Agilent NovoCyte Opteon Spectral Flow Cytometer Specifications

Technical Reference Summary

Covered Models & Document Scope

Citómetro de flujo espectral Agilent NovoCyte Opteon; configuraciones con cuatro láseres (349 nm 20 mW, 405 nm 130 mW, 488 nm 100 mW, 561 nm 100 mW, 637 nm 120 mW) y 8, 11, 14, 18 o 19 detectores; configuraciones personalizadas disponibles bajo pedido

Technical Data

Parameter Value Condition
Laser type Solid state laser with onboard thermal-electric cooling p.1 ↗
Laser beam configuration Spacially separated; fixed laser delay and no need for adjustment p.1 ↗
Photodetector Avalanche photodiode (APD) p.1 ↗
Optical alignment Fixed. No operator alignment required. p.1 ↗
Forward scatter detection laser 488 nm p.1 ↗ FSC
Side scatter detection laser BSSC 488 nm; VSSC 405 nm nm p.1 ↗ SSC
FSC sensitivity 0.4 µm p.1 ↗
VSSC sensitivity 0.08 µm p.1 ↗
BSSC sensitivity 0.2 µm p.1 ↗
FITC fluorescence threshold sensitivity ≤ 50 MESF p.1 ↗ B543 channel
PE fluorescence threshold sensitivity ≤ 10 MESF p.1 ↗ Y582 channel
APC fluorescence threshold sensitivity ≤ 10 MESF p.1 ↗ R661 channel
Pacific Blue fluorescence threshold sensitivity ≤ 30 MESF p.1 ↗ V455 channel
Fluorescence resolution < 3 % CV p.1 ↗ CEN
Flow cell dimensions 170 x 490 µm p.2 ↗ rectangular quartz flow cell
Sample acquisition rate 100,000 events/second p.2 ↗
Volumetric absolute count precision ≤ 5 % p.2 ↗
Sample flow rate 5 to 120 µL/min p.2 ↗ 1 µL increments
Sample acquisition volume 1 µL to 5 mL p.2 ↗
Sheath fluid container capacity 6 L p.2 ↗
Waste fluid container capacity 6 L p.2 ↗
Cleaning fluid container capacity 500 mL p.2 ↗
Decontamination fluid container capacity 500 mL p.2 ↗ automatic liquid volume sensing
Number of detectors 8, 11, 14, 18, 19 p.3 ↗ four-laser configuration

Operating Limits

  • Sample flow rate range: 5 to 120 µL/min p.2 ↗
  • Sample acquisition volume range: 1 µL to 5 mL p.2 ↗
  • Volumetric absolute count precision: ≤ 5% p.2 ↗