Induztr
Résumé technique Visionneuse de documents Modèles associés

Data Sheet Model UF V 500

Technical Reference Summary

Covered Models & Document Scope

Familia de congeladores de temperatura ultrabaja BINDER UF V 500: UFV500-230V (estándar, 230 V 1~ 50 Hz), UFV500-230V-W (refrigeración por agua, 230 V 1~ 50 Hz), UFV500UL-120V (115 V 1~ 60 Hz), UFV500UL-240V (208...230 V 2~ 60 Hz); volumen interior de 501 L

Technical Data

Parameter Value Condition
Temperature range -90...-40 °C p.2 ↗ UFV500-230V, 50 Hz
Temperature uniformity at -80°C 2.5 ± K p.2 ↗ UFV500-230V, 50 Hz
Temperature fluctuation at -80°C 1.5 ± K p.2 ↗ UFV500-230V, 50 Hz
Cooling down time from 22°C to -80°C 360 min p.2 ↗ UFV500-230V, 50 Hz
Rated voltage 230 V p.2 ↗ UFV500-230V, 50 Hz
Power frequency 50 Hz p.2 ↗ UFV500-230V
Nominal power 1,6 kW p.2 ↗ UFV500-230V, 50 Hz
Unit fuse 10,0 A p.2 ↗ UFV500-230V, 50 Hz
Phase (nominal voltage) 1~ p.2 ↗ UFV500-230V, 50 Hz
Interior volume 501 L p.2 ↗ UFV500-230V, 50 Hz
Net weight of the unit (empty) 247 kg p.2 ↗ UFV500-230V, 50 Hz
Permitted load 200 kg p.2 ↗ UFV500-230V, 50 Hz
Wall clearance back 100 mm p.2 ↗ UFV500-230V, 50 Hz
Wall clearance sidewise 250 mm p.2 ↗ UFV500-230V, 50 Hz
Width net 826 mm p.2 ↗ UFV500-230V, 50 Hz
Height net 1,966 mm p.2 ↗ UFV500-230V, 50 Hz
Depth net 938 mm p.2 ↗ UFV500-230V, 50 Hz
Interior width 606 mm p.2 ↗ UFV500-230V, 50 Hz
Interior height 1,300 mm p.2 ↗ UFV500-230V, 50 Hz
Interior depth 636 mm p.3 ↗ UFV500-230V, 50 Hz
Inner doors 2 p.3 ↗ UFV500-230V, 50 Hz
Unit doors 1 p.3 ↗ UFV500-230V, 50 Hz
Sound-pressure level 47 dB(A) p.3 ↗ UFV500-230V, 50 Hz
Number of shelves (std./max.) 3/11 p.3 ↗ UFV500-230V, 50 Hz
Cryoboxes, 50 mm 352 p.3 ↗ UFV500-230V, 50 Hz

Operating Limits

  • Temperature range: -90...-40 °C p.2 ↗
  • Permitted load: 200 kg p.2 ↗
  • Wall clearance back: 100 mm p.2 ↗
  • Wall clearance sidewise: 250 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-6 p.1 ↗ UFV500 Model family UF V 500 ultra-low temperature freezer, 501 L
suffix p.1 ↗ UL UL/NEMA variant for 60 Hz markets (absent = European 230 V 50 Hz version)
suffix p.1 ↗ -W Water cooling option model (absent = standard air-cooled)
voltage field p.1 ↗ 230V 230 V 1~ 50 Hz, CEE 7/7 plug, 10,0 A fuse
voltage field p.1 ↗ 120V 115 V 1~ 60 Hz, NEMA 5-15 plug, 15,0 A fuse
voltage field p.1 ↗ 240V 208...230 V 2~ 60 Hz, NEMA 6-15 plug, 10,0 A fuse

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
UFV500-230V p.1 ↗ 9020-0347 230 V 1~ 50 Hz, standard CEE 7/7 plug, 10,0 A fuse
UFV500-230V-W p.1 ↗ 9020-0349 230 V 1~ 50 Hz, water cooling CEE 7/7 plug, 10,0 A fuse
UFV500UL-120V p.1 ↗ 9020-0351 115 V 1~ 60 Hz, standard NEMA 5-15 plug, 15,0 A fuse
UFV500UL-240V p.1 ↗ 9020-0353 208...230 V 2~ 60 Hz, standard NEMA 6-15 plug, 10,0 A fuse

Modèles liés