0714 EN MK CO2 E6 UM 10 2026
Technical Reference Summary
Covered Models & Document Scope
MK (E6) series alternating climate chambers with program control: MK 240, MK 400, MK 720, MK 1020, each in 400 V / 50 Hz and 480 V / 60 Hz voltage variants. Art. nos. 9020-0513/9120-0513 (MK 240 400V/50Hz), 9020-0515/9120-0515 (MK 240 480V/60Hz), 9020-0518/9120-0518 (MK 400 400V/50Hz), 9020-0520/9120-0520 (MK 400 480V/60Hz), 9020-0522/9120-0522 (MK 720 400V/50Hz), 9020-0524/9120-0524 (MK 720 480V/60Hz), 9020-0527/9120-0527 (MK 1020 400V/50Hz), 9020-0529/9120-0529 (MK 1020 480V/60Hz). Options include RS485 interface, zero-voltage switching outputs, analog temperature outputs, flexible Pt 100 object temperature sensor, over/under temperature safety device class 2, and APT-COM 4 software.
Section-by-Section Technical Breakdown
Describes the chamber layout and operating elements: main power switch, signal triangle, ergonomically adjustable control terminal, lateral control panel and rear chamber view. The type plate is located on the left side of the chamber, bottom right-hand, above the refrigerating module.
- Main power switch, signal triangle, door handle, chamber door and refrigerating machine maintenance access are identified in the chamber overview.
- Control terminal is ergonomically adjustable; a lateral control panel is provided.
- Type plate is on the left side, bottom right-hand, above the refrigerating module.
- Safety labels include hot surface and cold surface on the chamber door and a label requiring operation only with residual current circuit breakers type B.
Covers installation location and ambient conditions, minimum distances, ventilation, and electrical connection. The chamber must not be installed in unventilated recesses or damp areas, and sufficient room volume must be calculated according to EN 378-1:2016+A1:2020.
- Observe prescribed minimum distances when installing the chamber (chap. 3.4).
- Ensure sufficient ventilation for heat dispersal; do not install in unventilated recesses.
- Do not install or operate in hazardous locations or potentially explosive areas.
- Set up the chamber in a splash-proof manner; do not install in damp areas or puddles.
- Electrical connection per chap. 4.1; operation only with residual current circuit breakers type B.
Describes the MB2 chamber controller: normal display operating functions, display views (normal, program, chart-recorder), controller icons, operating modes, menu structure including Main, Settings and Service submenus, entry principles, and behavior during power failures and when opening the door.
- Display views: normal display, program display, chart-recorder display.
- Operating modes include Fixed Value, timer program, time programs and week programs.
- Menu structure: Main menu, Settings submenu, Service submenu.
- Controller behavior during and after power failures and when opening the door is defined.
Covers turning on the chamber and controller settings upon start up.
- Turning on the chamber via main power switch.
- Controller settings upon start up are described in chap. 6.2.
Describes set-point entry through the Setpoints menu and direct setpoint entry via the Normal display, plus special controller functions.
- Set-point entry through the Setpoints menu.
- Direct setpoint entry via Normal display.
- Special controller functions available in Fixed Value mode.
Covers starting a timer program, performance during program delay time, cancelling a running timer program, and performance after the end of the program.
- Starting a timer program.
- Performance during program delay time.
- Cancelling a running timer program.
- Performance after the end of the program.
Covers starting, cancelling and creating time programs, program editor and section editor management, value entry for program sections including section duration, set-point ramp and step, tolerance range, repetition of sections, and saving programs.
- Program editor for program management and section editor for section management.
- Value entry includes section duration, set-point ramp and set-point step, tolerance range.
- Sections can be repeated within a time program.
- Performance during program delay time and after program end are defined.
Covers starting, cancelling and creating week programs, program and section editors, and value entry including set-point ramp/step modes, weekday, start time, setpoint and special controller functions.
- Program editor and section editor for week programs.
- Value entry includes weekday, start time, setpoint and special controller functions.
- Set-point ramp and set-point step modes are configurable.
Provides an overview of notifications and alarm messages, the state of alarm, resetting an alarm and the list of active alarms, and activating/deactivating the audible alarm buzzer.
- Notifications and alarm messages overview.
- State of alarm and resetting an alarm.
- List of active alarms.
- Activating/deactivating the audible alarm (alarm buzzer).
Describes the over temperature protective device (class 1), the overtemperature safety controller class 2 including safety controller modes, setting, messages and measures in the state of alarm, and function check, plus the optional over/under temperature safety device class 2.
- Over temperature protective device class 1.
- Overtemperature safety controller class 2: modes, setting, alarm message and measures, function check.
- Optional over/under temperature safety device class 2.
Covers authorization levels and password protection, log in/out, user change, password assignment and change, deleting passwords, new password assignment when the password function was deactivated, and activation codes.
- Authorization levels with password protection.
- Log in, log out and user change procedures.
- Password assignment, change and deletion per authorization level.
- Activation code handling.
Covers menu language selection, date and time setting, temperature unit selection, display configuration including display parameters and touchscreen calibration, network and communication (serial interfaces, Ethernet configuration and MAC address display, e-mail), USB data transfer, and automatic interior lighting turn-off.
- Select controller menu language, date/time and temperature unit.
- Display configuration and touchscreen calibration.
- Network and communication: serial interfaces, Ethernet configuration, MAC address display, e-mail.
- USB menu for data transfer via USB interface.
- Automatic turn-off of interior lighting.
Covers the service contact page, current operating parameters, event list and technical chamber information.
- Service contact page.
- Current operating parameters display.
- Event list.
- Technical chamber information.
Describes chart recorder views, showing/hiding the legend, switching legend pages, showing/hiding specific indications, history display, and setting the parameters.
- Views and legend handling.
- Show/hide specific indications.
- History display.
- Setting chart recorder parameters.
Provides notes on refrigerating operation of the chamber.
- Operating notes specific to the refrigeration system.
Describes anti-condensation protection for the chamber.
- Anti-condensation protection function.
Covers APT-COM 4 Multi Management Software (accessory) including Basic Edition, RS485 interface option, zero-voltage switching outputs option, analog outputs for temperature option, and object temperature display with flexible Pt 100 temperature sensor option.
- APT-COM 4 Multi Management Software and Basic Edition.
- RS485 interface option.
- Zero-voltage switching outputs option.
- Analog outputs for temperature option.
- Object temperature display with flexible Pt 100 temperature sensor option.
Covers cleaning and decontamination / chemical disinfection of the chamber.
- Cleaning procedures.
- Decontamination / chemical disinfection.
Covers general information and personnel qualification, maintenance intervals and service, simple troubleshooting, and sending the chamber back to BINDER GmbH. Annual maintenance is required; electrical repeat inspection is required annually and before initial startup and after maintenance or repair.
- Maintenance at least once a year; refrigeration system work documented in a service log book.
- Repairs to the refrigeration circuit may currently only be carried out at BINDER.
- Annual repeat electrical inspection recommended under EN 50678/VDE 0701 and EN 50699/VDE 0702.
- Replace door gasket only when cold.
- With increased dust, clean condenser fan several times a year; check fan grid weekly.
- After 8760 operating hours or two years a service message appears and repeats every two weeks until reset by BINDER Service.
Covers factory calibration and adjustment, over-current protection, definition of usable volume, MK (E6) technical data, equipment and options, accessories and spare parts, heating-up and cooling-down graphs, heat compensation, and dimensions.
- Factory calibration and adjustment.
- Over-current protection.
- Definition of usable volume.
- MK (E6) technical data and dimensions.
- Heating-up and cooling-down graphs, heat compensation.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | S'applique à |
|---|---|---|
| MK 240 nominal temperature 180 °C (356 °F); nominal power 5.20 kW; nominal current 8.9 A; nominal voltage 400 V / 50 Hz ±10%; current type 3 N ~. p.11 ↗ | Electrical supply and over-current protection must be sized for 8.9 A at 400 V / 50 Hz for the MK 240 50 Hz variant. | MK 240 (E6) 400V/50Hz, Art. 9020-0513/9120-0513 |
| Refrigerant is R744 (CO2) with filling weight 1.40 kg; maximum operating pressure in the refrigerating system is 120 bar. p.11 ↗ | Refrigeration work must account for high-pressure CO2; repairs to the refrigeration circuit may currently only be carried out at BINDER. | MK (E6) series |
| IP protection 20 according to EN 60529; temperature safety device according to DIN 12880, class 2.0. p.11 ↗ | Chamber is protected against solid objects >12.5 mm but not water; must be set up splash-proof and not in damp areas. | MK (E6) series |
| Operation only with residual current circuit breakers type B. p.1 ↗ | The site electrical installation must provide a type B RCD for the chamber supply. | MK (E6) series |
| Required room volume must be calculated according to EN 378-1:2016+A1:2020 (chap. 3.4). p.1 ↗ | Installation room volume must be verified for refrigerant safety before commissioning. | MK (E6) series |
| Prescribed minimum distances must be observed when installing the chamber (chap. 3.4); do not install in unventilated recesses. p.1 ↗ | Maintain clearance for heat dispersal to avoid overheating damage. | MK (E6) series |
| Annual repeat electrical inspection required, plus test prior to initial startup and prior to resuming operation after maintenance or repair; recommended under EN 50678/VDE 0701 and EN 50699/VDE 0702. p.21 ↗ | Plan periodic electrical safety testing in the maintenance schedule. | MK (E6) series |
| Maintenance work at least once a year; all work on the refrigeration system must be documented in a service log book. p.21 ↗ | Establish an annual service interval and equipment records. | MK (E6) series |
| Replace the door gasket only when cold. p.21 ↗ | Schedule door gasket replacement with the chamber at ambient/cold condition to avoid damage. | MK (E6) series |
| With increased dust in ambient air, clean the condenser fan several times a year; check the fan grid behind the left maintenance access flap every week and clean by suction if dirty. p.21 ↗ | Dusty environments require more frequent condenser fan maintenance. | MK (E6) series |
| After 8760 operating hours or two years a service message appears; after confirmation it reappears every two weeks until reset by BINDER Service. p.21 ↗ | Service message is a time-based reminder, not a fault; only BINDER Service can reset it. | MK (E6) series |
| Repairs to the refrigeration circuit may currently only be carried out at BINDER. p.21 ↗ | On-site refrigeration repair is not permitted; arrange return to BINDER for refrigeration circuit work. | MK (E6) series |
| Maintenance work at the refrigeration system must only be conducted by qualified personnel trained according to EN 13313:2010 (e.g. refrigeration technician with certified expert knowledge acc. to Regulation (EU) 303/2008). p.21 ↗ | Verify service personnel certification before refrigeration maintenance. | MK (E6) series |
| The chamber must NOT become wet during operation, cleaning, or maintenance; do not remove the rear panel. p.21 ↗ | Cleaning and maintenance procedures must avoid water ingress; rear panel removal is prohibited. | MK (E6) series |
| Chambers were produced in accordance with VDE regulations and routinely tested according to VDE 0411-1 (IEC 61010-1). p.1 ↗ | Electrical safety basis for the chamber is IEC 61010-1. | MK (E6) series |
| The temperature inside the chamber must lie below the flash point or below the sublimation point of the charging material; no explosive mixture with air must form. p.1 ↗ | Verify charging material properties before setting operating temperature. | MK (E6) series |
| During and shortly after operation, inner surfaces, window, access ports and inner chamber become hot and almost equal the set-point. p.1 ↗ | Provide burn protection and allow cool-down before touching inner surfaces. | MK (E6) series |
| Do not load the lower housing cover with heavy objects while the chamber door is open and do not climb on it. p.1 ↗ | Avoid tipping/breakaway hazard of the protruding lower housing cover. | MK (E6) series |
Faults & Troubleshooting
| Code / Symptôme | Possible Cause | Remedy | Note |
|---|---|---|---|
| Service message after 8760 operating hours or two years p.108 ↗ | Time-based service interval reached (8760 operating hours or two years). | Confirm with the Confirm icon; message reappears every two weeks until reset by BINDER Service. | Chap. 21.2 |