Selection
The unit is classified as safety class III / FL per DIN 12876, which determines the permitted flammable-liquid use and required protection concept.
Safety class governs whether the bath may be used with flammable media and what protective measures are mandatory.
Source
“Safety classification
III / FL”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, so the installation location must respect ignition-source and ventilation rules.
A3 flammability affects room classification, ventilation and prohibition of ignition sources near the unit.
Source
“R-290 (A3, H220)”
Installation
The refrigerating unit is air-cooled, so sufficient clearance and cool intake air must be provided at the installation site.
Air-cooled condensers need unobstructed airflow; poor clearance reduces cooling capacity and can trip the unit.
Source
“Refrigeration machine
air-cooled, natural
refrigerant”
Performance
Cooling capacity is only 0.05 kW at -20 °C, so the unit is unsuitable for sustained high heat loads at the low end of its range.
Knowing the low-temperature cooling capacity prevents oversizing expectations for exothermic reactions near -20 °C.
Source
“at -20°C
0,05 kW”
Installation
The power supply requirement is 208-240 V single or two-phase 50/60 Hz, so the site must provide the correct mains configuration.
Mismatched supply voltage or phase configuration prevents commissioning and may damage the unit.
Source
“Power supply requirement
208-240V 1~/2~ 50/60Hz”
Installation
The circuit must be protected with a fuse between 10 A and 16 A, matching the maximum current draw of the immersion thermostat.
Correct upstream fusing is required for safe operation and to avoid nuisance trips or inadequate protection.
Source
“min. Fuse
10A
max. Fuse
16A”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.
Sites with unstable mains must be checked against this combined tolerance rule before installation.
Source
“-5% voltage and +2% frequency -> not allowed!”
Operation
The permissible ambient temperature range is 5 °C to 40 °C, and cooling capacity may drop as ambient temperature rises.
Hot rooms degrade performance; the installation room must stay within the stated ambient limits.
Source
“If the ambient temperature rises, the cooling capacity may drop.”
Conformité
The equipment is EMC class A, group 1 per EN55011, so it is intended for industrial environments and may cause interference in residential areas.
Class A equipment has different EMC obligations and may not be suitable for residential or light-industrial locations.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.
The plug type must be stated on the order, otherwise the unit may arrive without a usable mains connection.
Source
“with power cable and country-specific plug (please specify when ordering)”
Safety
The degree of protection is IP20, so the unit must be kept away from water spray and conductive dust.
IP20 limits where the unit can be placed and rules out washdown or wet environments.
Source
“Degree of Protection
IP20”
Maintenance
The refrigerant charge is only 0.047 kg with a GWP of 0.02, giving a CO2 equivalent of 0.0 t and minimal leak-reporting burden.
Small charge and low GWP simplify refrigerant record-keeping and leak-check obligations.
Source
“Refrigerant quantity
0,047 kg”
Operation
The pressure/suction pump delivers up to 27 l/min at 0.7 bar, which sets the external circuit pressure-drop limit.
External loop resistance must be checked against this pump curve to avoid insufficient circulation.
Source
“max. delivery
27 l/min”
Selection
The bath volume is 4.5 l with a 140x120 mm opening, which constrains the size of immersed vessels and samples.
Bath volume and opening size determine whether the intended sample load fits and how fast temperature changes occur.
Source
“Bath volume
4,5 l”