Selection
The unit is a heating circulator intended for closed systems, so it cannot be used for open bath applications.
Determines whether the circulator matches the intended closed-loop process or requires a different open-bath model.
Source
“Heating Circulator for closed systems.”
Installation
The pump connection is M24x1.5 male, so mating hoses or adapters must match this thread.
Prevents ordering incompatible hoses or fittings and ensures leak-free connection at commissioning.
Source
“Pump connection
M24x1,5 male”
Installation
The electrical supply must be 200-240 V single or two-phase at 50/60 Hz with a 16 A fuse and 15.5 A maximum current.
Defines the branch circuit, protection device, and supply configuration needed before installation.
Source
“Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
15,5 A
Fuse
16 A”
Installation
Voltage and frequency tolerances are limited to ±5% voltage with simultaneous ±2% frequency, and some combinations are explicitly not allowed.
Site power quality must be checked; out-of-tolerance supply can prevent safe and reliable operation.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety
External branch circuit protection according to UL 489 is required.
The installer must provide a compliant external breaker or fuse; the unit does not include branch protection.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at order.
Avoids delivery of the wrong plug type and delays during commissioning.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The ambient temperature must remain between 5 °C and 40 °C for proper operation.
The installation location must be assessed for room temperature before commissioning.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Operation
The maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s.
Selecting a fluid with too high viscosity can overload the pump and degrade circulation.
Source
“max. permissible kin. viscosity
50 mm²/s”
Operation
The minimum filling capacity is 1.45 l and the expansion tank holds 2 l.
Determines the minimum system volume and expansion allowance needed for safe thermal expansion.
Source
“min. filling capacity
1,45 l
Filling capacity expansion tank
2 l”
Performance
The circulation pump delivers up to 45 l/min at a maximum delivery pressure of 0.9 bar.
Used to check whether the pump can overcome the pressure drop of the external circuit at the required flow.
Source
“max. delivery
45 l/min
max. delivery pressure
0,9 bar”
Conformité
The equipment is classified as pressure equipment category Art. 4.3 PED.
Affects documentation and conformity requirements for the pressure-containing circuit.
Source
“Pressure equipment category
Art. 4.3 PED”
Conformité
The disturbance classification is EN55011 Class A, Group 1.
Class A equipment may require additional EMC measures in residential or light-industrial environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The pump connection bore shape is Y (60°) per DIN 3863, with ball socket per DIN 3863 and sleeve nut per DIN 3870.
Ensures the correct hose end form and nut are used to achieve a proper seal.
Source
“Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN
3863, sleeve nut according to DIN 3870.”
Safety
The unit has adjustable overtemperature protection according to DIN 12876.
The overtemperature limit must be set appropriately during commissioning to protect the process and fluid.
Source
“With adjustable
overtemperature protection according to DIN 12876.”