Data Sheet Model WIC 1

Technical Reference Summary

Covered Models & Document Scope

Modèle WIC 1, chambre de test de stabilité à accès frontal selon ICH (Q1A) ; version standard WIC1-230V, Art.-No. 9800-0011, 200...230 V 1~ 50 Hz

Technical Data

Parameter Value Condition
Interior volume 12,530 L p.1 ↗ WIC 1
Rated voltage 200...230 V p.2 ↗ WIC 1
Power frequency 50 Hz p.2 ↗ WIC 1
Phase (nominal voltage) 1~ p.2 ↗ WIC 1
Nominal power 2,7 kW p.2 ↗ WIC 1
Unit fuse 16,0 A p.2 ↗ WIC 1
Temperature range with humidity 5...50 °C p.2 ↗ WIC 1
Humidity range 20...90 % RH p.2 ↗ WIC 1
Temperature uniformity at 25°C and 60% RH 1.5 ± K p.2 ↗ WIC 1
Temperature uniformity at 40°C and 75% RH 1.5 ± K p.2 ↗ WIC 1
Temperature fluctuation at 25°C and 60% RH 0.5 ± K p.2 ↗ WIC 1
Temperature fluctuation at 40°C and 75% RH 0.5 ± K p.2 ↗ WIC 1
Humidity fluctuation at 25°C and 60% RH 2.5 ± % RH p.2 ↗ WIC 1
Humidity fluctuation at 40°C and 75% RH 2.5 ± % RH p.2 ↗ WIC 1
Temperature accuracy ± 1.5 °C p.1 ↗ WIC 1
Relative humidity accuracy ± 2.5 % p.1 ↗ WIC 1
U value of room panels 0,21 W/m2K p.1 ↗ WIC 1
Surface load 50.000 N/m2 p.1 ↗ WIC 1
Refrigerant GWP value 547 p.1 ↗ WIC 1

Operating Limits

  • Temperature range with humidity: 5...50 °C p.2 ↗
  • Humidity range: 20...90 % RH p.2 ↗
  • Unit fuse: 16,0 A p.2 ↗
  • Surface load: 50.000 N/m2 p.1 ↗

Type Code Decoder

Position Token Meaning
1-4 p.1 ↗ WIC1 Model series WIC 1 walk-in chamber
6-8 p.1 ↗ 230 Rated voltage 200...230 V

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
WIC1-230V p.1 ↗ 9800-0011 Standard, 200...230 V 1~ 50 Hz Walk-in chamber for stability testing according to ICH (Q1A)