Product introduction and general description of the unit
Specp. 2
Technical data according to DIN 12876
p. 2table, 16 rows
p. 1table, 28 rows
Technical data according to DIN 12876
Parameter
Value
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,08 kg
CO2-Äquvivalent
0,0 t
UN-number
UN 3358
Circulation pump:
max. delivery
47 l/min
max. delivery pressure
0,9 bar
Pump connection
M24x1,5 male
max. permissible kin. viscosity
50 mm²/s
min. filling capacity
1,5 l
Overall dimensions WxDxH **
295x530x570 mm
Power supply requirement
208-240V 1~/2~ 50/60Hz
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/25
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Characteristic
Grande Fleur-eo
Order-No.
1041.0018.01
Operating temperature range
-40...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power at 240V
2 kW
Heating power at 230V
1,8 kW
Heating power at 220V
1,6 kW
Heating power at 208V
1,5 kW
Heating power convertible
1,5 - 2 kW
Cooling power with Thermooil at 200°C
0,6 kW
Cooling power with Thermooil at 150°C
0,6 kW
Cooling power with Thermooil at 100°C
0,6 kW
Cooling power with Thermooil at 50°C
0,6 kW
Cooling power with Thermooil at 20°C
0,6 kW
Cooling power with Ethanol at 0°C
0,6 kW
Cooling power with Ethanol at -20°C
0,35 kW
Cooling power with Ethanol at -30°C
0,2 kW
Cooling power with Ethanol at -40°C
0,04 kW
Refrigeration machine
air-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit is designed only for external open operation and has no expansion tank, so the application loop must accommodate fluid filling and thermal expansion.
Determines whether an external reservoir or expansion vessel must be added to the installation.
Source
“The machine is developed exclusively for external open operation, and does not possess an expansion tank.”
Installation
Fluid filling and compensation for temperature-dependent volume changes must be provided through the externally connected application.
Requires the customer loop to be sized and arranged to accept fill and expansion volume.
Source
“The filling of the thermofluid, and allowance for the temperature dependent volume changes is only possible via an externally connected application.”
Safety
The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, which affects site classification and handling.
Flammable refrigerant classification drives ventilation, ignition-source and safety requirements at the installation site.
Source
“R-290 (A3, H220)”
Installation
The unit is water-cooled and requires cooling water intake between 5 °C and 20 °C.
Cooling water supply temperature must be verified before installation; higher temperatures reduce cooling capacity.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
If cooling water temperature falls below the condensation point, the cooling water pipes in the unit must be insulated.
Prevents condensation and corrosion on internal cooling water piping.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation
Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so suitable cooling water quality is required.
Guides cooling water quality selection to avoid corrosion or fouling of the circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Power supply quality must be checked against these limits before commissioning.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Conformité
The example states that -5% voltage with +2% frequency is not allowed, while -5% voltage with -2% frequency is allowed.
Clarifies that voltage and frequency deviations must be evaluated together, not independently.
Source
“-5% voltage and +2% frequency -> not allowed!”
Conformité
The equipment is EMC classified as Class A, Group 1 according to EN55011.
Class A equipment may require additional EMC measures in residential or sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.
The correct plug type must be ordered to match the site power outlet.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation
Three-phase devices below 63 A are supplied with cable but without plug.
Site electrical termination must be planned separately for three-phase units.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Pump connections use bore shape Y (60°) per DIN 3863, with ball socket per DIN 3863 and sleeve nut per DIN 3870.
Ensures compatible hoses and fittings are selected for the pump connection.
Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation
The maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s.
Limits the choice of heat transfer fluid, especially at low temperatures.
Source
“max. permissible kin. viscosity 50 mm²/s”
Operation
The minimum filling capacity is 1.5 l.
The external application and piping must provide at least this fluid volume for safe operation.