Induztr
Résumé technique Visionneuse de documents Modèles associés

Agilent Cary 620 FTIR Microscope User's Guide

Sommaire

  1. Warning pp. 7-18
    Safety warnings for UV, electrical, laser, and other hazards
  2. Overview pp. 19-24
    System overview, environmental requirements, purge gas, and training
  3. Procedure pp. 25-36
    Software and hardware installation, cable connections, and setup steps
  4. Procedure pp. 37-42
    Turning on system, filling Dewars, performing collects, and shutdown
  5. Procedure pp. 43-46
    Routine maintenance procedures and warnings
  6. pp. 47-49
    Troubleshooting guidance for common issues
  7. p. 22 table, 1 rows
  8. p. 7 table, 12 rows
  9. p. 21 table, 2 rows

Table 2. Cary 620 FTIR microscope weights and dimensions

Width (cm, in)Depth (cm, in)Height (cm, in)Weight (kg, lb)
39.6, 15.668.9, 26.366.8, 26.350, 110

Table

РазделСтраница
General8
Verifying Safe State8
Ultraviolet radiation9
Cryogenic cooling9
Compressed gas cylinders9
Electrical hazards10
Laser safety11
Other precautions13
Warnings or symbols14
Color coding15
CE compliance15
Electromagnetic compatibility15

Оглавление раздела 1. Safety Practices and Hazards; таблица восстановлена как напечатано (без транспонирования).

Table 1. Fuse specifications

InstrumentFuse type
Agilent Cary 600 Series FTIR spectrometerT4 AH 250 V, IEC601272-2 Sheet 5, 5x20 mm, Littelfuse 0215004 or equivalent
Agilent Cary 610/620 FTIR microscope (typical)1FS1: T2.0 AH 250 V, IEC601272-1 Sheet 5, 5 x 20 mm, Littelfuse 0215020 or equivalent

Engineering Facts

14 facts · every claim linked to a page
Safety

The instrument contains a Class 2 helium-neon laser at 632.8 nm, so laser safety precautions and interlocks must be respected during operation and service.

Laser classification drives site safety signage, eyewear assessment, and prohibition of bypassing interlocks.

Source
“a helium-neon Class 2 laser operating in the visible region at 632.8 nanometers”
Safety

Liquid nitrogen used for detector cooling is at -196 °C and requires appropriate protective equipment.

Cryogenic handling PPE and fill procedures must be planned before installation and operation.

Source
“The liquid nitrogen used in this process is extremely cold (-196 °C, -320.8 °F) and can cause damage to the human body.”
Installation

The microscope requires a dedicated clean 10 A grounded circuit with no other instruments on it.

Electrical infrastructure must be provisioned before the instrument arrives.

Source
“Dedicated, clean 10 A power circuit with grounded receptacles. No other instruments should be powered from the same circuit.”
Installation

Purge gas must be water-free air dried to -70 °C dew point or dry nitrogen, with flow monitored by a flow meter.

Purge supply quality and metering are installation prerequisites for optical performance.

Source
“Purge gas must be water-free air (dried to a dew point of -70 °C) or dry nitrogen.”
Installation

Purge flow must not exceed 10 L/min and requires a flow meter plus a regulator if source pressure exceeds 60 psi.

Over-pressure or unmonitored purge can damage optics or fail to protect the optical path.

Source
“Flow rate for purge: 10 L/min (20 ft3/hr) maximum. You must have a flow meter to monitor the flow of the purge gas.”
Installation

The bench must be level and rigid with a loaded flatness tolerance of 0.4 mm per 300 mm and at least 30 cm clearance behind the microscope.

Bench flatness and rear clearance affect alignment stability and service access.

Source
“Loaded flatness tolerance: 0.4 millimeters per 300 millimeters of length (1/64 inches per foot of length).”
Installation

The microscope is for indoor use only and is classified Pollution degree 2 and Installation Category II.

Site environment and overvoltage category must match the installation category for safe operation.

Source
“The Agilent Cary 620 FTIR Series microscope is suitable for indoor use only and is classified Pollution degree 2 and Installation Category II (EN 61010-1).”
Performance

Operating environment must be 20–26 °C with a maximum gradient of 1 °C/hr and 20–50% non-condensing RH.

Temperature and humidity control are required for stable FTIR measurements.

Source
“Temperature: 20 °C to 26 °C (68 °F to 80 °F)”
Performance

The system is tested for correct operation at sea level only; higher altitudes require Agilent advice.

Altitude affects cooling and purge performance, so site altitude must be confirmed before selection.

Source
“FTIR spectrometers and microscopes have been thoroughly tested for correct operation at sea level only.”
Operation

The FPA detector Dewar takes about 475 mL of liquid nitrogen initially, then about 200 mL top-up after 10–20 minutes equilibration.

LN2 consumption and fill procedure must be planned for routine operation.

Source
“This takes about 475 milliliters (16 ounces) of liquid nitrogen.”
Operation

Do not fill the Dewar to the top; overfilling can freeze the elastomer o-ring seal and cause loss of vacuum requiring detector return.

Improper LN2 filling can permanently damage the detector assembly.

Source
“Do not fill the Dewar to the top. This can freeze the elastomer o-ring Dewar seal causing a loss of vacuum.”
Operation

The microscope shutter should always be in the open position when using the microscope.

Shutter state directly affects whether spectra are collected.

Source
“The shutter should always be in the open position when using the microscope.”
Maintenance

Only maintenance procedures specifically mentioned may be performed by the operator; all others require Agilent-trained personnel.

Defines the boundary of user-serviceable maintenance and warranty-safe intervention.

Source
“Any maintenance procedures not specifically mentioned should be carried out only by Agilent-trained, Agilent-qualified or Agilent-authorized representatives.”
Conformité

The product is ITAR-regulated and requires a U.S. export license for export from the United States.

Export control affects procurement, shipment, and international deployment planning.

Source
“An export license from the U.S. government is therefore required to export this product from the United States”