the unit (empty) 41 kg p.2 ↗ FP056-230V / FP056UL-120V
Load per rack 30 kg p.2 ↗ FP056-230V / FP056UL-120V
Permitted load 70 kg p.2 ↗ FP056-230V / FP056UL-120V
Wall clearance back 160 mm p.2 ↗ FP056-230V / FP056UL-120V
Wall clearance sidewise 100 mm p.2 ↗ FP056-230V / FP056UL-120V
Width net 560 mm p.2 ↗ FP056-230V / FP056UL-120V
Height net 622 mm p.2 ↗ FP056-230V / FP056UL-120V
Depth net 596 mm p.2 ↗ FP056-230V / FP056UL-120V
Sound-pressure level 43 dB(A) p.3 ↗ FP056-230V / FP056UL-120V
Energy consumption at 100°C 180 Wh/h p.3 ↗ FP056-230V / FP056UL-120V

Operating Limits

  • Temperature range: +10 °C above ambient temperature to 300 °C p.2 ↗
  • Permitted load: 70 kg p.2 ↗
  • Load per rack: 30 kg p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-2 p.1 ↗ FP Series: drying and heating chamber with forced convection and program functions
3-5 p.1 ↗ 056 Model size / interior volume 60 L (FP 56)
suffix p.1 ↗ UL UL variant for 120 V / 60 Hz supply with NEMA 5-20 plug (absent = standard 230 V variant)
suffix p.1 ↗ 230V Mains voltage variant 230 V 1~ 50/60 Hz, CEE 7/7 plug
suffix p.1 ↗ 120V Mains voltage variant 120 V 1~ 60 Hz, NEMA 5-20 plug

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
FP056-230V p.1 ↗ 9010-0363 Standard, 230 V 1~ 50/60 Hz, CEE 7/7 plug Standard model version
FP056UL-120V p.1 ↗ 9010-0367 Standard, 120 V 1~ 60 Hz, NEMA 5-20 plug UL model version