Data Sheet Model KBF S ECO 400
Technical Reference Summary
Covered Models & Document Scope
Cámara climática constante modelo KBF-S ECO 400 con refrigeración termoeléctrica. Dos variantes estándar: KBFSECO400-230V (n.º de art. 9020-0463, 200...230 V 1~ 50/60 Hz, enchufe CEE 7/7) y KBFSECO400UL-120V (n.º de art. 9020-0464, 100...120 V 1~ 50/60 Hz, enchufe NEMA 5-20).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Temperature range | 0...70 °C p.2 ↗ | max. 26 °C under ambient temperature |
| Temperature range with humidity | 5...70 °C p.2 ↗ | max. 26 °C under ambient temperature |
| Humidity range | 10...80 % RH p.2 ↗ | |
| Max. heat compensation at 40°C | 300 W p.2 ↗ | |
| Temperature uniformity at 25°C and 60% RH | 0.1 ± K p.2 ↗ | |
| Temperature uniformity at 40°C and 75% RH | 0.5 ± K p.2 ↗ | |
| Temperature fluctuation at 25°C and 60% RH | 0.1 ± K p.2 ↗ | |
| Temperature fluctuation at 40°C and 75% RH | 0.1 ± K p.2 ↗ | |
| Humidity fluctuation at 25°C and 60% RH | 0.3 ± % RH p.2 ↗ | |
| Humidity fluctuation at 40°C and 75% RH | 0.3 ± % RH p.2 ↗ | |
| Rated voltage | 200...230 V p.2 ↗ | KBFSECO400-230V |
| Rated voltage | 100...120 V p.2 ↗ | KBFSECO400UL-120V |
| Power frequency | 50/60 Hz p.2 ↗ | |
| Nominal power | 1,1 kW p.2 ↗ | |
| Unit fuse | 16,0 A p.2 ↗ | |
| Phase (nominal voltage) | 1~ p.2 ↗ | |
| Interior volume | 400 L p.2 ↗ | |
| Net weight of the unit (empty) | 192 kg p.2 ↗ | |
| Load per rack | 30 kg p.2 ↗ | |
| Permitted load | 120 kg p.2 ↗ | |
| Sound-pressure level | 45 dB(A) p.3 ↗ | |
| Energy consumption at 25°C and 60% RH | 60 Wh/h p.3 ↗ | |
| Energy consumption at 40°C and 75% RH | 70 Wh/h p.3 ↗ | |
| Number of shelves (std./max.) | 2/15 p.3 ↗ | |
| Width net | 925 mm p.3 ↗ |
Operating Limits
- Temperature range: 0...70 °C (max. 26 °C under ambient temperature) p.2 ↗
- Temperature range with humidity: 5...70 °C (max. 26 °C under ambient temperature) p.2 ↗
- Humidity range: 10...80 % RH p.2 ↗
- Max. heat compensation at 40°C: 300 W p.2 ↗
- Permitted load: 120 kg p.2 ↗
- Load per rack: 30 kg p.2 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-10 p.1 ↗ | KBFSECO400 | Model series KBF-S ECO, size 400 |
| suffix p.1 ↗ | -230V | Mains voltage variant 200...230 V 1~ 50/60 Hz, CEE 7/7 plug |
| suffix p.1 ↗ | UL-120V | Mains voltage variant 100...120 V 1~ 50/60 Hz, NEMA 5-20 plug |
Ordering Data
| Ordering Code | Part No. | Variant / Code | S'applique à | Note |
|---|---|---|---|---|
| KBFSECO400-230V p.1 ↗ | 9020-0463 | 200...230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard | ||
| KBFSECO400UL-120V p.1 ↗ | 9020-0464 | 100...120 V 1~ 50/60 Hz, NEMA 5-20 plug, Standard |