The cooling water circuit needs a minimum differential pressure of 3 bar and must not exceed 6 bar, so the building water supply must be checked.
Insufficient or excessive cooling water pressure will prevent proper operation or damage the unit.
Source
“min. cooling water differential pressure 3 bar”
Installation
Cooling water intake temperature must be between 5 °C and 20 °C, and pipes must be insulated if water temperature can fall below the condensation point.
Defines the acceptable cooling water supply conditions and condensation protection measures.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops significantly at low temperatures, from 2 kW at 20 °C to only 0.13 kW at -90 °C, so process heat loads must be checked against the actual operating temperature.
Prevents oversizing expectations and ensures the unit can remove the process heat at the required low temperature.
Source
“at -90°C 0,13 kW”
Installation
The unit is water-cooled and uses refrigerants R-452A and R-508B with CO2 equivalents of 1.9 t and 5.8 t, which matters for refrigerant regulations and leak monitoring.
Affects compliance with F-gas regulations and the need for leak detection or reporting.
Source
“CO2 equivalent 2nd stage 5,8 t”
Operation
The pressure pump delivers up to 25 l/min at 0.7 bar and the suction pump up to 18.5 l/min at 0.4 bar, so external circuit pressure losses must be evaluated.
Determines whether the pump can overcome the pressure drop of the external application loop.
Source
“max. delivery 25 l/min”
Operation
The maximum permissible kinematic viscosity of the bath fluid is 50 mm²/s, so the heat transfer fluid must be selected within this limit.
Ensures the pump can circulate the fluid and that heat transfer remains adequate.
Source
“max. permissible kin. viscosity 50 mm²/s”
Installation
The bath requires a minimum filling capacity of 19 l and holds 26 l, with only 15 l when a displacement rack is used, so the fluid volume and expansion must be planned.
Prevents insufficient filling and overflow during temperature changes.
Source
“min. filling capacity 19 l”
Installation
The unit has IP20 protection and must be operated between 5 °C and 40 °C ambient temperature, so it cannot be placed in a wet or hot environment.
Restricts installation location regarding moisture and ambient temperature.
Source
“Degree of Protection IP20”
Installation
The overall dimensions are 605x706x1136 mm and the net weight is 175 kg, so floor space and load-bearing capacity must be verified.
Needed for site planning, access routes and floor loading.
Source
“Overall dimensions WxDxH ** 605x706x1136 mm”
Conformité
Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, but the combination of -5% voltage and +2% frequency is not allowed.
Defines the acceptable power quality window for reliable operation.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Three-phase devices with current consumption less than 63 A are delivered with cable but without plug, so a suitable plug must be provided on site.
Affects electrical installation scope and commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL, which is relevant for unattended operation and flammable fluids.
Confirms built-in safety functions required for protected operation.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation
The pressure equipment category is Art. 4.3 PED, which determines the applicable pressure vessel compliance requirements.
Relevant for compliance documentation and acceptance at site.