Eukaryote Total RNA Nano Assay Qualitative range 5 to 500 ng/µL
RNA 6000 Nano LabChip Kit
mRNA Nano Assay* Qualitative range 25 to 250 ng/µL
RNA 6000 Pico LabChip Kit
Eukaryote Total RNA Pico Assay Qualitative range 50 to 5000 pg/µL in water
RNA 6000 Pico LabChip Kit
mRNA Pico Assay* Qualitative range 250 to 5000 pg/µL in water
* The mRNA assays are suitable for analysis of cRNA as well.
Table 31
Ozone level in your lab
Wash Procedure
Ozone-Barrier Slide Cover
< 5 ppb
Wash Procedure without Stabilization and Drying Solution. "Step 4. Wash the microarray slides" on page 44
No
> 5 ppb and < 10 ppb
Wash Procedure without Stabilization and Drying Solution. "Step 4. Wash the microarray slides" on page 44
Yes
> 10 ppb
Wash Procedure with Stabilization and Drying Solution. See "Step 1. Prepare the Stabilization and Drying Solution" on page 80 and "Step 2. Wash with Stabilization and Drying Solution" on page 81.
Yes
Wash procedure to follow
Related Microarray Reagents
Description
Vendor and part number
Universal Human Reference RNA
Agilent p/n 740000
Universal Mouse Reference RNA
Agilent p/n 740100
Universal Rat Reference RNA
Agilent p/n 740200
Table 9 Required Reagents
Description
Vendor and part or catalog number
Low Input Quick Amp Labeling Kit, Two-Color
Agilent p/n 5190-2306
RNA Spike-In Kit, Two-Color
Agilent p/n 5188-5279
Gene Expression Hybridization Kit
Agilent p/n 5188-5242
Gene Expression Wash Buffer Kit
Agilent p/n 5188-5327
DNase/RNase-free distilled water
Invitrogen p/n 10977-015
RNeasy Mini Kit
Qiagen p/n 74104
ethanol (95% to 100% molecular biology grade)
Sigma-Aldrich p/n E7023-6×500ML
Milli-Q water or equivalent
isopropyl alcohol (molecular biology grade)
Engineering Facts
14 facts · every claim linked to a page
Selection
The kit supports a total RNA input range of 10–200 ng or a minimum of 5 ng poly A+ RNA for two-color processing, which defines the minimum sample you can load.
Determines whether your available sample amount is compatible with this labeling kit before you commit to the workflow.
Source
“with a sample input RNA range between 10 ng and 200 ng of total RNA or a minimum of 5 ng of poly A+ RNA for two-color processing”
Performance
Impure template RNA (low A260/A230) degrades results and must be purified before use.
Sets a purity acceptance criterion for input RNA that must be checked before labeling.
Source
“RNA that is not pure, as measured by A260/A230 ratio, can lead to poor results and must be purified.”
Safety
The Stabilization and Drying Solution is toxic and flammable and must be handled in a fume hood with gloves and eye/face protection.
Dictates lab ventilation and PPE requirements for the drying step installation.
Source
“Stabilization and Drying Solution is toxic and flammable and must be used in a suitable fume hood.”
Safety
LiCl in the 2× Hi-RPM Hybridization Buffer is toxic and a potential teratogen, requiring suitable PPE.
Requires PPE planning when handling hybridization buffer.
Source
“LiCl is toxic and a potential teratogen.”
Operation
Spike A must always be labeled with Cy3 and Spike B with Cy5, and this orientation is hard-coded in the software.
Prevents dye-swap errors that would invert the LogRatio output and require manual correction.
Source
“Always label Spike A Mix with cyanine 3 and Spike B Mix with cyanine 5.”
Operation
Fragmentation must be stopped after exactly 30 minutes at 60°C by cooling and adding hybridization buffer.
Over-fragmentation degrades cRNA and compromises hybridization signal.
Source
“Do not incubate sample in the next step for more than 30 minutes.”
Operation
Do not vortex the hybridization mix because the surfactant forms bubbles; mix only by careful pipetting.
Bubbles in the hybridization mix cause array artifacts and uneven hybridization.
Source
“Do not mix on a vortex mixer; mixing on a vortex mixer introduces bubbles.”
Maintenance
Adding 0.005% Triton X-102 to wash buffers reduces array wash artifacts and is highly recommended.
Improves wash quality and reproducibility of array results.
Source
“The addition of 0.005% Triton X-102 (10%) to the Gene Expression wash buffers reduces the possibility of array wash artifacts.”
Operation
For uniform washing, disassemble no more than eight microarray slides per wash group.
Sets a batch-size limit that affects throughput planning and wash consistency.
Source
“For uniform washing, do up to a maximum of eight disassembly procedures yielding eight microarray slides.”
Operation
Do not scan G3 microarrays with HD format settings because resolution will be too low for analysis.
Prevents selecting the wrong scanner profile that would yield unusable image resolution.
Source
“Do not scan G3 microarrays with HD format settings.”
Selection
HD formats require 5 µm scan resolution while G3 formats require 3 µm.
Determines scanner resolution setting matched to the array format chosen.
Source
“Scan resolution”
Performance
Default Feature Extraction protocols may not be optimal for custom arrays lacking enough replicated or validated negative control probes.
Custom array designs may need protocol tuning for reliable data extraction.
Source
“If custom arrays without enough replicated probes are used, or arrays with custom probes designated as Negative Control probes are used, the default FE Protocols may not be optimal.”
Operation
Using a thermocycler instead of water baths can slightly lower cRNA yield.
Informs the trade-off when choosing thermocycler versus water bath incubation.
Source
“The use of a thermocycler can slightly lower the yield of cRNA when compared to the use of water baths.”
Operation
Avoid pipetting volumes below 2 µL to ensure accuracy when preparing spike mixes.
Sets a minimum pipetting volume to maintain dilution accuracy.
Source
“Avoid pipetting volumes less than 2 µL to ensure accuracy.”