Induztr
Résumé technique Visionneuse de documents Modèles associés

Fermentors Bioreactors Brochure BioBLU Family Single Scalability

Technical Reference Summary

Covered Models & Document Scope

Familia de biorreactores de un solo uso BioBLU®: BioBLU 0.3c, 1c, 3c, 10c, 50c (cultivo celular, 'c'); BioBLU 0.3f, 1f, 3f (fermentación microbiana, 'f'); BioBLU 5p (lecho empaquetado, células adherentes); BioBLU 0.3sc (células madre); y variantes HNQ (0.3c, 1c, 3c, 5p, 10c, 50c, 1f, 3f) para fabricación comercial. Rango de volumen de trabajo de 65 mL a 40 L.

Technical Data

Parameter Value Condition
Working volume 100 – 250 mL p.10 ↗ BioBLU 0.3c
Working volume 320 mL – 1.25 L p.10 ↗ BioBLU 1c
Working volume 1.25 – 3.75 L p.10 ↗ BioBLU 3c
Working volume 3.3 – 10 L p.10 ↗ BioBLU 10c
Working volume 18 – 40 L p.10 ↗ BioBLU 50c
Working volume 3.75 L p.10 ↗ BioBLU 5p
Working volume (total) 380 mL p.10 ↗ BioBLU 0.3c
Working volume (total) 1.8 L p.10 ↗ BioBLU 1c
Working volume (total) 5 L p.10 ↗ BioBLU 3c
Working volume (total) 13.3 L p.10 ↗ BioBLU 10c
Working volume (total) 50 L p.10 ↗ BioBLU 50c
Working volume (total) 5 L p.10 ↗ BioBLU 5p
Working volume 65 – 250 mL (380 mL) p.11 ↗ BioBLU 0.3f
Working volume 250 mL – 1.25 L (1.8 L) p.11 ↗ BioBLU 1f
Working volume 1.25 L – 3.75 L (5 L) p.11 ↗ BioBLU 3f
Recommended agitation 20 – 500 rpm p.10 ↗ BioBLU 0.3c
Recommended agitation 30 – 600 rpm p.10 ↗ BioBLU 1c
Recommended agitation 25 – 200 rpm p.10 ↗ BioBLU 3c
Recommended agitation 25 – 200 rpm p.10 ↗ BioBLU 10c
Recommended agitation 25 – 150 rpm p.10 ↗ BioBLU 50c
Recommended agitation 25 – 200 rpm p.10 ↗ BioBLU 5p
Recommended agitation 20 – 2,000 rpm p.11 ↗ BioBLU 0.3f
Recommended agitation 100 – 1,500 rpm p.11 ↗ BioBLU 1f
Recommended agitation 25 – 1,200 rpm p.11 ↗ BioBLU 3f
Max. operating temperature 45 °C p.11 ↗ BioBLU 0.3f / 1f / 3f

Operating Limits

  • Working volume range (whole family): 65 mL to 40 L p.3 ↗
  • Max. operating temperature (BioBLU f): 45 °C p.11 ↗
  • Sterilization (BioBLU 0.3f/1f): Irradiated by > 25 kGy (X-ray) p.11 ↗
  • Sterilization (BioBLU 3f): Autoclavable, no pre-sterilization p.11 ↗
  • Gassing (BioBLU c): up to 2 vvm p.7 ↗
  • Gassing (BioBLU 0.3sc): up to 1.5 vvm p.7 ↗

Type Code Decoder

Position Token Meaning
1-6 p.10 ↗ BioBLU Product family: rigid-wall single-use bioreactor
7-10 p.10 ↗ 0.3c / 1c / 3c / 10c / 50c Nominal size and cell-culture ('c') variant
7-10 p.11 ↗ 0.3f / 1f / 3f Nominal size and microbial fermentation ('f') variant
7-10 p.10 ↗ 5p Packed-bed bioreactor for adherent cells (Fibra-Cel® disks)
7-10 p.10 ↗ 0.3sc Stem cell variant with 8-blade impeller
suffix p.12 ↗ HNQ Variant suitable for commercial manufacturing
order no. digit p.10 ↗ 13861xxxxx Eppendorf order number; digits encode size, sparger, impeller, pH sensor and pack quantity

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
BioBLU® 0.3c p.10 ↗ 1386101000 Open pipe sparger, 1x pitched blade, standard pH, 4-pack
BioBLU® 0.3c p.10 ↗ 1386101500 Open pipe sparger, 1x pitched blade, optical pH, 4-pack
BioBLU® 0.3sc p.10 ↗ 1386102000 Open pipe sparger, 1x 8-blade impeller, optical pH, 4-pack
BioBLU® 1c p.10 ↗ 1386111000 Open pipe sparger, 1x pitched blade, standard pH, 4-pack
BioBLU® 1c p.10 ↗ 1386112000 Open pipe sparger, 1x pitched blade, optical pH, 4-pack
BioBLU® 1c p.10 ↗ 1386111100 Open pipe sparger, 2x pitched blade, standard pH, 4-pack
BioBLU® 1c p.10 ↗ 1386112100 Open pipe sparger, 2x pitched blade, optical pH, 4-pack
BioBLU® 3c p.10 ↗ 1386124000 Microsparger, 1x pitched blade, optical pH, 1-pack
BioBLU® 3c p.10 ↗ 1386125000 Macrosparger, 1x pitched blade, optical pH, 1-pack
BioBLU® 3c p.10 ↗ 1386126000 Microsparger, 2x pitched blade, optical pH, 1-pack