Covered Models & Document Scope

Gama de incubadores de CO2 CellXpert, centrada en el CellXpert C170i (con interfaz táctil VisioNize, control de O2/tri-gas, HTD de 180 °C) y el CellXpert C170 más económico (pantalla LED estándar, HTD de 140 °C, sin control de O2). Las tablas de accesorios/pedidos cubren configuraciones de puerta (1/4/8 segmentos, manilla derecha/izquierda), monitor de humedad, sensor de nivel de agua, O2 1–20 %, paquete de cobre y variantes de red 230 V 50/60 Hz (Europa, UK/HKG, Australia, China) y 100–120 V 50/60 Hz (EE. UU./Japón).

Technical Data

Parameter Value Condition
Display VisioNize touch interface p.16 ↗ CellXpert C170i
Display standard LED display p.16 ↗ CellXpert C170
High temperature disinfection (HTD) 180 °C p.16 ↗ CellXpert C170i, standard
High temperature disinfection (HTD) 140 °C p.16 ↗ CellXpert C170, standard
Water/Humidity monitoring optional p.16 ↗ CellXpert C170i
Water/Humidity monitoring – p.16 ↗ CellXpert C170
O2 range (tri-gas control) 1 – 20 % p.16 ↗ CellXpert C170i
O2 range (tri-gas control) – p.16 ↗ CellXpert C170
O2 control increment 0.1 % p.16 ↗ CellXpert C170i
O2 stability ±0.1 % p.16 ↗ CellXpert C170i
O2 accuracy ±0.25 % p.16 ↗ CellXpert C170i
O2 sensor platinum-coated ZrO2 (high temperature r...) p.16 ↗ CellXpert C170i
Number of shelves (included/max) 4/8 p.16 ↗ CellXpert C170i
Number of shelves (included/max) 3/8 p.16 ↗ CellXpert C170
Copper options chamber, water tray, shelves p.16 ↗ CellXpert C170i
Copper options water tray, shelves p.16 ↗ CellXpert C170
On-board data logging standard p.16 ↗ CellXpert C170i
On-board data logging – p.16 ↗ CellXpert C170
Data export interface USB, Ethernet, VisioNize onboard p.16 ↗ CellXpert C170i
Data export interface Ethernet, VisioNize ready (requires Visi...) p.16 ↗ CellXpert C170
Width (w) 71.8 cm (28.3 in) p.19 ↗ CellXpert C170i
Height (h) 90.0 cm (35.4 in) p.19 ↗ CellXpert C170i
Depth (d) 71.5 cm (28.1 in) p.19 ↗ CellXpert C170i
Height (h1) 194.6 cm (76.7 in) p.19 ↗ CellXpert C170i, with stacking stand
Height (h2) 107.3 cm (42.3 in) p.19 ↗ CellXpert C170i, with stacking stand

Operating Limits

  • High temperature disinfection (HTD), CellXpert C170i: 180 °C p.16 ↗
  • High temperature disinfection (HTD), CellXpert C170: 140 °C p.16 ↗
  • O2 range (tri-gas control), CellXpert C170i: 1 – 20 % p.16 ↗
  • O2 control increment, CellXpert C170i: 0.1 % p.16 ↗
  • O2 stability, CellXpert C170i: ±0.1 % p.16 ↗
  • O2 accuracy, CellXpert C170i: ±0.25 % p.16 ↗

Type Code Decoder

Position Token Meaning
field 1 p.18 ↗ 6731 CellXpert C170i accessory/configuration order-number prefix
field 2 p.18 ↗ 6734 CellXpert C170 accessory/configuration order-number prefix
field 3 p.18 ↗ 000 / 001 000 = without O2 control; 001 = with O2 1–20 % tri-gas control
field 4 p.18 ↗ 0 / 1 0 = without humidity monitor; 1 = with humidity monitor
field 5 p.18 ↗ 0 / 2 / 3 / 5 / 8 0 = no sensor; 2 = water level sensor; 3 = humidity monitor + water level sensor; 5 = humidity monitor (with O2); 8 = O2 with both sensors
field 6 p.18 ↗ 1 / 4 / 8 number of door segments
field 7 p.18 ↗ 1 / 2 / 3 / 4 / 5 mains/plug variant: 1 = European 230 V, 2 = UK/HKG 230 V, 3 = Australia 230 V, 4 = China 230 V, 5 = USA/Japan 100–120 V (all 50/60 Hz)
suffix p.18 ↗ * variant marked with asterisk in the ordering table

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
6731 000.011* p.18 ↗ 6731 000.011 1 door segment, right handle, European 230 V 50/60 Hz asterisk as printed
6731 000.511 p.18 ↗ 6731 000.511 1 door segment, right handle, with humidity monitor, European 230 V 50/60 Hz
6731 001.011* p.18 ↗ 6731 001.011 1 door segment, right handle, O2 1–20 %, European 230 V 50/60 Hz asterisk as printed
6731 001.511 p.18 ↗ 6731 001.511 1 door segment, right handle, O2 1–20 % with humidity monitor, European 230 V 50/60 Hz
6731 001.811 p.18 ↗ 6731 001.811 1 door segment, O2 1–20 % with humidity monitor, European 230 V 50/60 Hz
6731 000.021* p.18 ↗ 6731 000.021 1 door segment, left handle, European 230 V 50/60 Hz asterisk as printed
6731 000.521 p.18 ↗ 6731 000.521 1 door segment, left handle, with humidity monitor, European 230 V 50/60 Hz
6731 001.021* p.18 ↗ 6731 001.021 1 door segment, left handle, O2 1–20 %, European 230 V 50/60 Hz asterisk as printed
6731 001.521 p.18 ↗ 6731 001.521 1 door segment, left handle, O2 1–20 % with humidity monitor, European 230 V 50/60 Hz
6731 001.821 p.18 ↗ 6731 001.821 1 door segment, left handle, O2 1–20 % with both sensors, European 230 V 50/60 Hz