Induztr
Résumé technique Visionneuse de documents Modèles associés

Data Sheet Model MKF 240

Technical Reference Summary

Covered Models & Document Scope

Cámara climática dinámica modelo MKF 240 para cambios rápidos de temperatura con control de humedad. Dos variantes: MKF240-400V (estándar, 400 V 3~/N/PE 50 Hz, n.º de art. 9020-0380) y MKF240-480V-C (con convertidor de tensión y frecuencia, 480 V 3~ 60 Hz, n.º de art. 9020-0358).

Technical Data

Parameter Value Condition
Temperature range -40...180 °C p.3 ↗ MKF240-400V, 50 Hz
Temperature range -40...180 °C p.3 ↗ MKF240-480V-C, 60 Hz
Temperature uniformity dependent on set 0.1...1.5 ± K p.3 ↗ MKF240-400V, 50 Hz
Temperature fluctuation dependent on set 0.1...0.5 ± K p.3 ↗ MKF240-400V, 50 Hz
Cooling down time from 180°C to -40°C 120 min p.3 ↗ MKF240-400V, 50 Hz
Temperature range with humidity 10...95 °C p.3 ↗ MKF240-400V, 50 Hz
Humidity range 10...98 % RH p.3 ↗ MKF240-400V, 50 Hz
Max. heat compensation at 25°C and 90% RH 400 W p.3 ↗ MKF240-400V, 50 Hz
Rated voltage 400 V p.3 ↗ MKF240-400V, 50 Hz
Rated voltage 480 V p.3 ↗ MKF240-480V-C, 60 Hz
Power frequency 50 Hz p.3 ↗ MKF240-400V
Power frequency 60 Hz p.3 ↗ MKF240-480V-C
Nominal power 6,8 kW p.3 ↗ MKF240-400V, 50 Hz
Unit fuse 16,0 A p.3 ↗ MKF240-400V, 50 Hz
Phase (nominal voltage) 3~/N/PE p.3 ↗ MKF240-400V, 50 Hz
Phase (nominal voltage) 3~ p.3 ↗ MKF240-480V-C, 60 Hz
Interior volume 228 L p.3 ↗ MKF240-400V, 50 Hz
Net weight of the unit (empty) 360 kg p.3 ↗ MKF240-400V, 50 Hz
Load per rack 30 kg p.3 ↗ MKF240-400V, 50 Hz
Permitted load 70 kg p.3 ↗ MKF240-400V, 50 Hz
Sound-pressure level 65 dB(A) p.4 ↗ MKF240-400V, 50 Hz
Sound-pressure level 67 dB(A) p.4 ↗ MKF240-480V-C, 60 Hz
Energy consumption at 25°C and 60% RH 1,500 Wh/h p.4 ↗ MKF240-400V, 50 Hz
Width net 1,115 mm p.4 ↗ MKF240-400V, 50 Hz
Height net 1,715 mm p.4 ↗ MKF240-400V, 50 Hz

Operating Limits

  • Temperature range: -40...180 °C p.3 ↗
  • Temperature range with humidity: 10...95 °C p.3 ↗
  • Humidity range: 10...98 % RH p.3 ↗
  • Max. heat compensation at 25°C and 90% RH: 400 W p.3 ↗
  • Permitted load: 70 kg p.3 ↗
  • Load per rack: 30 kg p.3 ↗

Type Code Decoder

Position Token Meaning
1-3 p.1 ↗ MKF Series: dynamic climate chamber for rapid temperature changes with humidity control
4-6 p.1 ↗ 240 Model size (interior volume class, 228 L)
suffix p.2 ↗ 400V Standard mains variant: 400 V 3~/N/PE 50 Hz
suffix p.2 ↗ 480V-C Mains variant with voltage and frequency converter: 480 V 3~ 60 Hz

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
MKF240-400V p.2 ↗ 9020-0380 Standard, 400 V 3~/N/PE 50 Hz Standard model version
MKF240-480V-C p.2 ↗ 9020-0358 with voltage and frequency converter, 480 V 3~ 60 Hz Option model with voltage and frequency converter

Modèles liés