UF V (E3) 09 2026

Technical Reference Summary

Covered Models & Document Scope

Covers UF V (E3) series ultra-low temperature freezers with RD4 controller and HIT insulation: UF V 350, UF V 500, UF V 700 in 230 V versions; UF V 350-UL, UF V 500-UL, UF V 700-UL in 120 V and 208-240 V versions; and water-cooled UF V 500-230V-H2O and UF V 700-230V-H2O. Options include door access system (UF V 500/700), battery-buffered alarm system, CO2 emergency cooling (UF V 500/700), advanced voltage booster, analog output, and lockable door handle.

Section-by-Section Technical Breakdown

pp. 19-28
Chamber Description and Door/Controller Housing

Describes the chamber overview, door lock and controller housing, main power switch, chamber rear, outer and compartment doors, and the optional lockable door handle. Also covers the optional door access system NumPad and electromechanical locking for UF V 500 and UF V 700.

  • Door access system option available for UF V 500 and UF V 700
  • Compartment doors and outer door described separately
  • Drain well for condensate during defrosting is an accessory
pp. 34-42
Installation and Connections

Covers spacers for rear wall distance, support feet (UF V 350 only), adjustable shelves, cooling water connections for water-cooled models, electrical connection, and the optional advanced voltage booster. Includes connection kit and hose burst protection safety kit for water cooling.

  • Support feet apply to UF V 350 only
  • Water cooling requires inlet/outlet connections and optional connection kit
  • Hose burst protection with reflux protection available via BINDER Customized Solutions
  • Advanced voltage booster is an option
pp. 43-44
Functional Overview of the RD4 Chamber Controller

Introduces the RD4 controller menu structure and access levels. Provides the basis for set-point entry, special functions, password protection, safety controller, general settings, tolerance/alarm settings, and alarm functions.

  • Menu structure and access levels described
  • Controller supports password-protected user and admin levels
pp. 45
Start Up

Describes preset factory parameters and chamber behavior after turning on. Provides the baseline for commissioning and set-point entry.

  • Preset factory parameters are defined
  • Behavior after power-on is described
pp. 46
Temperature Set-Point Entry

Explains how to enter the temperature set-point on the RD4 controller. This is the primary operating adjustment for the freezer.

  • Set-point entry via RD4 controller
  • Applies to all UF V models
pp. 48
Setting Special Controller Functions

Covers special controller functions available on the RD4 controller beyond basic set-point entry. Includes configuration of special operating modes.

  • Special functions configured via controller menu
  • Access may require password
pp. 49-51
Password Protection

Describes password request, assignment and modification of User and Admin passwords, and performance during and after power failure and shut down.

  • Separate User and Admin passwords
  • Password request behavior described
  • Power failure and shutdown behavior covered
pp. 51-53
Safety Controller (Temperature Safety Device)

Covers setting the safety controller mode and value, messages and measures in alarm state, and function check. The safety controller is a temperature safety device according to DIN 12880.

  • Safety controller mode and value settable
  • Alarm state messages and measures defined
  • Function check procedure provided
pp. 54-57
General Controller Settings

Covers menu language selection, temperature unit, date and time setting, language selection at restart, chamber address, and display brightness.

  • Language and temperature unit selectable
  • Date and time settable
  • Chamber address configurable
  • Display brightness adjustable
pp. 58-59
Tolerance Range and Alarm Delay Settings

Covers delay time for door open alarm, delay time for tolerance range alarm, temperature tolerance range, and alarm delay for continuous operating alarm.

  • Door open alarm delay configurable
  • Tolerance range alarm delay configurable
  • Temperature tolerance range settable
  • Continuous operating alarm delay settable
pp. 60-68
Alarm Functions

Describes alarm messages, information messages, audible alarm activation/deactivation, required actions for safety controller temperature alarm, tolerance range alarm, door open alarm, power failure alarm, battery management messages, temperature sensor failure messages, CO2 emergency cooling messages, and the zero-voltage relay alarm output.

  • Alarm and information messages distinguished
  • Audible alarm can be activated/deactivated
  • Specific actions defined for each alarm type
  • Zero-voltage relay alarm output available
pp. 69-74
Ethernet Network Settings

Covers showing network settings (MAC address, IP address, subnet mask, standard gateway, DNS server address, DNS chamber name) and changing network configuration (automatic/manual IP and DNS assignment).

  • MAC address displayable
  • IP address, subnet mask, gateway, DNS configurable
  • Automatic or manual IP assignment selectable
pp. 74-75
Access Codes (Door Access System Option)

Covers assigning access codes and opening the chamber door with an access code for the optional door access system on UF V 500 and UF V 700.

  • Access codes assignable
  • Door can be opened with access code
  • Option for UF V 500 and UF V 700
pp. 76-77
Data Recorder

Describes recorded data, storage capacity, setting the storage rate for the DL1 recorder data, and deleting the data recorder.

  • DL1 recorder data storage rate settable
  • Storage capacity defined
  • Data recorder can be deleted
pp. 78-80
USB-Menu: Data Transfer via USB Interface

Covers connecting a USB stick, import and export functions, ongoing data transfer, error during data transmission, and removing the USB stick.

  • USB stick connection procedure
  • Import and export functions available
  • Error handling during data transmission described
pp. 80-81
Battery Management System (Battery-Buffered Alarm System Option)

Covers battery operation and charging voltage for the optional battery-buffered alarm system.

  • Battery operation described
  • Charging voltage specified
pp. 81-82
Setting and Activating the Service Setpoint

Covers setting and activating the service setpoint on the RD4 controller. Intended for service purposes.

  • Service setpoint settable
  • Service setpoint activation described
pp. 83-89
CO2 Emergency Cooling (Option for UF V 500, UF V 700)

Covers connecting and exchanging the pressurized CO2 cylinder, operating the CO2 emergency cooling system, and controller settings including temperature setpoint, activation, and test run.

  • CO2 cylinder connection and exchange procedure
  • CO2 emergency cooling temperature setpoint settable
  • Activation and test run described
  • Option for UF V 500 and UF V 700
pp. 90
Data Monitoring and Recording

Covers the Ethernet interface, APT-COM 4 Multi Management Software (accessory), and analog output for temperature (option).

  • Ethernet interface for monitoring
  • APT-COM 4 software available as accessory
  • Analog temperature output optional
pp. 91
Chamber Inventory: Storage Rack Systems and Cryo Boxes (Option)

Covers storage rack systems with or without cryo boxes and cryo boxes as optional inventory equipment.

  • Storage rack systems optional
  • Cryo boxes optional
pp. 92-94
Cleaning and Decontamination

Covers cleaning procedures and decontamination/chemical disinfection of the chamber.

  • Cleaning procedure described
  • Chemical disinfection procedure described
pp. 95-101
Maintenance and Service, Troubleshooting, Repair, Testing

Covers general information and personnel qualification, customer maintenance work (condenser air filter, condenser cleaning, de-icing/defrosting, door lock maintenance), simple troubleshooting, maintenance intervals and service, service reminder, and returning the chamber to BINDER.

  • Condenser air filter checking/cleaning/replacing
  • Condenser cleaning procedure
  • De-icing and defrosting procedure
  • Door lock maintenance at least once a year with medical vaseline
  • Simple troubleshooting table provided
  • Annual electrical repeat inspection recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702
pp. 106-114
Technical Description

Covers factory calibration and adjustment, over current protection, technical data, equipment/options/accessories, optional equipment and spare parts, and dimensions for UF V 350, UF V 500, and UF V 700.

  • Technical data table
  • Over current protection specified
  • Dimensions for each model provided
  • Options and accessories listed

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication S'applique à
Nominal temperature is -90 °C / -130 °F for all UF V models. p.10 ↗ Defines the rated operating temperature for selection and validation. Whole series
UF V 500 standard 230 V model: nominal power 1.60 kW, nominal current 7.0 A, 230 V / 50 Hz, 1 N~. p.10 ↗ Electrical supply must be sized for at least 7.0 A at 230 V. UF V 500-230V
UF V 500-UL 208-230 V model: nominal power 1.80 kW, nominal current 8.2 A, 208-230 V / 60 Hz, 2~. p.11 ↗ UL 208-230 V variants require a 2-phase supply rated for 8.2 A. UF V 500-UL (208-240 V)
Maximum operating pressure of the refrigerating system is 28 bar. p.10 ↗ Service and connection work must respect this pressure limit. Whole series
Refrigerant stage 1: R290 0.150 kg; stage 2: R170 0.150 kg; contains hydrocarbon gases. p.10 ↗ Flammable refrigerants require appropriate ventilation and no ignition sources. Whole series
IP protection is IP 20 according to EN 60529. p.11 ↗ Chamber must not be installed in damp areas or exposed to water. Whole series
Temperature safety device according to DIN 12880, class 3.1. p.10 ↗ Safety controller classification per DIN 12880 must be considered for application risk assessment. Whole series
Nominal voltage tolerance is +/- 10% at the indicated power frequency. p.11 ↗ Supply voltage must remain within +/-10% of nominal for safe operation. Whole series
Door handle, door hinges and locking counterpart must be cleaned at least once a year and lubricated with medical vaseline. p.98 ↗ Annual maintenance schedule must include door lock lubrication. Whole series
Annual repeat electrical inspection and test prior to initial startup and after maintenance/repair are required; recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702. p.95 ↗ Electrical safety testing must be planned annually and after service. Whole series
Support feet apply to UF V 350 only. p.34 ↗ Installation leveling for UF V 350 uses support feet; other models differ. UF V 350
Water-cooled models UF V 500-230V-H2O and UF V 700-230V-H2O require cooling water inlet and outlet connections; connection kit and hose burst protection with reflux protection are available. p.37 ↗ Water cooling installation must include inlet/outlet plumbing and optional safety kit. UF V 500-230V-H2O, UF V 700-230V-H2O
CO2 emergency cooling is an option for UF V 500 and UF V 700 only. p.83 ↗ CO2 emergency cooling cannot be specified for UF V 350. UF V 500, UF V 700
Door access system option is available for UF V 500 and UF V 700. p.24 ↗ Access control option limited to larger models. UF V 500, UF V 700
Chambers with optional CO2 emergency cooling do not bear the GS mark; chambers with optional door access system do not bear the GS mark; CO2-equipped chambers do not bear the cUL test mark. p.12 ↗ Certification marks are affected by certain options and must be considered for compliance. Whole series with options
Advanced voltage booster is an option. p.42 ↗ Voltage booster can be specified for unstable supply conditions. Whole series
Analog output for temperature is an option. p.90 ↗ External monitoring can be implemented via optional analog output. Whole series
Data recorder DL1 storage rate is settable; storage capacity is defined. p.76 ↗ Data logging duration depends on configured storage rate. Whole series
Ethernet interface supports MAC address, IP address, subnet mask, gateway, and DNS configuration, with automatic or manual assignment. p.69 ↗ Network integration can be configured for remote monitoring. Whole series
Zero-voltage relay alarm output is provided. p.68 ↗ Alarm can be wired to external monitoring systems via volt-free contact. Whole series

Faults & Troubleshooting

Code / Symptôme Possible Cause Remedy Note
Chamber without function. p.98 ↗ No power supply. Check connection to power supply. 25.3
Chamber without function. p.98 ↗ Wrong voltage. Check power supply for correct voltage (chap.4.6). 25.3
Chamber without function. p.98 ↗ Chamber fuse has responded. Check chamber fuse and replace it if appropriate. If it responds again, contact BINDER service. 25.3
Chamber without function. p.98 ↗ Controller defective. Contact BINDER service. 25.3
Alarm message “Door open” p.98 ↗ Chamber door open. Close chamber door. 25.3
No refrigerating performance after turning on the chamber. p.98 ↗ Alarm message “Safety controller” Check setting of temperature set-point and of safety controller. If appropriate, select suitable limit value. 25.3
No refrigerating performance after turning on the chamber. p.98 ↗ Limit temperature reached. Check setting of temperature set-point and of safety controller. If appropriate, select suitable limit value. 25.3
No refrigerating performance after turning on the chamber. p.98 ↗ Controller defective. Contact BINDER service. 25.3
No refrigerating performance after turning on the chamber. p.98 ↗ Safety controller defective. Contact BINDER service. 25.3
No refrigerating performance after turning on the chamber. p.98 ↗ Semi-conductor relay defective. Contact BINDER service. 25.3
No refrigerating performance after turning on the chamber. p.98 ↗ Too much external heat load. Reduce heat load. 25.3